Vocabulary flashcards covering medical gas cylinder identification, charging procedures, duration of flow calculations, bulk oxygen installations, central piping architecture, and therapeutic gas mixtures.
Asterisk (*) Cylinder Marking
A stamp on a medical gas cylinder indicating that the cylinder is approved for hydrostatic retesting every 10 years rather than the standard 5-year interval.
Plus (+) Cylinder Marking
A stamp on a gas cylinder indicating that the cylinder is approved to be filled to 10% above its rated service pressure.
Gas Cylinder Charging
The process of filling a compressed gas cylinder directly to its specified service pressure at a reference temperature of 70∘F.
Liquid Cylinder Charging
The process of filling a liquefied gas cylinder according to specified filling density and critical temperature rather than pressure alone.

Measuring Cylinder Contents (Gas vs. Liquid)
Gas cylinder contents are monitored by checking gauge pressure against full pressure; liquid cylinder contents must be weighed because vapor pressure remains constant until all liquid is depleted.
E Cylinder Factor
The factor used for calculating flow duration for an E-size oxygen cylinder, equal to 0.28L/psig.
H Cylinder Factor
The factor used for calculating flow duration for an H-size oxygen cylinder, equal to 3.14L/psig.
Duration of Gas Flow Formula
The formula used to calculate remaining delivery time: Duration of flow (min)=Flow (L/min)Cylinder Pressure (psig)×Cylinder Factor.
Change Out Pressure
A required safety margin of 500psig at which a medical gas cylinder must be removed from patient use and replaced with a full cylinder.
Usable Cylinder Gas
The available pressure in a cylinder when factoring in patient safety, calculated by subtracting the 500psig change out pressure from the current gauge pressure (e.g., 2200psig−500psig=1700psig).
Liquid Oxygen Expansion Ratio
The volume expansion property where 1cu.ft. of liquid oxygen expands to produce 860cu.ft. of oxygen gas.
Bulk Oxygen System
A centralized supply system used in large healthcare facilities holding at least 20,000cubic feet of gas (including reserves), most commonly stored in liquid form.
Alternating Supply System (Cylinder Manifold System)
A medical gas supply system consisting of banks of large compressed gas cylinders manifolded together into a primary side and a reserve side, primarily utilized in small facilities.

Cylinder Supply System with Reserve
A gas supply system featuring a primary supply, a secondary supply, and an insulated vacuum container reserve supply of compressed or liquid gas.
Critical Temperature of Oxygen
The physical threshold of −118.8∘C (−181.4∘F) below which oxygen must be kept in large bulk stand tanks to remain liquefied at a working pressure of 250psig.

Liquid Oxygen Stand Tank
A bulk storage device consisting of an inner insulated vessel with a near vacuum, a vaporizer coil system, and a pressure-reducing valve to convert cold liquid into gaseous oxygen.
NFPA Reserve Requirement
A National Fire Protection Agency safety standard requiring the emergency reserve of a bulk oxygen supply system to equal at least one full day's average hospital usage.
Central Piping System
A hospital distribution network that conveys gases from bulk supply through main alarm panels, risers, zone valves, and terminal wall station outlets.

Zone Valves
Piping valves installed throughout hospital units that allow immediate shutoff of medical gases to specific wings, surgical suites, or floors during fire emergencies or maintenance.
HeliOx
A medical gas mixture of helium and oxygen (standardized as 80%/20% or 70%/30% He/O2) whose low gas density decreases airway resistance to assist breathing in acute asthmatics.