FDA

fluoro is regulated by what part of the US government

2
New cards

2.0mm Pb eq

tower housing must have what amount of lead equivalent

3
New cards

0.25mm Pb eq

the bucky slot cover & curtain must have what amount of lead equivalent

4
New cards

21 mGy/min per mA at 80 kVp

tabletop intensity limit is what

5
New cards

100 mGy/min

max exposure rate for normal fluoro

6
New cards

200mGy/min

max exposure rate for high dose fluoro

7
New cards

15” (38cm)

source-to-skin distance minimum for stationary

8
New cards

12” (30cm)

source-to-skin distance minimum for mobile

9
New cards

2.5mm Al eq. over 70kVp

fluoro tube filtration minimum

10
New cards

3%

collimators must restrict the beam input size to within ____

11
New cards

intensifying screens

part of the film world, film/screen contact tests were conducted to limit blur

12
New cards

protective lead apparel

need to be checked annually, fluoro or high kVp radiography or CT

13
New cards

film illuminators

viewbox illumination- measured by luminance

14
New cards

NCRP (national council of radiation protection)

sets dose limits & protective equipment standards

15
New cards

0.5mm Pb eq

lead aprons & thyroid collars must meet what requirement

16
New cards

50mSv/year

what is the yearly dose limit set by the NCRP

17
New cards

Society of Motion Picture and Television Engineers

SMPTE

18
New cards

SMPTE (Society of Motion Picture and Television Engineers)

pattern is used to observe luminance patches

19
New cards

National Electrical Manufacturers Association

NEMA

20
New cards

Digital Imaging and Communication

DICOM

21
New cards

NEMA-DICOM

gray scale display function (GSDF)- consistent grayscale appearance; GSDF (is mandated for all digital display devices in 1998 (including flat panel displays)

22
New cards

Deutsches Institut

DIN 2001

23
New cards

DIN 2001

publication-developed to address requirements for digital displays

24
New cards

video electronics standard association

VESA

25
New cards

VESA

set of instructions that can be used to evaluate system performance

26
New cards

American Association of Physicists in Medicine (Task group 18)

AAPM TG 18

27
New cards

AAPM TG 18 tests

developed by the AAPM and ACR for teleradiology, AAPM TG patterns can be used for QC - daily monitoring

28
New cards

26

AAPM Report No. OR-03 has _____ test patterns to assess display characteristics- distortion, reflection, luminance, noise, and spatial resolution of the monitors

29
New cards

daily tests

visual inspection of IR and monitors, CR-PSP reader check by erasing plate, FPD IR’s as recommended by manufacturer

30
New cards

monthly tests

some facilities erase all PSPs monthly, workstation calibrations

31
New cards

quarterly tests

clean and inspect all PSP plates and cassettes, more thorough QC tests, ______ repeat rate

32
New cards

annual tests

medical physicist will conduct and review test data

33
New cards

General Image Quality: AAPM TG18-QC

looks for general image quality, distortion & artifacts; assesses for variations in the shape of the displayed image from the original (distortions appear blurry)

34
New cards

Reflection: AAPM TG18 - AD

  • evaluates how ambient light affects ability to see structures on the monitor

  • ideally, the light intensity from the monitor is the only light it produces, as additional lights can degrade the ability to see clearly

  • monthly or quarterly


35
New cards

Display Luminance: when viewed at perpendicular angle

  • assesses the displayed luminance values vs the input values from the display system

  • max luminance > 171 cd/m2

  • monthly or quarterly


36
New cards

Luminance Dependencies: AAPM TG18-UNL10(A) and TG18UNL80 (B)

  • this test evaluates the image for non-uniformity and effects of viewing at different angles

  • monthly or quarterly


37
New cards

Display Spatial Resolution: AAPM TG18-QC & AAPM TG18-CX

assesses the system’s ability to display images of different parts of an image with high fidelity - sharpness

  • measures in Lp/mm

  • monthly or quarterly


38
New cards

Display Noise: AAPM TG18-AFC

  • pattern contains large number of regions with changing target positions with constant size and contrast

  • looks at white dot series and visibility

  • monthly or quarterly


39
New cards

1-megapixel (Mp)

general viewing monitor minimum requirement (like tech displays)

40
New cards

2 Mp

general radiologist interpretation/quality control min. requirement

41
New cards

5 Mp

interpretation of digital mammograms min. requirement

42
New cards

veiling glare

light that is scattered inside the display monitor that negatively affects image quality

43
New cards

chromaticity

is whether a color monitor’s display is uniform, or uniformly gray across the test pattern