F. Film/screen compatibility—Color of light emission test

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/10

encourage image

There's no tags or description

Looks like no tags are added yet.

11 Terms

1
New cards

100 cm

set the FFD to blank

2
New cards

dim

(blank) the room lights to clearly observe the screen.

3
New cards

80 kV and 10s

Make an exposure using:

• High KV:

• Long exposure time (max time possible: blank)

4
New cards

note the color of emitted light and note the intensity

Observe the screen during exposure:

  • note blank

  • note blank


5
New cards

color of emitted light should match film sensitivity and intensity should be consistent across similar screen type

Evaluation: (2)

6
New cards

screen consistency test

Corrective Actions:

• If color mismatch is found:

→ Replace the film with one compatible with the screen's emission.

• If significant variation in intensity is observed between screens of the same type:

→ Check cassette labels and/or conduct blank

→ Re-label cassettes accurately as needed.



7
New cards

Calcium Tungstate and Rare Earth

two classification of IS

8
New cards

Blue light, has crystals, Higher than 0.2s or higher than 20 mAs

Calcium Tungstate (CaWO4)

Light color emitted:

Physical structure: (suspended in a phosphor layer)

Exposure Requirements: Needs longer exposure time (s) and higher dose



9
New cards

Green light, Gadolinium Oxysulfide and Lanthanum Oxybromide, lower than 0.05s or lower than 5 mAs

Rare Earth 

Light color emitted:

Chemical Compositions:

Exposure Requirements: Needs shorter exposure



10
New cards

color of light emitted = film sensitivity

SPECTRAL MATCHING FORMULA

11
New cards

unexposed film or image blur

if spectral mismatch occur, this leads to (2):