4 factors that affect image acquisition
mA
time
kV
distance
kV
kilovoltage
kV affects?
primary = quality
secondary = quantity (bc INC kV = INC speed of e- / penetrating power = more XR photons hit IR, doesn't actually generate more xray photons)
quality synonyms
beam energy, penetrability
quantity synonyms
beam intensity, exposure, amount of xr
how does kV work?
INC kV = INC speed of electrons as they hit anode with more force
hitting anode with more force = higher energy xr photons produced
higher energy beam/quality will cause more photons to penetrate thru patient and hit IR, therefore increasing quantity
INC kV = INC speed of electrons = INC energy = INC quality AND quantity
kV controls?
radiographic contrast
% change to perceive a change in optical density?
need at least a 4% change in kV to perceive a change in optical density
(multiply by 0.04 to find out the amount of kV needed to make a perceivable change)
INC kV = --- image contrast
INC kV = lower contrast (long scale, more grays)
DEC kV = --- image contrast
DEC kV = higher contrast (short scale, less grays)
? allows us to have good images with bad technical factors
digital post processing
prefer over or underexposed images?
overexposed because post processing cannot add more xr photons to an image
long scale
low contrast
more grays
short scale
high contrast
less grays, more black and white
low contrast desired for ?
chest and abdomen exams bc the organs are different densities
high contrast desired for ?
extremities because mostly just skin and bone, not much soft tissue
--- kV, --- mAs better for tube life and pt dose?
HIGH kV
LOW mAs
better for tube life because it uses less heat
reduces patient exposure
15% rule
to double the exposure of an image, increase kV by 15% (same as x2 mAs) MULTIPLY BY 1.15
to half the exposure of an image, decrease kV by 15% (same as 1/2 the mAs) MULTIPLY BY 0.85
when to use 15% rule vs 4% change kV?
15% rule when need drastic exposure change (double or half exposure)
4% rule when image is almost optimal, but need a minor tweak
maintaining exposure
if kV is INC 15%, DEC mAs by half to maintain same exposure
used only when image is diagnostic quality but exposure number is incorrect or kV values increased for reduction in patient exposure
high kV techniques
reduce pt exposure
reduce the xr tube heat production
heat unit equation for more efficient generators (3 phase, 12 pulse)
HU = mAs x time x kV x 1.41
inverse square law
the intensity of radiation is inversely proportional to the square of the distance
inverse square law eq
i1/i2 = d2^2/d1^2
xr in air
roentgen
R or mR
INC distance = --- intensity
INC distance = DEC intensity
DEC distance = --- intensity
DEC distance = INC intensity
direct square law
mAs is directly proportional to the square of the distance
if you increase distance you will increase mAs to maintain same amount of photons hitting the IR
direct square law equation
mAs1/mAs2 = d1^2/d2^2
how to solve for greatest amount of exposure
rank each technical factor from smallest to greatest
add the rank numbers
greatest number = greatest amt of exposure
smallest number = smallest amt of exposure
low mA + high kV = --- HU
low mA + high kV = low HU
high mA + low kV = --- HU
high mA + low kV = high HU