Scatter Radiation
X-rays that have interacted with an electron of an atom which causes the direction and energy of the x-ray photon to be altered
What does scatter do to contrast?
Decreases it
The 3 things that can happen when an x-ray photon interacts with a patient
It can be absorbed
It can pass through unaffected
It can scatter
2 major controllers of scatter radiation
Beam restriction, Grid
2 factors that affect the amount of scatter radiation produced
kVp, irradiated material
Higher kVp =
Greater the scatter, highly penetrating, less photons being absorbed by the patient, less contrast on image
The amount of scatter created during an interaction is affected by
The volume and the atomic number of the material being irradiated
The greater the volume =
More scatter
Volume increases as…
Field size increases, patient size increases
The higher the atomic mass the greater..
The chance of x-ray interaction due to photoelectric absorption
Less scatter occurs
Bone absorbs more radiation than…
Soft tissue
The larger the field size, the more x-ray photons allowed to interact with the patient which…
Creates more scatter
What should you do with collimation to decrease scatter?
Use the smallest field size possible
The 4 types of beam restricting devices
Aperture Diaphragm
Cones and Cylinders
Collimator
Lead Blocker/Mask
Aperture Diaphragm defintion and characteristics
A flat sheet of metal, usually lead, with a hole cut in the center and attached to the x-ray tube
Simplest to use
Cheapest to buy
Can be made into any size or shape
Disadvantages of Aperture Diaphragm
Different sizes and shapes needed to accommodate different film sizes
Increased penumbra
Penumbra
Geometric unsharpness around the periphery of the image
What causes penumbra?
Due to x-rays not having a “point source”
Penumbra is reduced when…
Beam restrictor is used, using small focal spot
Decrease penumbra =
Increase detail
What radiation does aperture diaphragms cause an increase in?
Off-focus radiation
Off-Focus Radiation
Photons not created at the focal spot but elsewhere in the tube. Produces ghost-like shadow on the radiograph
Determining the diaphragm field size aids in determining….
The projected image on the image receptor
Formula for Diaphragm Field Size
Projected image size = SID × diameter of diaphragm / distance from focal spot to diaphragm
What do cones affect?
Does not reduce penumbra
Does not increase recorded detail
Improves contrast
Only cleans up scatter
What does a cylinder affect?
Reduces penumbra
Reduces off-focus radiation
Cylinder Disadvantages
Difficult to use due to weight
Fixed field size
Cone Field Size Equation
Cone Field Size = SID × lower diameter of cone / distance from focal spot to bottom of cone
What is the most common type of beam restrictor used in radiography?
Collimator
Collimator advantages & disadvantages
More complex and expensive
Allows numerous field sizes
Provides a light source to aid in positioning
A collimator should never be opened greater than…
The size of the image receptor
Upper shutters on collimators reduce…
Off-Focus Radiation
Bottom shutters on collimators reduce…
Penumbra
Positive Beam Limitation
Sensing device that determines the size and placement of cassette in bucky tray. Automatically collimates to size of cassette by moving shutters.
US required by law from 1974 to 1993 that all fixed x-ray equipment have…
Positive Beam Limitation Device
Lead Blockers
Sheet of lead-impregnated rubber that can be cut into any size or shape. Absorbs scatter from large body parts and aids in positioning
Where are lead blockers most commonly used?
Lateral L-Spine, Lateral T-Spine
Lead Blocker affects…
Cleans up scatter
Does not decrease patient dose
Lead Mask
Cut to field size being used, secured at end of collimator, most common in cerebral angiography
Increasing field size also means…
Opening the collimator
Decreasing beam restriction
Increasing field size affects…
Increased scatter
Increased IR exposure
Decreased contrast
Decreasing field size also means….
Closing collimation
Increasing beam restriction
Decreased field size affects….
Decreases beam quantity
Decreases scatter
Increases contrast
Decreases IR exposure
Increase in mAs must accompany a significant reduction in field size to…
Maintain IR exposure
Distortion
The misrepresentation of the size or shape of the structures being examined
The 2 types of distortion
Size distortion
Shape distortion
Size distortion type
Magnification
What does magnification reduce?
Reduces recorded detail
2 factors that affect size distortion
OID
SID
The greater the SID, the _____ the magnification
Less
The greater the OID, the ____ the magnification
More
Magnification Factor Equation
SID/SOD
Magnification Factor - the 3 ways to write it
40”/30”
1.33 x's larger than it’s size
The magnification will be 33%
The image will be 133% of the object size
Object Size Equation
Image Size / Magnification Factor
Percent Magnification Equation
IS - OS / OS × 100
Shape Distortion Definition
An unequal magnification of the object
2 types of shape distortion
Elongation, Foreshortenining
What causes shape distortion?
Tube/part/receptor alignment
Shape Distortion Angulation
Direction: cephalic and caudad will distort image
Degree: greater angle on tube causes more distortion or greater elongation
Shape Distortion Alignment
Central Ray: beam divergence is greater at edges due to angle of beam. Center of beam more perpendicular (no angle), MATCH IR
Anatomical part: part aligned correctly to IR and CR
Image Receptor: flat, not angled, perpendicular to beam
What causes elongation?
Caused by misalignment (angulation) of the central ray and the image receptor
Greater angle = greater elongation
Foreshortening
Object appears to be shorter than it really is
What causes foreshortenining?
Part is improperly aligned to IR
Tube or IR angle will only cause what?
Elongation NEVER FORESHORTENING
3 ways to minimize distortion
Use a small OID
Use a large SID
Ensure proper tube/part/receptor alignment
5 (Tube angulation)
39.8 (Overhead scale)
10 (Tube angulation)
39.4 (Overhead scale)
15 (Tube angulation)
38.6 (Overhead Scale)
20 (Tube angulation)
37.6 (Overhead Scale)
Filtration
The process of eliminating undesirable low energy x-ray photons by the insertion of absorbing materials into the primary beam.
Filtration is sometimes called “…… .. … ….”
Hardening of the Beam
Because it removes the low-energy or soft photons
Soft-x-rays or low energy x-rays are mostly _________ by the first few cms of tissue
Absorbed
What do soft-x-rays contribute to the x-ray image?
Nothing
Soft-x-rays ________ patient dose without adding anything to the image.
Increase
What is a filter and where is it placed?
A filter is anything that selectively absorbs photons from the x-ray beam.
It is placed between the tube and the patient.
What is the most common filter material used?
Aluminum
What are some other filter materials used?
Glass, oil, copper, tin, plastic
What are filters expressed in?
Terms of thickness - Al/Eq (Aluminum Equivalency)
Measured in mm
The Half-Value Layer (HVL)
The amount of absorbing material that will reduce the intensity of the primary beam to ½ its original value
What are the units of measurement for the Half-Value Layer?
mm Al/Eq
1 HVL =
2 mm Al/Eq
An x-ray tube that operates above 130 kVp must have a minimum of…
3.5 mm Al/Eq
An x-ray tube above 150 kVp must have..
4.1 mm Al/Eq
Mammography machines use below 50 kVp and use…
.3 mm Al/Eq
What are the 2 types of filtration?
Inherent, Added
Inherent Filtration
Built into the x-ray tube and housing.
Examples: glass envelope of the tube, glass window of tube, insulating oil surrounding the tube (dielectric oil)
Average inherent filtration?
.5 to 1 mm Al/Eq
Where is most inherent filtration found?
In the window of the glass envelope
Added Filtration
Any filtration that occurs outside the x-ray tube and above the image receptor.
Examples: mirror and collimator (1 mm Al/Eq)
Low atomic numbers function well as a…
Low energy absorber
Total Filtration Equation
Total Filtration = Inherent + Added Filtration
Below 50 kVp
.5 mm Al/Eq
50 to 70 kVp
1.5 mm Al/Eq
Above 70 kVp
2.5 mm Al/Eq
Compound Filters
Use 2 or more materials to complement one another in their absorbing abilities
The second filter absorbs scatter before it…
Hits the patient
Compound filters are also called…
K-edged filters
What materials do the 2 layers of a compound filter contain?
One layer of Aluminum and one layer of copper
The filter with the highest atomic number is placed closest to the ____ and the lowest number would be placed closest to the ________
Tube, Patient
Copper is closest to tube and aluminum is closest to patient
Aluminum Atomic Number
13