advantages of beam limitation (3)
reduces patient exposure
increases image quality by decreasing scatter to the IR
increases visibility of detail
% of density on the radiograph resulting from scatter radiation
50-90%
where is penumbra greater at on the beam?
at the edges
types of beam limiting devices (3)
aperture diaphragm, cones, variable aperture
what are beam limiting devices made of?
lead
what is advantage of lead?
has the ability to absorb x-rays
aperture diaphragms
simplest, least expensive
change for each size
large amount of blur
dedicated units
types of cones (2)
flared and cylinder
flared cone
matches the divergence of the beam; flare is usually larger than the beam
TRUE or FALSE: flared portion of a flare cone does not contribute to beam restriction
TRUE
cylinder cone
circular field; blur is reduced because it limits the field size
TRUE or FALSE: decreased beam restriction is obtained with the use of a cone with extension
FALSE
disadvantage of cones
excessive radiation for larger body parts
variable apertures (3)
collimator
automatic collimator
positive beam limiting device (PBL)
positive beam limiting device (PBL)
assures the size of field does not exceed the size of the IR
automatic collimator housing
2 sets of horizontal lead shutters
adjustable; allowed for infinite # of field sizes
contributes to “added” filtration
provides a light field
what do upper shutters in an automatic collimator control?
stem radiation
what do lower shutters in an automatic collimator reduce?
blur
stem radiation
refers to photons not produced at the focal spot of the anode
stem radiation housing
tube housing; evaporate tungsten on envelope
field light congruency
pennies must align with border
what will poor field to light congruency lead to?
parts of the anatomy being cut off