Scatter Radiation

X-rays that have interacted with an electron of an atom which causes the direction and energy of the x-ray photon to be altered

2
New cards

What does scatter do to contrast?

Decreases it

3
New cards

The 3 things that can happen when an x-ray photon interacts with a patient

  • It can be absorbed

  • It can pass through unaffected

  • It can scatter


4
New cards

2 major controllers of scatter radiation

Beam restriction, Grid

5
New cards

2 factors that affect the amount of scatter radiation produced

kVp, irradiated material

6
New cards

Higher kVp =

Greater the scatter, highly penetrating, less photons being absorbed by the patient, less contrast on image

7
New cards

The amount of scatter created during an interaction is affected by

The volume and the atomic number of the material being irradiated

8
New cards

The greater the volume =

More scatter

9
New cards

Volume increases as…

Field size increases, patient size increases

10
New cards

The higher the atomic mass the greater..

The chance of x-ray interaction due to photoelectric absorption

Less scatter occurs

11
New cards

Bone absorbs more radiation than…

Soft tissue

12
New cards

The larger the field size, the more x-ray photons allowed to interact with the patient which…

Creates more scatter

13
New cards

What should you do with collimation to decrease scatter?

Use the smallest field size possible

14
New cards

The 4 types of beam restricting devices

Aperture Diaphragm

Cones and Cylinders

Collimator

Lead Blocker/Mask

15
New cards

Aperture Diaphragm defintion and characteristics

A flat sheet of metal, usually lead, with a hole cut in the center and attached to the x-ray tube

  • Simplest to use

  • Cheapest to buy

  • Can be made into any size or shape


16
New cards

Disadvantages of Aperture Diaphragm

  • Different sizes and shapes needed to accommodate different film sizes

  • Increased penumbra


17
New cards

Penumbra

Geometric unsharpness around the periphery of the image

18
New cards

What causes penumbra?

Due to x-rays not having a “point source”

19
New cards

Penumbra is reduced when…

Beam restrictor is used, using small focal spot

20
New cards

Decrease penumbra =

Increase detail

21
New cards

What radiation does aperture diaphragms cause an increase in?

Off-focus radiation

22
New cards

Off-Focus Radiation

Photons not created at the focal spot but elsewhere in the tube. Produces ghost-like shadow on the radiograph

23
New cards

Determining the diaphragm field size aids in determining….

The projected image on the image receptor

24
New cards

Formula for Diaphragm Field Size

Projected image size = SID × diameter of diaphragm / distance from focal spot to diaphragm

25
New cards

What do cones affect?

  • Does not reduce penumbra

  • Does not increase recorded detail

  • Improves contrast

  • Only cleans up scatter


26
New cards

What does a cylinder affect?

  • Reduces penumbra

  • Reduces off-focus radiation


27
New cards

Cylinder Disadvantages

  • Difficult to use due to weight

  • Fixed field size


28
New cards

Cone Field Size Equation

Cone Field Size = SID × lower diameter of cone / distance from focal spot to bottom of cone

29
New cards

What is the most common type of beam restrictor used in radiography?

Collimator

30
New cards

Collimator advantages & disadvantages

  • More complex and expensive

  • Allows numerous field sizes

  • Provides a light source to aid in positioning


31
New cards

A collimator should never be opened greater than…

The size of the image receptor

32
New cards

Upper shutters on collimators reduce…

Off-Focus Radiation

33
New cards

Bottom shutters on collimators reduce…

Penumbra

34
New cards

Positive Beam Limitation

Sensing device that determines the size and placement of cassette in bucky tray. Automatically collimates to size of cassette by moving shutters.

35
New cards

US required by law from 1974 to 1993 that all fixed x-ray equipment have…

Positive Beam Limitation Device

36
New cards

Lead Blockers

Sheet of lead-impregnated rubber that can be cut into any size or shape. Absorbs scatter from large body parts and aids in positioning

37
New cards

Where are lead blockers most commonly used?

Lateral L-Spine, Lateral T-Spine

38
New cards

Lead Blocker affects…

  • Cleans up scatter

  • Does not decrease patient dose


39
New cards

Lead Mask

Cut to field size being used, secured at end of collimator, most common in cerebral angiography

40
New cards

Increasing field size also means…

Opening the collimator

Decreasing beam restriction

41
New cards

Increasing field size affects…

  • Increased scatter

  • Increased IR exposure

  • Decreased contrast


42
New cards

Decreasing field size also means….

  • Closing collimation

  • Increasing beam restriction


43
New cards

Decreased field size affects….

  • Decreases beam quantity

  • Decreases scatter

  • Increases contrast

  • Decreases IR exposure


44
New cards

Increase in mAs must accompany a significant reduction in field size to…

Maintain IR exposure

45
New cards

Distortion

The misrepresentation of the size or shape of the structures being examined

46
New cards

The 2 types of distortion

  • Size distortion

  • Shape distortion


47
New cards

Size distortion type

Magnification

48
New cards

What does magnification reduce?

Reduces recorded detail

49
New cards

2 factors that affect size distortion

  • OID

  • SID


50
New cards

The greater the SID, the _____ the magnification

Less

51
New cards

The greater the OID, the ____ the magnification

More

52
New cards

Magnification Factor Equation

SID/SOD

53
New cards

Magnification Factor - the 3 ways to write it

40”/30”

1.33 x's larger than it’s size

The magnification will be 33%

The image will be 133% of the object size


54
New cards

Object Size Equation

Image Size / Magnification Factor

55
New cards

Percent Magnification Equation

IS - OS / OS × 100

56
New cards

Shape Distortion Definition

An unequal magnification of the object

57
New cards

2 types of shape distortion

Elongation, Foreshortenining

58
New cards

What causes shape distortion?

Tube/part/receptor alignment

59
New cards

Shape Distortion Angulation

  • Direction: cephalic and caudad will distort image

  • Degree: greater angle on tube causes more distortion or greater elongation


60
New cards

Shape Distortion Alignment

Central Ray: beam divergence is greater at edges due to angle of beam. Center of beam more perpendicular (no angle), MATCH IR

Anatomical part: part aligned correctly to IR and CR

Image Receptor: flat, not angled, perpendicular to beam

61
New cards

What causes elongation?

Caused by misalignment (angulation) of the central ray and the image receptor

Greater angle = greater elongation

62
New cards

Foreshortening

Object appears to be shorter than it really is

63
New cards

What causes foreshortenining?

Part is improperly aligned to IR

64
New cards

Tube or IR angle will only cause what?

Elongation NEVER FORESHORTENING

65
New cards

3 ways to minimize distortion

  • Use a small OID

  • Use a large SID

  • Ensure proper tube/part/receptor alignment


66
New cards

5 (Tube angulation)

39.8 (Overhead scale)

67
New cards

10 (Tube angulation)

39.4 (Overhead scale)

68
New cards

15 (Tube angulation)

38.6 (Overhead Scale)

69
New cards

20 (Tube angulation)

37.6 (Overhead Scale)

70
New cards

Filtration

The process of eliminating undesirable low energy x-ray photons by the insertion of absorbing materials into the primary beam.

71
New cards

Filtration is sometimes called “…… .. … ….”

Hardening of the Beam

Because it removes the low-energy or soft photons

72
New cards

Soft-x-rays or low energy x-rays are mostly _________ by the first few cms of tissue

Absorbed

73
New cards

What do soft-x-rays contribute to the x-ray image?

Nothing

74
New cards

Soft-x-rays ________ patient dose without adding anything to the image.

Increase

75
New cards

What is a filter and where is it placed?

A filter is anything that selectively absorbs photons from the x-ray beam.

It is placed between the tube and the patient.

76
New cards

What is the most common filter material used?

Aluminum

77
New cards

What are some other filter materials used?

Glass, oil, copper, tin, plastic

78
New cards

What are filters expressed in?

Terms of thickness - Al/Eq (Aluminum Equivalency)

Measured in mm

79
New cards

The Half-Value Layer (HVL)

The amount of absorbing material that will reduce the intensity of the primary beam to ½ its original value

80
New cards

What are the units of measurement for the Half-Value Layer?

mm Al/Eq

81
New cards

1 HVL =

2 mm Al/Eq

82
New cards

An x-ray tube that operates above 130 kVp must have a minimum of…

3.5 mm Al/Eq

83
New cards

An x-ray tube above 150 kVp must have..

4.1 mm Al/Eq

84
New cards

Mammography machines use below 50 kVp and use…

.3 mm Al/Eq

85
New cards

What are the 2 types of filtration?

Inherent, Added

86
New cards

Inherent Filtration

Built into the x-ray tube and housing.

Examples: glass envelope of the tube, glass window of tube, insulating oil surrounding the tube (dielectric oil)

87
New cards

Average inherent filtration?

.5 to 1 mm Al/Eq

88
New cards

Where is most inherent filtration found?

In the window of the glass envelope

89
New cards

Added Filtration

Any filtration that occurs outside the x-ray tube and above the image receptor.

Examples: mirror and collimator (1 mm Al/Eq)

90
New cards

Low atomic numbers function well as a…

Low energy absorber

91
New cards

Total Filtration Equation

Total Filtration = Inherent + Added Filtration

92
New cards

Below 50 kVp

.5 mm Al/Eq

93
New cards

50 to 70 kVp

1.5 mm Al/Eq

94
New cards

Above 70 kVp

2.5 mm Al/Eq

95
New cards

Compound Filters

Use 2 or more materials to complement one another in their absorbing abilities

96
New cards

The second filter absorbs scatter before it…

Hits the patient

97
New cards

Compound filters are also called…

K-edged filters

98
New cards

What materials do the 2 layers of a compound filter contain?

One layer of Aluminum and one layer of copper

99
New cards

The filter with the highest atomic number is placed closest to the ____ and the lowest number would be placed closest to the ________

Tube, Patient

Copper is closest to tube and aluminum is closest to patient

100
New cards

Aluminum Atomic Number

13