what are the two types of radiation?

primary and remnant

2
New cards

what is primary radiation?

-The original X-ray beam that exits the X-ray tube

- Not as strong as original photon entering patient, but

still strong enough to go straight through

- Carries vital anatomical information used to form the

visible image

3
New cards

what are the subsets of primary beam?

secondary and scatter

4
New cards

how does primary radiation travel?

travels in a straight line toward the patient

5
New cards

what is the purpose of primary radiation?.

starts the imaging process and carries the main energy needed to take the X-ray

6
New cards

within the primary beam, what happens in the patient?

secondary and scatter

-so beam is still primary until it exits the patient

7
New cards

do secondary and scatter interact?

yes, in a negative way to differing degrees

8
New cards

when is the primary beam no longer considered and is now called secondary/scatter?

when any part of the beam is affected, meaning interacts and changes direction

9
New cards

how much of the primary beam is off-focused?

25%

10
New cards

where does off-focused radiation originate?

in tube

11
New cards

where does scatter originate?

in patient and xray table

12
New cards

primary radiation reaaches the image receptor as what?

only as the remnant (exit) beam

13
New cards

what is secondary radiation?

interact inside the patient but do not scatter

14
New cards

how does secondary radiation work?

When some of the primary photons hit atoms inside the patient's body or equipment, they transfer energy and trigger new, weaker rays.

15
New cards

what is the purpose of the secondary radiation?

generally a byproduct that does not help form a clear image, especially if these photons scatter.

16
New cards

what does secondary radiation refer to?

any radiation resulting from interactions WITHIN the patient

17
New cards

what is scatter radiation?

secondary radiation that bounces off or deflects in random directions but keeps traveling

18
New cards

How does scatter radiation affect image quality?

strikes IR at incorrect angles

-fogs the image, lowers contrast, and creates noise

19
New cards

what are the 3 causes of scatter?

-high kVp

-large field sizes

-large tissue thickness

20
New cards

what is remnant(exit) radiation?

radiation that exits the patient

-it exposes the IR to create the X-ray image

21
New cards

how does remnant(exit) radiation work?

strikes the digital sensor or image receptor

22
New cards

what is the purpose of remnant (exit) radiation?

carries the actual internal pattern of bones and soft tissues to create the final diagnostic picture

23
New cards

what is remnant radiation made of?

-unattenuated primary radiation

-scatter radiation

24
New cards

what percent of the primary beam travels straight through the patient without interacting with atoms?

less than 5%

25
New cards

what percent that reaches the IR is secondary?

about 1%

26
New cards

what happens to the remaining % of radiation?

either absorbed by patient or exits in form of scatter and some reaches IR

27
New cards

what is attenuation?

the partial absorption of the x-ray beam

28
New cards

what is attenuation also referred to as?

partial absorption

29
New cards

what is the primary to remnant path of a photon?

penetrate part and go directly to expose IR

30
New cards

what is the primary to secondary to remnant path of a photon?

interact in patient, do NOT change direction and may get to IR but usually very weak photon once it interacts in patient

31
New cards

what is the primary to scatter to remnant path of a photon?

interact in patient, change direction (materially altered), scatter, and some may eventually expose IR

32
New cards

what does it mean if the photons never leave patient?

interact in patient and become totally absorbed

33
New cards

what is radiographic noise?

any unwanted information on the image that interferes with the visibility

34
New cards

what is the msot common cause of noise?

scatter

35
New cards

what is fog?

positive unwanted density characterized by a dull gray appearance, usually resulting from scatter

36
New cards

effects of fog:

_____ contrast?

decreases

37
New cards

effects of fog:

_____ density?

increased

38
New cards

effects of fog:

_____ visibility?

decreases

39
New cards

how can fog be controlled?

1. proper use of exposure factors, especially proper kVp

2. increase collimation/use of lead blockers

3. use of grids

4. compression of the part

40
New cards

whatr is a primary method to prevent scatter?

collimation

41
New cards

as thickness increases, fog?

increases; due to scatter

42
New cards

what is the density (composition) of subject matter?

bone, muscle, fat, tissue, air

43
New cards

as composition increases, fog?

increases

44
New cards

as kVp ______, scatter increases and fog from scatter _____?

-increases

-increases

45
New cards

as area ______, scatter increases

increases

46
New cards

as collimation increases, scatter _________?

decreases

47
New cards

what are artifacts?

form of image noise, any unwanted info on image

48
New cards

SID affect on amount of exposure reaching the patient and the image receptor?

inversely

49
New cards

SID affects magnification ?

inversely

50
New cards

SID affects patient dose?

inversely

51
New cards

SID affects both visibility and resolution?

directly

52
New cards

If SID increases but OID stays the same, the x-rays from the focal spot spread ______ after they pass the object on their way to the IR.

less

53
New cards

As OID increases, magnification ______?

increases (directly)

54
New cards

As OID increases, resolution

decreases (inversely)

55
New cards

As OID increases, density/receptor exposure?

decreases (inversely)

56
New cards

As OID increases, contrast

increases (directly)

57
New cards

As OID increases, size of part?

increases (directly)

58
New cards

what is the inverse square law?

the intensity of radiation is inversely proportional to the square of the distance

59
New cards

as SID increases, resulting density?

decreases (inversely)

60
New cards

as SID increases, dose ?

decreases (inversely)

61
New cards

what is the inverse square law formula?

I1/I2 = D2^2/D1^2

62
New cards

what is the square law formula?

mAs 1/mAs 2 = D1^2/D2^2

63
New cards

if SID is increased, mAs must be ? which raises dose

increased

64
New cards

What happens to ESD when SID increases without changing mAs?

ESD decreases

65
New cards

What happens to ESD when SID increasesandmAs is adjusted using the Direct Square Law?

ESD is approximately maintained, assumingotherexposure conditions remain constant.

66
New cards

As the SID increases the Source to

Skin Distance (SSD) _______ which provides the

________ patient ESD

increases

-decreased

67
New cards

as SID and collimation increases, dose ?

decreases

68
New cards

why should studies be done at 48"?

have decreased magnification and better resolution

69
New cards

how to find SOD?

SID-OID

70
New cards

what is the primary controlling factor over magnification?

SID/SOD

71
New cards

for effects on resolution and magnification, SID/SOD ratio must ?

be considered in conjuction with each other in the form of a ratio

72
New cards

magnification is _________ proportional to the SID/SOD ratio?

directly

73
New cards

as SID/SOD ratio increases, magnification?

increases (directly)

74
New cards

by either decreasing OID or increasing SID, the ratio will ? and magnification ?

decrease

-decrease

75
New cards

whats the % mag formula?

(SID/SOD- 1 ) x 100

76
New cards

what is the mag factor?

SID/SOD

77
New cards

when is SID and OID considered separately when discussing their effects on?

density and contrast

78
New cards

when can SID and SOD NOT be considered separately in controlling their effects on?

magnification and resolution

79
New cards

what is magnification?

ratio between size of image and size of object

80
New cards

how to calculate the size of part on x-ray compared to actual size?

(image size)- (object size)/ (object size) x 100

81
New cards

if OID increases, sharpness ?

decreases

82
New cards

If SOD increases, sharpness ?

increases

83
New cards

how to calculate unsharpness?

FS x OID/SOD

84
New cards

how to calculate sharpness?

SOD/ FS xOID

85
New cards

Focal spot size and OID have an _______ on sharpness/resolution?

inverse

86
New cards

SOD has a ______ on sharpness/resolution?

direct

87
New cards

As SID increases, contrast?

not affected