what is atomic absorption methodology?
used to determined the concentration of elements in their solution by vaporizing them and measuring their absorption of characteristic wavelengths
what is used to convert sample to atomic vapor?
high energy source
when sample is converted to a atomic vapor, what state are the electron in?
mostly ground state but some excited state electron
what is being measured in atomic absorption?
wavelengths are produced of the element being measured
what is the relationship between amount of light emitted and concentration of sample?
inversely proportional
______ concentration of sample the ______ intensity of light that is emitted
greater; lesser
how is the change of intensity of light measured by?
photodetector
What is the actual measurement taken that reflects the concentration of the analyte that you are trying to quantify (are you measuring absorption or transmittance)?
The light is inversely proportional to the concentration. (you are measuring transmittance)
what type of cation does the sample need to be to atomized easily?
divalent or trivalent
what are divalent cations?
any cation in the 2+ state
Ca2+, Cu2+, Mg2+
what are trivalent cations
any cation in the 3+ state
Fe3+, Au3+
what happens when molecules are burned?
complexes are disassociated, chemical bonds are broken and atoms are released in an unexcited ground state
what happens when light passes through ground-state atoms?
atoms absorb light at characteristic wavelengths and become excited (transmission)
what is a ground-state atom?
atoms whose electrons are at their lowest energy levels
what happens when an atom absorbs light?
electron gains energy and moves to a higher energy level
what is an excited-state atoms?
an atom with an electron raised to a higher energy level
what type of wavelength do atoms absorb?
characteristic wavelength specific to the element
how is light absorbed related to analyte concentration?
directly proportional
more atoms = more light absorbed
what happen to absorbance as analyte concentration increases?
absorbance increases
how is light emitted related to atom concentration?
inversely proportional
what does atomic absorption spectroscopy measured?
the amount of characteristic light absorbed by ground state atoms
if fewer atoms are present what happen to absorbance?
absorbance decreases because fewer atoms absorb light
What is the state of the atoms (ground or excited) produced after being vaporized by the heat source?
the ground state
What is the state of the atoms (ground or excited) when they absorb light?
a transition between the ground and excited states
What is the state of the atoms (ground or excited) when light is being emitted?
an excited state
what are the components of atomic absorption?
light sources
nebulizer
energy sources (atomizer)
monochromator
detector
read-out device
what are some examples of light sources
hollow cathode lamp
laser
what is the most common light source for atomic absorption?
hollow cathode lamp
what are some examples of energy sources (atomizer)
graphite furnance or burner
flame
what happens in the nebulizer?
where samples are introduced into the atomizer
direct injection point
creating aerosol of speciment
what happens in the atomizer?
converting aerosolized sample into a vapor (gas)
what is the most common atomizer (energy source)?
graphite furnace or burner
very consistent high energy source
what is a monchromator
isolates the wavelength of the light that is needed to be measured
exit slit = bandwidth, ideal wavelength of light is going to pass out
bandwidth side can be adjusted.
what does a wider bandwidth mean?
increases the intenisty of light that is pass through but decrease resolution
what does a smaller bandwidth mean?
resolution will increase but decrease the intensity
what first happens to sample in atomizer?
aspirated into the heat source and then converted to a gaseous state
the ______ temperature produces atoms in ______ state
high; ground
what is used to produce a high temperature (2700C)
graphite furnace or burner
what are advantages of graphite furnace than a flame?
uniform heating
increased sensitivity
lower detection limit
what is needed for each analyte?
serparate hollow cathode
what does each cathode lamp incude?
filament containing the same metal being tested (The reason a separate hollow cathode lamp is needed for each analyte)
rare gas, argon helium or neon
why is a rare gas needed in each cathode lamp?
electrical discharged produced by the power source ionizes the rare gas atoms, causing collision of the gas and metal atoms
the element of interest is separated and placed in its unexcited ground state
what causes collision of gas and metal atoms?
electrical discharged produced by the power sources ionzes the rare gas atoms
what happens when metal is present?
some of the light will absorbed
what is being measured from atomizer
intensity of light emitted
the _____ concentrated the metal in sample the _____ light will be absorbed
more; more
________ _________ curves are compared to the sample result
standard calibration
what is used to analyze many metals?
atomic absorption spectrophotometry (AAS)
what heavy metals are present in body fluids?
iron (Fe), copper (Cu), aluminum (Al), lead (Pb), arsenic (Ar), calcium (Ca), zinc (Zn), and cadmium (Cd)
what else can heavy metal analysis used for
fertilizers and industrial waste
what are advantages of AAS?
quantitative
AAS is a sensitive technique that determine the concentration of most elements at the part-per-million (ppm) level
what are disadvantages of AAS?
furnace may not be able to separate all elements in the sample, necessary for absorption to occur
matrix interference due to the composition of the standard varying greatly from the test sample
If instruments use flame as a heat source it is an inconsistent heat source due to flickering.
How is resolution and light intensity may be affected by adjusting the bandwidth to be larger?
greater intensity of light and lower resolution
How is resolution and light intensity may be affected by adjusting the bandwidth to be smaller?
Less intensity of light but increase the resolution
What is the most common energy (atomizer) source used in atomic absorption methods?
a graphite furnace or burner.
What are the advantages of using a graphite furnace vs. a flame as the atomizer component in atomic absorption methods?
the furnace has uniform heating, increased sensitivity, and a lower detection limit