What is the purpose for protective tube housing in diagnostic testing?
reduce leakage radiation
In diagnostic testing, leakage radiation must not measure more than.....
1 mGya/hr at 1 meter from protective housing
In diagnostic testing, what does the control panel positively indicate?
conditions of exposure
What can you see on the control panel?
mAs and kVp
During exposure when the beam is energized, you notice_________during exposure.
audible sound and visual signal
A tape measure and laser are two ways to measure....
SID (source to image distance)
The tape measurer or laser must be accurate within ___ percent of the SID?
2%
What is another name for the xray beam or light beam?
collimation
What must coincide within 2% of SID?
x-ray beam / collimation
PBL stands for.....
positive beam limitation
T or F: Positive Beam Limitation (PBL) is currently mandatory in diagnostic testing?
false
A PBL must be accurate to within ___% of the SID.
2%
What is it called when the center of image receptor and the x-ray beam are aligned?
beam alignment
What is true about filtration in diagnostic testing?
all-purpose rooms must have at least 2.5mm Al equivalent
What decreases exposure of the patient in diagnostic testing?
filtration
What is the required filtration equivalence when the kVp is set above 70?
2.5mm Al equivalence
1.5mm Al equivalence is used when the tube uses what kVp, regarding filtration?
kVp between 50 and 70
What is the required filtration equivalence when the kVp is set below 50?
0.5mm Al equivalence
How do we measure filtration for diagnostic testing?
HVL - half value layer
How many micro-meters of Molybdenum (Mo) is used for filtration in Mammography?
30
How many micro-meters of Rhodium (Rh) is used for filtration in Mammography?
60
Repeated exposures at a given technique must provide consistent intensity.
Reproducibility
Referring to reproducibility the difference in intensity during reproducibility should only change within........
5% or less
what is the formula linearity
mA x s (time)
what does linearity mean
different combinations but the same amount of intensity
Linearity must not deviate more than____ in diagnostic testing?
10%
How is intensity expressed in linearity?
mGya/mAs
how long/short is the cord for the exposure switch in the imaging department?
as short as possible
A 6 foot long cord is used with what kind of machine?
portable / mobile machine
When using a portable xray machine the cord must be at least 6 feet, ____cm, or _____meters long.
180cm / 2m
What does SSD mean?
source to skin distance
SSD for fluoroscopy must be_____cm for stationary or in room examinations
no less than 38 cm
SSD for a portable (C-arm) fluoroscopy must be_______cm for a portable examinations.
no less than 30 cm
In fluoroscopy, the dose is ______ when the tube is closer to the patient's skin
higher
In fluoroscopy, the dose is lower when the tube is moved ______ from the patient's skin
further
PPB means....
primary protective barrier
The image intensifier is part of the PPB and must be _______ equivalent
2mm PB equivalent
In fluoroscopy, the _____ and _____ must be connected and able to move together
Image Intensifier and tube
When the image intensifier is in park, xrays_______
cannot be produced
Filtration in fluoroscopy must be at least________equivalent
2.5mm Al
Anything that is between the _____ and the _____counts as part of the 2.5mm Al equivalent for fluoroscopy filtration
xray tube and patient
Filtration in fluoroscopy is measured by______
HVL
When collimating in fluoroscopy, what should you be able to see on the image monitor?
an unexposed (dark) visible border
What type of switch is an exposure switch?
dead-man type
When your foot in not touching the exposure (dead-man) switch, xrays ______ be produced
cant
In fluoroscopy, what is the opening on the table where the IR is placed?
bucky slot cover
All area that the IR does not occupy is considered to be_______
open space
The bucky slot cover is covered in Pb during fluoroscopy exams so that the _____ and _____ are protected
radiologist and technologist
The bucky slot cover must be at least _________ in fluoroscopy
0.25 Pb equivalent
The protective curtain must be _______ equivalent and reduces exposure to the _____
0.25 Pb; technologist
What indicates how long the fluoro has been on or energized for?
cumulative timer
In what intervals does the timer go off for?
5 minutes
The timer produces audible sound and ______ exposure
interrupts
The x-ray beam intensity at the tabletop during fluoroscopy must not exceed _____mGya/minute for each mA at _______kVp
20; 80
The x-ray beam intensity at tabletop with NO High-Level Control should not exceed__________/minute
100 mGya/minute
The x-ray beam intensity at tabletop WITH High-Level Control should not exceed__________/minute
200 mGya/minute
T of F: There is no limit on xray intensity when the image is recorded?
true