Biomedical Equipment Technician (BMET)
A technician who supports the safety, reliability, functionality, and maintenance of medical equipment used in healthcare.
Healthcare Technology Management (HTM)
The management of healthcare technology throughout its lifecycle, including acquisition, inventory, maintenance, risk management, service, and retirement.
Incoming inspection
The initial evaluation of newly received equipment before it is placed into service, including applicable electrical safety, operational, functional, and documentation checks.
Original Equipment Manufacturer (OEM)
The company that originally manufactures a medical device or piece of equipment.
Instructions for Use (IFU)
The manufacturer's instructions describing how a device is intended to be used, including relevant safety, operation, cleaning, and maintenance information.
Preventive Maintenance (PM)
Scheduled maintenance intended to help maintain equipment performance, reliability, and safety and to reduce the likelihood of failure.
Corrective Maintenance (CM)
Maintenance performed to address a malfunction, defect, or other identified problem with equipment.
Medical Equipment Management Plan (MEMP)
An organizational plan for managing medical equipment, including policies, inventory, maintenance, safety, risk management, and related responsibilities.
Alternative Equipment Maintenance (AEM)
A maintenance program that allows an organization, where permitted, to use an alternative maintenance strategy or schedule instead of the manufacturer's recommendations under applicable requirements and risk-based justification.
Clinical engineering
An engineering and technical discipline focused on healthcare technology and its safe, effective use in clinical environments.
Medical equipment lifecycle
The stages through which equipment passes, including planning, acquisition, acceptance, inventory, maintenance, use, evaluation, and retirement.
Equipment inventory
An organized record of healthcare equipment, including identifying information, location, status, and relevant maintenance or management information.
CMMS (Computerized Maintenance Management System)
A computerized system used to manage maintenance activities, equipment records, work orders, service history, and related operational information.
Work order
An organized record of a maintenance request or service activity, including reported concerns, work performed, findings, status, and applicable documentation.
Equipment recall
An action involving a medical device identified as potentially unsafe, defective, or otherwise requiring corrective action, often initiated or communicated by the manufacturer or regulator.
Medical device hazard alert
Communication about a potential risk, safety issue, or device-related concern that may require evaluation or action.
Equipment incident reporting
The process of documenting and communicating events involving equipment that may have contributed to injury, risk, malfunction, or other safety concerns.
Medical device intended use
The purpose and function for which a medical device is designed and designated to be used.
Clinical device versus diagnostic device
A clinical device may support, treat, or monitor patient care, while diagnostic equipment is used to evaluate or identify physiological conditions or disease; some devices overlap categories.
Life-support equipment
Equipment that supports or sustains critical physiological functions, where failure may create serious or immediate patient risk.
Life-sustaining versus life-supporting equipment
Life-sustaining equipment supports essential functions required to sustain life; life-supporting equipment may assist critical functions but the terms and classifications can vary by organizational or regulatory context.
Equipment risk classification
Categorization of equipment based on factors such as function, potential consequences of failure, patient impact, and organizational risk criteria.
Why is equipment classification important in HTM?
It helps guide maintenance priorities, risk assessment, inspection, monitoring, and resource allocation.
Equipment status
The documented condition or disposition of equipment, such as available, in service, under evaluation, out of service, or otherwise designated according to organizational procedures.
Why must equipment status be accurate?
Incorrect status may result in inappropriate use, confusion, missed maintenance, or failure to communicate safety concerns.
Why is manufacturer documentation important?
It provides device-specific specifications, instructions, safety information, maintenance requirements, and applicable procedures that support appropriate servicing and evaluation.
Why does incoming inspection matter?
It helps establish that newly acquired equipment has been appropriately evaluated, documented, and prepared before entering service.
Medical device post-market surveillance
Monitoring medical devices after they enter the market to identify safety concerns, adverse events, trends, or other issues requiring attention.
FDA medical device oversight
The U.S. Food and Drug Administration regulates medical devices through applicable requirements involving safety, effectiveness, manufacturing, labeling, reporting, and other regulatory processes.
FDA 510(k) clearance
FDA clearance based on a device demonstrating substantial equivalence to a legally marketed predicate device through the applicable 510(k) pathway.
FDA PMA (Premarket Approval)
An FDA regulatory pathway generally used for certain higher-risk medical devices that requires evidence supporting safety and effectiveness.
FDA device recall versus device clearance
Clearance or approval relates to a device's regulatory pathway before or during market authorization; a recall is an action addressing a device-related issue after distribution or use.
Safe Medical Devices Act (SMDA)
The 1990 law associated with strengthening medical device safety oversight, reporting, tracking, and post-market surveillance.
Patient-connected medical equipment
Equipment that interacts directly or indirectly with a patient through monitoring, treatment, sensing, or another function, potentially requiring special attention to patient safety and electrical or mechanical hazards.
Equipment safety versus equipment effectiveness
Safety concerns whether a device presents unacceptable risk; effectiveness concerns whether it appropriately performs its intended function. A device can have issues with one, the other, or both.
Medical device reliability
The ability of equipment to perform its intended function consistently under specified conditions over time.
ECG (Electrocardiograph)
A medical device that records the electrical activity of the heart to help evaluate cardiac rhythm and electrical patterns.
ECG simulator
An instrument that generates simulated cardiac electrical signals to evaluate an ECG monitor's response to defined inputs.
Patient monitor
Equipment used to monitor physiological parameters such as heart rate, ECG, oxygen saturation, blood pressure, or other parameters depending on configuration.
Pulse oximeter
An instrument that estimates arterial oxygen saturation (SpO₂) and pulse rate using optical measurement techniques.
SpO₂
Peripheral oxygen saturation estimated by pulse oximetry.
NIBP (Noninvasive Blood Pressure)
A blood pressure measurement method that uses a cuff and an appropriate measurement system without requiring direct access to an artery.
Infusion pump
Medical equipment designed to control the delivery of fluids, medications, or other solutions into a patient.
Infusion pump analyzer
An instrument designed to evaluate applicable infusion pump performance characteristics, such as delivery-related measurements.
Ventilator
Medical equipment that supports or assists ventilation by moving gas into and out of the lungs.
Defibrillator
Medical equipment designed to deliver electrical therapy for certain cardiac emergencies.
Defibrillator analyzer
Test equipment used to evaluate applicable defibrillator performance characteristics, such as delivered electrical energy.
Electrosurgical unit (ESU)
Equipment that uses electrical energy to cut tissue or control bleeding during surgical procedures.
ESU analyzer or simulator
Test equipment used to evaluate applicable electrosurgical unit performance and electrical output characteristics.
Electrical safety analyzer
An instrument designed to evaluate applicable electrical safety characteristics of medical equipment.
Digital multimeter (DMM)
An instrument used to measure electrical quantities such as voltage, current, and resistance, depending on the instrument and configuration.
Oscilloscope
An instrument used to display and analyze electrical signals over time, allowing evaluation of characteristics such as waveform shape, amplitude, and timing.
Receptacle tester
An instrument designed to identify certain wiring or electrical outlet conditions, depending on its capabilities.
ECG simulator versus defibrillator analyzer
An ECG simulator provides simulated cardiac electrical signals for evaluating monitor response; a defibrillator analyzer evaluates characteristics of defibrillator output.
Electrical safety analyzer versus DMM
An electrical safety analyzer is designed for applicable medical electrical safety assessments; a DMM measures general electrical quantities but does not automatically perform a complete medical electrical safety assessment.
Oscilloscope versus DMM
A DMM typically provides numerical measurements of selected electrical quantities; an oscilloscope displays signal behavior over time, helping reveal waveform characteristics.
Infusion pump analyzer versus ECG simulator
An infusion pump analyzer evaluates applicable infusion delivery characteristics; an ECG simulator supplies simulated cardiac electrical signals to an ECG monitoring system.
Device analyzer versus simulator
An analyzer typically measures or evaluates device performance; a simulator generates defined inputs or signals to evaluate how a device responds.
Medical equipment tester selection
Test equipment should be selected according to the device, the characteristic being evaluated, the applicable specification, and the instrument's capabilities.
Medical equipment failure categories
Potential failure categories include electrical, mechanical, software, sensor, accessory, communication, environmental, and user-related factors, among others.
Equipment malfunction versus user error
A malfunction involves an equipment-related failure or abnormality; a use error involves incorrect or unintended use, although the two may coexist or contribute to one another.
Hazard
A condition, source, or situation that has the potential to cause harm.
Risk
The combination or consideration of the likelihood of harm occurring and the severity of its consequences.
Hazard versus risk
A hazard is a potential source of harm; risk evaluates the potential harm and its likelihood or consequences.
Hierarchy of controls
A framework for reducing hazards through progressively less effective categories of control, generally prioritizing elimination, substitution, engineering controls, administrative controls, and PPE.
Elimination
Removing a hazard entirely.
Substitution
Replacing a hazardous process, material, or item with a less hazardous alternative.
Engineering control
A physical or engineered measure designed to reduce exposure to or control a hazard.
Administrative control
A policy, procedure, training, scheduling, warning, or other organizational measure intended to reduce risk.
Personal Protective Equipment (PPE)
Equipment worn to reduce exposure to hazards, such as gloves, masks, protective eyewear, gowns, or other protective equipment.
Why are engineering controls generally preferred over PPE?
Engineering controls reduce or isolate hazards at their source or through physical systems, whereas PPE relies more heavily on correct individual use.
OSHA (Occupational Safety and Health Administration)
A U.S. agency responsible for workplace safety and health regulation and enforcement.
NIOSH (National Institute for Occupational Safety and Health)
A U.S. organization focused on occupational safety and health research and recommendations.
SDS (Safety Data Sheet)
A document providing information about a chemical's hazards, handling, storage, exposure controls, first aid, and related safety considerations.
Hazard Communication Standard (HCS)
OSHA's framework for communicating workplace chemical hazards through labeling, safety data sheets, and related requirements.
Lockout/Tagout (LOTO)
A safety process for controlling hazardous energy during servicing or maintenance to prevent unexpected energization, startup, or release of stored energy.
Stored energy hazard
Energy remaining in a system after power or another energy source has been disconnected, which may still create a hazard.
Electrical shock
The physiological effect of electrical current passing through the body.
Electrical hazard
An electrical condition that may create a risk of injury, damage, or unsafe operation.
Electrical safety
The prevention or reduction of electrical hazards through appropriate design, controls, procedures, equipment, and evaluation.
Grounding
An electrical connection to ground intended to support safety or system operation, depending on the design and application.
Protective earth
An electrical safety connection intended to help protect against electric shock by providing a protective path for fault current.
Ground fault
An unintended electrical connection between a conductor and ground or another grounded part.
Insulation
Material or a system intended to prevent or limit unwanted electrical conduction.
Leakage current
Electrical current that flows through unintended or specified leakage pathways rather than the intended normal current path, relevant to medical electrical safety evaluation.
Patient leakage current
Current that may flow through patient-connected pathways under applicable conditions and is evaluated according to relevant medical electrical safety requirements.
NFPA 70 (National Electrical Code)
A standard associated with electrical installation requirements and electrical safety in buildings and installations.
NFPA 99 (Health Care Facilities Code)
A standard addressing safety-related requirements in healthcare facilities, including applicable healthcare electrical and other systems.
NFPA 70 versus NFPA 99
NFPA 70 focuses on electrical installations and wiring requirements; NFPA 99 addresses safety requirements specific to healthcare facilities and systems.
Electrical safety analyzer purpose
An instrument used to evaluate applicable electrical safety characteristics of medical equipment under relevant testing requirements.
Infection control
Practices and measures intended to prevent or reduce the transmission of infectious agents.
Standard precautions
Infection-prevention measures applied broadly to patient care based on the possibility of exposure to infectious materials.
Transmission-based precautions
Additional infection-control measures used for patients with infections or conditions requiring specific precautions based on transmission risk.
Contact precautions
Measures designed to reduce transmission through direct or indirect contact.
Droplet precautions
Measures designed to reduce transmission through respiratory droplets.
Airborne precautions
Measures designed to reduce transmission through infectious particles that can remain suspended in air.
PPE selection principle
PPE should be selected according to the hazard, potential exposure, applicable precautions, and organizational policies.
Hand hygiene
Cleaning hands to reduce microorganisms and prevent transmission of infectious agents.
Cleaning
Removing dirt, debris, and organic or inorganic material from surfaces or objects.
Disinfection
Reducing or eliminating many pathogenic microorganisms on surfaces or objects, generally not including all bacterial spores.