purpose of microscopy
magnification; resolution, contrast
magnification
makes fine details more visible = increase resolution
resolution
minimum difference between 2 points that can be observed as separate distinct points
empty magnification
magnification with no increased resolution
introducing contrast to samples
staining, special optics to enhance small amount of existing contrast in unstained specimens
light microscope 2 systems
light and glass lens system
properties that change how light travels
RI of a medium, wavelength
diffraction
light scatters as it passes by and contacts objects
shorter wavelengths diffract less
diffraction leads to
coloured halos = decreased resolution
refraction
bend light as it passes through one medium to another
refractive index
measure of how much medium affects speed of light
amount of refraction through curved glass lens is affected by
lens shape, thickness, wavelength of light
focal point
point at which refracted rays converge on other side of lens
focal length
distance from midpoint of lens to focal point
convex lens
positive lens
light rays converge to focal point
concave lens
negative lens
upon exit, rays diverge/no focal point
optical aberrations
deviation from what is expected
chromatic aberration
due to different wavelengths refracting to different degrees as they pass through lens → image won’t be in complete focus
spherical aberration
light rays refract different depending on where in the lens they hit → image not focused
field curvature
image appears curved → image field can’t be focused simultaneously
correcting for aberrations
combine positive and negative lenses → cancels out aberrations
use different lens material
virtual image
image seen relative to specimen on slide
will be: enlarged, inverted, reversed
ocular
further magnifies real image of specimen
objective
creates magnified real image of specimen, primary determinant of image resolution, does most of image correction
types of microscopes
brightfield, phase contrast, inverted, polarized, fluorescent
plan corrects for
field curvature
highest level of objective lens correction
Plan apochromat
increasing correction
Numerical Aperture and resolution increases
field lens
directs column of light to substage condenser
field diaphragm
atop field lens; controls amount of light
sub stage condenser
gathers light from field lens and focuses it onto specimen; contains aperture diaphragm
numerical aperture (NA)
measure of light gathering capacity of objective lens
10x objective AD setting
0.2
20X objective ad setting
0.4
40X objective ad setting
0.5
100X objective ad setting
1
if ad diameter is too wide
light goes past objective → information is lost
kohler illumination
technique to align illumination and optic systems
sets correct amount of light entering objective = maximized resolution; provides sufficient and even illumination of FOV
minimizes artifact and misinformation
3 main components involved in kohler
AD, sub stage condenser, field diaphragm
when magification increases
NA, AD, and resolution increases
FOV, depth of field, working distance decreases
field of view
sample area diameter → how much of sample is seen
depth of field
number of specimen layers simultaneously in focus
shallower DOF = less layers focused at one time
working distance
distance between specimen and objective lens
cover glass
protects slide mounted specimen; needs to match objective
microscope slide
thin flat piece of glass used to support specimen
medium
what is between the cover glass and objective
immersion oil
RI similar to glass → minimizes light loss due to internal refraction
for higher magnification objectives to maximize resolving power