Absorbed dose
A measure of the biologically significant effect produced by ionizing radiation
Applied/Given dose
The absorbed dose at Dmax
_____ is used to signify the dose delivered beyond the PTV to OARS
Exit dose
Absorbed dose rate
Output specified in terms of absorbed dose rate in a patient or phantom at a given SSD for a specified field size and reference depth
Target dose
Dose prescribed at a point of interest (specific D)
The abosorbed dose at a point within the volume to be treated
Integral dose
Summation of the dose to the entire irradiated area
TOTAl energy deposited within the whole body
Sources of secondary radiation are?
Patient scatter
Collimator scatter
Leakage radiation
All of the above
All of the above
Dose deposition in medium is affected by
Attenuation of the beam
ISL
Scatter within medium
All of the above
All of the above
Cobalt units are set in terms of time output tables are given in terms of cGy/min at Dmax
Linacs are set in terms of Monitor Units and output tables are given in cGy/MU at Dmax
Give formula for treatment time
GD / Dose rate = tx time or MUs
Steps to a single SSD field treatment time calculation
1. Check for dose fractionation, field size, SSD, field depth or pt separation, number of fields
2. Determine daily target dose
3. Check PPD data
4. Calculate daily given dose
Daily target dose x 100% / PDD
5. Determine maximum absorbed dose rate
6. Calculate treatment time
Daily given dose / MADR
Write out equation
Daily treatment time = Total target dose / # fractions x PDD x output
The output of a machine is also called the _____?
Dose rate
Given dose is dose at ______.
Dmax
You can calculate TT with _____ and _____.
PDD and ADR
Negative vs positive stand-off
Negative: the cone is directly on the skin
Positive: Cone is not directly on the skin surface, is at a distance
Given dose, applied dose, and max absorbed dose are all used to describe the dose received at dmax? T/F
True
When we increase the SSD, what happens to the PDD?
PDD with be higher ALWAYS
What is the mayneord factor used for? Give the equation
To convert PDDs from one SSD to another
New PDD = (SSD1 + d)^2 (SSD2 +dm)^2
PDD ------------------ x ------------
SSD1 +dm)^2 (SSD2 +d)^2
The dose variation will get ______ as the patient thickness increases or the energy of the beam decreases
Worse
____ is a measure of the machine's radiation output and is used to control the dose delivered during radiation therapy. One _____ represents a specific amount of radiation which is standardized to deliver a set dose under defined reference conditions. _____ counts the # delivered and the treatment is programmed to stop after a preset number of MUs have been delivered.
A monitor unit
MU
Internal ionization chamber