What is the purpose for protective tube housing in diagnostic testing?

reduce leakage radiation

2
New cards

In diagnostic testing, leakage radiation must not measure more than.....

1 mGya/hr at 1 meter from protective housing

3
New cards

In diagnostic testing, what does the control panel positively indicate?

conditions of exposure

4
New cards

What can you see on the control panel?

mAs and kVp

5
New cards

During exposure when the beam is energized, you notice_________during exposure.

audible sound and visual signal

6
New cards

A tape measure and laser are two ways to measure....

SID (source to image distance)

7
New cards

The tape measurer or laser must be accurate within ___ percent of the SID?

2%

8
New cards

What is another name for the xray beam or light beam?

collimation

9
New cards

What must coincide within 2% of SID?

x-ray beam / collimation

10
New cards

PBL stands for.....

positive beam limitation

11
New cards

T or F: Positive Beam Limitation (PBL) is currently mandatory in diagnostic testing?

false

12
New cards

A PBL must be accurate to within ___% of the SID.

2%

13
New cards

What is it called when the center of image receptor and the x-ray beam are aligned?

beam alignment

14
New cards

What is true about filtration in diagnostic testing?

all-purpose rooms must have at least 2.5mm Al equivalent

15
New cards

What decreases exposure of the patient in diagnostic testing?

filtration

16
New cards

What is the required filtration equivalence when the kVp is set above 70?

2.5mm Al equivalence

17
New cards

1.5mm Al equivalence is used when the tube uses what kVp, regarding filtration?

kVp between 50 and 70

18
New cards

What is the required filtration equivalence when the kVp is set below 50?

0.5mm Al equivalence

19
New cards

How do we measure filtration for diagnostic testing?

HVL - half value layer

20
New cards

How many micro-meters of Molybdenum (Mo) is used for filtration in Mammography?

30

21
New cards

How many micro-meters of Rhodium (Rh) is used for filtration in Mammography?

60

22
New cards

Repeated exposures at a given technique must provide consistent intensity.

Reproducibility

23
New cards

Referring to reproducibility the difference in intensity during reproducibility should only change within........

5% or less

24
New cards

what is the formula linearity

mA x s (time)

25
New cards

what does linearity mean

different combinations but the same amount of intensity

26
New cards

Linearity must not deviate more than____ in diagnostic testing?

10%

27
New cards

How is intensity expressed in linearity?

mGya/mAs

28
New cards

how long/short is the cord for the exposure switch in the imaging department?

as short as possible

29
New cards

A 6 foot long cord is used with what kind of machine?

portable / mobile machine

30
New cards

When using a portable xray machine the cord must be at least 6 feet, ____cm, or _____meters long.

180cm / 2m

31
New cards

What does SSD mean?

source to skin distance

32
New cards

SSD for fluoroscopy must be_____cm for stationary or in room examinations

no less than 38 cm

33
New cards

SSD for a portable (C-arm) fluoroscopy must be_______cm for a portable examinations.

no less than 30 cm

34
New cards

In fluoroscopy, the dose is ______ when the tube is closer to the patient's skin

higher

35
New cards

In fluoroscopy, the dose is lower when the tube is moved ______ from the patient's skin

further

36
New cards

PPB means....

primary protective barrier

37
New cards

The image intensifier is part of the PPB and must be _______ equivalent

2mm PB equivalent

38
New cards

In fluoroscopy, the _____ and _____ must be connected and able to move together

Image Intensifier and tube

39
New cards

When the image intensifier is in park, xrays_______

cannot be produced

40
New cards

Filtration in fluoroscopy must be at least________equivalent

2.5mm Al

41
New cards

Anything that is between the _____ and the _____counts as part of the 2.5mm Al equivalent for fluoroscopy filtration

xray tube and patient

42
New cards

Filtration in fluoroscopy is measured by______

HVL

43
New cards

When collimating in fluoroscopy, what should you be able to see on the image monitor?

an unexposed (dark) visible border

44
New cards

What type of switch is an exposure switch?

dead-man type

45
New cards

When your foot in not touching the exposure (dead-man) switch, xrays ______ be produced

cant

46
New cards

In fluoroscopy, what is the opening on the table where the IR is placed?

bucky slot cover

47
New cards

All area that the IR does not occupy is considered to be_______

open space

48
New cards

The bucky slot cover is covered in Pb during fluoroscopy exams so that the _____ and _____ are protected

radiologist and technologist

49
New cards

The bucky slot cover must be at least _________ in fluoroscopy

0.25 Pb equivalent

50
New cards

The protective curtain must be _______ equivalent and reduces exposure to the _____

0.25 Pb; technologist

51
New cards

What indicates how long the fluoro has been on or energized for?

cumulative timer

52
New cards

In what intervals does the timer go off for?

5 minutes

53
New cards

The timer produces audible sound and ______ exposure

interrupts

54
New cards

The x-ray beam intensity at the tabletop during fluoroscopy must not exceed _____mGya/minute for each mA at _______kVp

20; 80

55
New cards

The x-ray beam intensity at tabletop with NO High-Level Control should not exceed__________/minute

100 mGya/minute

56
New cards

The x-ray beam intensity at tabletop WITH High-Level Control should not exceed__________/minute

200 mGya/minute

57
New cards

T of F: There is no limit on xray intensity when the image is recorded?

true