22 Terms

1
New cards

detector efficiency

efficiency of rad detector is its ability to detect certain isotopes or radiation energies

2
New cards

efficiency can vary based on what factors

source-detector geometry

source-detector distance

source strength

3
New cards

contamination limits

tied to a regulatory limit or facility set limit

unrestricted area is area where nonradiation workers have access

4
New cards

low: ALI…

over 5000 uCi

5
New cards

moderate: ALI…

bw 500-5000 uCi

6
New cards

high: ALI…

bw 5-500 uCi

7
New cards

very high: ALI…

less than 5 uCi

8
New cards

tc99m ALI

80000

9
New cards

f18 ALI

50000

10
New cards

i131 ALI

30

11
New cards

lu177 ALI

2000

12
New cards

ac225 ALI

0.3

13
New cards

trigger levels

predefined count rates or measured counts that indicate decontamination must be performed

14
New cards

minimum detectable amount (MDA)

level at which you can say with 95% confidence that you can distinguish a count/count rate from background

aka, smallest count that your detector can “see”

15
New cards

basis of MDA

rad detection roughly follows normal distribution; there is some randomness in decay of source and mechanisms of detector

16
New cards

when decontaminating, we must be sure that MDA is…

lower than our contamination limit

17
New cards

MDA: as background increases…

MDA increases

18
New cards

MDA: as efficiency increases,

MDA decreases

19
New cards

MDA: as count time increases…

MDA decreases

20
New cards

NIST sources

national institute of standards and technology

traceable sources are calibrated in reference to this company

21
New cards

NIST sources have very…

low uncertainty in their measurements; you can trust that they represent the activity that it says

22
New cards

NIST traceable sources are required…

for certain calibrations, such as dose calibrator QA/QC tests