issue: All the analyte spots are in the upper region of the TLC plate & close together
the eluent used was too polar
issue: Analyte spots moved up the plate diagonally instead of vertically
the TLC plate was touching the filter paper wick
While spotting the analyte spot became too large
TLC plate didn’t dry between drops
Rf values differ for 2 identically spotted plates from the same developing chamber containing 2:1 acetone:heptane eluent
eluent chamber was not capped, resulting in evaporation
why shouldn’t you use pen to mark the TLC plate
pen ink will bleed up the plate
approximately how far should you let the eluent travel?
75-85% up the plate
you should NEVER touch the plate…
in between the solvent and origin line after its been run
stationary phase
the fixed material (solid or liquid) that stays in place and interacts with the mixture (silica in this project)
mobile phase
moving fluid (liquid or gas) that carries a sample mixture through a system (heptane/acetone ratio that is also the eluent)