26 Terms

1
New cards

acceptance testing

performed upon equipment installation, by physicists and vendor representative

2
New cards

commissioning

performed by qualified physicist, establishes baselines

3
New cards

QA testing

performed by physicist, equipment calibrated by ADL (accredited dosimetry callibration laboratory)

4
New cards

spot checks

  • performed by therapist, equipment less precise

  • array detectors

  • multiple tests in 1 exposure (output, symmetry, constancy)


5
New cards

daily QA dosimetry

  • x-ray output constancy: 3%

  • electron output constancy: 3% (all energies at least 2x weekly)


6
New cards

daily QA mechanical

  • lasers: 2mm (non IMRT), 1.5 (IMRT), 1mm (SBRT)

  • ODI: 2mm

  • collimator size: 2mm (non IMRT/IMRT), 1mm (SBRT)


7
New cards

daily QA safety

door interlock & audiovisual monitor: functional

8
New cards

daily QA checks

  • water level/temp/pressure

  • gas pressure SF6!!

  • general conditions

  • tx devices

  • supplies


9
New cards

monthly QA check - dosimetry

  • x-ray output constancy: 2%

  • x-ray beam profile constancy: 1%

  • electron output constancy: 2%

  • electron beam profile constancy: 1%

  • backup monitor: 2%

  • dose rate constancy: 2%


10
New cards

monthly QA check safety

emergency off & wedge interlocks: functional

11
New cards

monthly QA check mechanical

  • lasers 2mm (non IMRT), 1mm (IMRT, SBRT, SRS)

  • light vs x-ray: 2mm/1mm (asym)

  • FS indicators: 2mm/1mm (asym)

  • tx couch position (2mm/1 degree

  • gantry/collimator: 1 degree

  • cross hair center: 1 mm

  • light field intensity & tray/wedge latches: functional


12
New cards

temperature and pressure corrections

  • most ion chambers unsealed

  • chamber sensitivity changes with temp and pressure

  • increased temp, reading goes up

  • increased pressure, reading goes down


13
New cards

monthly QA respiratory gating

  • beam output constancy: 2%

  • phase, amplitude beam control: functional

  • in room respiratory monitoring system & gaing interlock: functional


14
New cards

annual QA check dosimetry

  • x-ray flatness.symmetry: 1%

  • electron flatness/symmetry: 1%


15
New cards

annual QA check safety

per manufacturer

16
New cards

annual QA check mechanical

  • collimator rotation iso: 1mm

  • coincidence of radiation and mechanical iso: 2mm

  • table top sag: 2mm

  • table angle: 1 degree

  • electron applicator interlocks: functional


17
New cards

stability of isocenter

  • stable: convergence @ iso

  • wobble: non-convergence @ iso

  • split field test: used to QA POP fields


18
New cards

split field test

materials: film, plastic tray on 4 feet, 2 shielding blocks

19
New cards

split field technique

  • gantry @ 0, film on tray

  • shielding block on top of film

  • shielding block under tray, opposite side/jaw

  • expose film @ 0 and 180


20
New cards

couch isocenter

  • gantry @ 0, ref. marj w. table at iso height, rotate through various angles

  • not exceed 1mm

  • error: bad ref. mark, collimator, light crosshair alignment


21
New cards

machine dose rate constancy

  • constancy: machine output, cGy/MU

  • spot check confirms baseline parameters

  • often combined with flatness, symmetry, beam penetragtion

  • possible reasons: magnetron current, electron gun current, change in monitor ion chamber

  • instruments: ion chamber, diodes, film, TLD


22
New cards

flatness

  • maximum difference from CA dose over 80% field dimension

  • defined at a depth of 10cm


23
New cards

symmetry

  • difference in dose rate between any 2 points within 80% of field of dimension


24
New cards

FFF

ff removed in beam to attenuate photons


25
New cards

beam penetrative quality

  • PDD/HVL - established upon commissioning using precision instruments

  • spot check - standardize SSD and FS

  • materials: tissue equivalent, ion chamber/diods

  • technique: 2 measurements, compare ratio to established baseline usually at dmax and some depth <70%


26
New cards

imagers QA - kV and MV

  • daily: collision/door interlock, warning lights.sounds, laser/image.tx iso coincidence (2mm), phantom localization and couch shift

  • monthly: kV/MV/laser alignment (1mm), high contrast spatial resolution, CT # accuracy

  • annual: imaging dose and x-ray generator performance