1/40
Comprehensive vocabulary flashcards covering the physical principles, formulas, contrast agents, and comparative clinical applications of X-ray, CT, Ultrasound, and MRI from the ENGIN 114 Bioimaging lecture.
X-rays
A form of high-energy ionizing electromagnetic radiation used in medical imaging.
Wilhelm Röntgen
The physicist who discovered X-rays in 1895.
X-ray Photon Energy Formula
E=hν=λhc, where energy increases as photon wavelength decreases.
Planck's Constant (h)
h=6.626×10−34J⋅s.
Speed of Light (c)
c=3.00×108m/s.
Ionizing Radiation
Radiation with sufficient energy to remove electrons from atoms and potentially damage DNA.
Conventional X-ray Image Formation
An imaging process where X-rays pass through tissue and a detector measures the transmitted radiation.
Tissue Appearance on X-ray
Bone appears white because it attenuates more X-rays; soft tissue appears gray; air appears dark.
Attenuation
The reduction of X-ray beam intensity as it passes through matter.
Beer's Law (X-ray Attenuation)
I=I0e−μx, where I is transmitted intensity, I0 is initial intensity, μ is the attenuation coefficient, and x is material thickness.
X-ray Attenuation Coefficient Factors
The three physical factors governing the attenuation coefficient (μ): photon energy, atomic number, and physical density.
Half-Value Layer (HVL)
The thickness of material that reduces transmitted beam intensity to 50%, calculated as HVL=μln(2).
Transmitted Fraction Formula (HVL)
(21)HVLthickness; after three HVLs, 12.5% of the beam is transmitted and 87.5% is removed.
Computed Tomography (CT)
An imaging modality that collects X-ray projections from many angles using a rotating gantry to reconstruct cross-sectional slices and 3D representations.
CT Scanner Components
The four main physical hardware components: X-ray tube, detectors, rotating gantry, and patient table.
CT Projection
A set of X-ray attenuation measurements acquired along paths through the body at a specific angle.
Filtered Back-Projection
A standard mathematical CT reconstruction method that filters and back-projects measured attenuation profiles.
CT Radiation Dose
Radiation exposure that is higher than a single conventional radiograph (e.g., approximately 7mSv for a chest CT compared to approximately 0.1mSv for a chest X-ray).
Iodine-Based Contrast
A contrast agent commonly administered during CT scans to evaluate blood vessels, such as in CT pulmonary angiography.
Ultrasound
A non-ionizing imaging modality that forms real-time images using high-frequency sound waves and returning echoes.
Medical Ultrasound Frequency Range
Typically between 2MHz and 15MHz.
Piezoelectric Transducer
A device containing piezoelectric material that converts electrical energy into sound waves and converts returning echoes back into electricity.
Lead Zirconate Titanate (PZT)
A widely used piezoelectric material in ultrasound transducers.
Ultrasound Depth Formula
d=2ct, where c is sound speed and t is round-trip echo time, divided by 2 because sound travels to the reflector and back.
Speed of Sound in Soft Tissue
Standard assumed velocity of approximately 1540m/s, yielding an echo round-trip time of about 130μs for a structure at a depth of 10cm.
Acoustic Impedance (Z)
The product of physical density and sound velocity (Z=ρc); boundaries with large differences in impedance produce strong reflections.
Ultrasound Acoustic Gel
A coupling medium placed on the skin to eliminate air gaps, preventing the approximately 99.9% beam reflection that occurs at a soft tissue-air boundary.
Ultrasound Frequency Tradeoff
Higher frequencies provide better spatial resolution but poorer penetration depth; lower frequencies provide deeper tissue penetration but poorer resolution.
Microbubbles
Gas-filled microscopic spheres used as intravenous contrast agents in ultrasound imaging.
Magnetic Resonance Imaging (MRI)
A non-ionizing imaging technique based on nuclear magnetic resonance, utilizing strong static magnetic fields, gradient fields, and radiofrequency pulses.
Main Magnetic Field (B0)
The primary static magnetic field in an MRI scanner that establishes a net longitudinal alignment among hydrogen magnetic moments.
Radiofrequency (RF) Pulse
An electromagnetic pulse used in MRI to tip net proton magnetization away from the static field (B0) into the transverse plane.
Larmor Resonance Frequency Formula
f=γB, defining the precession frequency of nuclear spins proportional to magnetic field strength.
Hydrogen Gyromagnetic Ratio (γ)
Approximately 42.58MHz/T, resulting in resonance frequencies of approximately 63.9MHz at 1.5T and 127.7MHz at 3T.
T1 Relaxation
The process describing the recovery of longitudinal magnetization back along the axis of B0.
T2 Relaxation
The process describing the decay and loss of phase coherence among spins in the transverse plane.
T1-Weighted MRI
An MRI contrast weighting where fluids appear dark and fat appears bright, preferred for demonstrating anatomical detail.
T2-Weighted MRI
An MRI contrast weighting where water and fluids appear bright ('water is white on T2'), preferred for demonstrating edema, fluid collections, and inflammation.
Gadolinium-Based Contrast
A paramagnetic chemical contrast agent administered intravenously to alter local relaxation rates in MRI.
MRI Projectile Hazard
The safety risk where external ferromagnetic objects experience intense translational force and are violently pulled toward the bore of the magnet.
RF Tissue Heating
A biological safety concern in MRI resulting from the absorption of radiofrequency power by patient tissues.