stiffer tissue

With shear wave elastography, faster shear wave propagation generally indicates:

2
New cards

wall

The control that selectively removes low-frequency Doppler shifts from moving tissue like cardiac valve motion, vessel walls, or any other moving tissues is the _______ filter.

3
New cards

10 MHz

A 5 MHz transducer is used for harmonic imaging. Which frequency is received for the second harmonic?

4
New cards

reduces amplitude differences while preserving their relative order

What does compression do to echo amplitudes?

5
New cards

is halved

If vascular resistance doubles while the pressure gradient remains constant, volume flow rate:

6
New cards

+3000; towards

If transmitted frequency is 5 MHz and received frequency is 5,003,000 Hz, the Doppler shift is _____Hz and flow is moving ______ the transducer

7
New cards

PACS stores and provides access to studies; DICOM defines standards for exchanging imaging information.

What best distinguishes PACS from DICOM?

8
New cards

highest in the center and lowest near the wall

in parabolic laminar flow, blood flow velocity is:

9
New cards

80

a receiver signal has a dynamic range of 110 dB and undergoes 30 dB of compression. The resulting dynamic range is ____ dB

10
New cards

the cosine of 90 degrees is zero

why is there no Doppler frequency shift at a beam incidence angle of 90 degrees?

11
New cards

spatial compounding

combines images acquired from different beam angles

12
New cards

autocorrelation

analyzes color doppler data rapidly

13
New cards

temporal compounding

combines earlier frames with the current frame

14
New cards

doppler phantom

uses controlled moving targets to evaluate doppler performance

15
New cards

fast fourier transform

analyzes signals to create a spectral Doppler waveform

16
New cards

more gray shades and lower contrast

which image appearance is expected when the displayed dynamic range is increased?

17
New cards

tissue equivalent phantom

which QA device is most appropriate for evaluating grayscale tissue appearance and uniformity?

18
New cards

continuous wave doppler

which doppler technique measures very high velocities without aliasing but lacks range specificity?

19
New cards

4000

at a PRF of 8,000 Hz, the Nyquist limit is ____ Hz.

20
New cards

40

For voltage amplitudes of 100 V and 1 V, dynamic range is _____ dB