What factors can the operator control?
Milliamperage (mA)
Scan time
Kilovoltage peak (kVp)
Slice thickness
Field of view
Reconstruction algorithm
Pitch (helical scanning)
What factors affect the total beam exposure in CT
mA setting
Scan time
kVp
For the factors that affect the total beam exposure in CT, which factor is held constant
kVp’s are generally held constant
For the factors that affect the total beam exposure in CT, which factor is varied
mAs is varied to control patient dose and image quality
What is mA
The current of electrons flowing from the cathode to the anode
How does mA work?
Electrons are boiled off when the filament is heated
X-ray photons are produced when electrons strike the anode
Increasing mA will increase the number of photons
What is scan time and how does it work
The time the x-ray beam is on for the collection of data for each slice
Most often, it is the time required for the gantry to make a 360-degree rotation
Shorter scans help to reduce motion on the image
What is the relationship between scan time and mAs
mAs = mA x scan time
Using a higher mA setting allows for a shorter scan time while maintaining the same mAs
Shorter scan time helps reduce patient motion
mAs is limited by heat
Recommended mAs settings are programmed into each exam protocol
What happens to SNR when mAs is doubled
SNR increases by approximately 40%
How do 5 mm and 2.5 mm slices compare
A 5 mm slice contains twice as many photons as a 2.5 mm slice, improving contrast resolution
What is the formula for pixel size
Pixel size = DFOV / matrix size
What is kVp and how does it work
AKA tube voltage
Controls the quality and energy of the beam
What standard kVp values are used in CT
Adults: 120-140 kVp
Pediatrics: Usually 80 kVp
What is the uncoupling effect and how does it affect dose
When using digital technology, the image quality is not directly linked to the dose, so even when mA or kVp setting that is too high is used, a good image still results
This effect can make it difficult to identify when a dose that is higher than necessary is used
Dose Creep: Gradual increase in techniques over time because high dose images still look acceptable
What is Automatic Tube Current Modulation (ATCM) and how does it work
An equipment (software) option that will make changes in tube current (mA) based on the estimated attenuation of the patient at a specific location
These estimations are derived from scout views done in both the AP and lateral projections or from the previous slices. From these views, the mA will be programmed to vary by location along the length of the patient
The exact details of the option vary by manufacturer
What is slice thickness and what is its impact on image quality?
The slice thickness is how the data was acquired
Slice thickness has a significant impact on image quality
What is the difference between the slice thickness and image thickness
Slice thickness is how the data was acquired
Image thickness is how the data is reconstructed (processed)
Looks at raw data versus image data
What is Scan Field of View
Determines the area, within the gantry for which all raw data are acquired
Acquired around the gantrys isocenter
What is the Display Field of View
Determines how much and what section, of the collected raw data are used to create an image
What are Reconstruction Algorithms
Depending on the manufacturer, is also called algorithm, convolution filter, FC filter or filter
Applied only to raw data
Changing the reconstruction algorithm will change the way the raw data are manipulated to reconstruct the image
Windowing does not change the algorithm; it will only change the way the image is viewed
What is pitch
The relation of table speed to slice thickness
The travel distance of the CT scan table per 360 rotation of the x-ray tube, divided by the x-ray beam collimation width
Know about tube arc and what is the most common tube arc
360 degrees is the most common tube arc
This is data collected from 1 360-degree rotation of the x-ray tube (2 samples 180 degrees apart)
400 degrees is used for overscanning
This was used in 4th generation scanners to reduce motion artifact by collecting overlapping data
Define Image Quality
A comparison of the image to the actual object
In CT, image quality is directly related to its usefulness in providing an accurate diagnosis
What 2 features are used to assess how well the image represents the real anatomy
Spatial Resolution (detail) - the ability to resolve (as separate objects) small, high-contrast objects
Contrast Resolution - the ability to differentiate between objects with very similar densities as their background (low contrast detectability)
What is spatial resolution?
The ability to resolve (as separate objects) small, high-contrast objects
Can also be called detail
What is contrast resolution
The ability to differentiate between objects with very similar densities as their background (low contrast detectability)
What are the 2 ways to measure spatial resolution
Direct:
by using a line pairs phantom
Counts the number of strips
Spatial Frequency- the number of line pairs visible per unit length
Small objects=high spatial frequency
Large objects=low spatial frequency
Modulation Transfer Function:
The ratio of the accuracy of the image compared with the actual object being scanned
Most commonly used method in CT and DR
Indicated image fidelity
Scale from 0-1
0= blank image, no info
1= object reproduced exactly
As the object size increases, MTF increases
What factors affect spatial resolution
Pixel size (matrix size, DFOV)
Slice thickness
Reconstruction algorithm
Focal spot size
Pitch (because of slice thickness)
Patient motion
How does a smaller pixel size affect spatial resolution
Increased spatial resolution
How does a bigger pixel affect spatial resolution
Decreased Spatial Resolution
How does using a thinner slice affect spatial resolution
Increased Spatial Resolution
How does using a wider slice thickness affect spatial resolution
Decreased Spatial Resolution
How does using the appropriate detail enhancing algorithm affect spatial resolution
Increases Spatial Resolution
How does the use of a larger focal spot affect spatial resolution
Decreases Spatial Resolution
How does the use of a smaller focal spot affect spatial resolution
Increases Spatial Resolution
How does increased pitch affect spatial resolution
Decreases Spatial Resolution
How does decreased pitch affect spatial resolution
Increases Spatial Resolution
How does patient motion affect spatial resolution
Decreases Spatial Resolution
What is in-plane resolution
Resolution in the x and y direction
What is longitudinal resolution
Resolution in the Z direction
What is an isotropic voxel and how does it affect information when data are reformatted in a different plane
A cube with equal dimensions in the X, Y, and Z directions
When the imaging voxel is equal in size in all dimensions, there is no loss of information when data are reformatted in a different plane
What is the Nyquist Sampling Theorem and how is it related to pixel dimension
Theorem that states because an object may not lie entirely within a pixel, the pixel dimension should be half the size of the object to increase the likelihood of that object being resolved
What is Iterative Reconstruction
A different way to apply an algorithm
Begins with an assumption and then improve the image by continually analyzing scan data and making adjustments until the assumption and real time data are in agreement
Creates less noise and better spatial resolution then filtered back projection
What are deep learning reconstruction methods
Deep learning reconstruction produces significantly lower noise while preserving low-contrast detectability and spatial resolution
Know about low-contrast detectability
The ability to distinguish an object that is nearly the same density as its background
High contrast differences are easier to see
Low contrast differences are more difficult to see
What can contrast resolution also be called
Sensitivity of the system
What is % linear attenuation coefficient and the % difference in CT related to contrast change
The contrast difference between objects
1% contrast difference = 10 HU
CT can differentiate an object with a 0.5% contrast variation
CT has better contrast resolution than conventional x-ray, but MRI is considered superior for contrast resolution
Know about the phantom that is used to measure small differences in densities
Cylindrical phantom
Contains objects with small differences in density
Used to evaluate the system’s ability to distinguish low-contrast objects
What is noise in CT
The undesirable fluctuation of pixel values in an image of a homogeneous material
Appears grainy or like salt and pepper
What is the most common cause of noise
Quantum Mottle (insufficient number of photons)
Fewer photons=more noise
What is SNR
Signal to noise ratio
Higher SNR is better (5:1 is better than 4:1)
What does standard deviation indicate
The amount of variance among pixel values in a designated ROI
The SD of a phantom indicates the degree of noise
Lower SD=less noise and better contrast resolution
What 5 factors affect contrast resolution
mAs/Dose
Pixel Size
Slice Thickness
Reconstruction Algorithm
Patient Size
How does increasing mAs affect contrast resolution
Increases SNR and improves contrast resolution, but increases patient dose
How does pixel size affect contrast resolution
Smaller pixels= more noise and worse contrast resolution
Larger pixels= less noise and better contrast resolution
How does slice thickness affect contrast resolution
Thicker slices = more photons and better contrast resolution, but worse Z-axis spatial resolution
How do reconstruction algorithms affect contrast resolution
Bone (Edge, high pass) = better spatial resolution, worse contrast
Soft tissue (Smooth, low pass) = better contrast and worse spatial resolution
How does patient size affect contrast resolution
Larger patient = lower SNR, more noise, and worse contrast resolution (same technique)
What exam can use a low dose despite reduced contrast resolution
Low-dose CT lung cancer screening because nodules and masses can still be seen
What is subject contrast
The inherent properties of the object scanned (Tech cannot control this)
How does windowing help contrast resolution
For low contrast objects to be visible, the window level must be set near the average density of the object, and the window width must be narrow enough so that only small groups of HU are represented in each shade of gray displayed
What is temporal resolution?
How rapidly data are acquired
What 3 factors control temporal resolution
Gantry rotation speed
The number of detector channels in the system
The speed with which the system can record changing signals
What is temporal resolution measured in
Milliseconds (ms)
When is temporal resolution important
Heart
Angiography
Perfusion studies (contrast studies)
How do manufacturers report spatial resolution
MTF
How do manufacturers report contrast resolution
% linear attenuation coefficient
Sensitivity of the system
SNR (signal to noise ratio)
Standard deviation
What is the main difference between spatial and contrast resolution
Spatial: Detail (small, high contrast objects)
Contrast: Density differences (similar-density objects)