acceptance testing
performed upon equipment installation, by physicists and vendor representative
commissioning
performed by qualified physicist, establishes baselines
QA testing
performed by physicist, equipment calibrated by ADL (accredited dosimetry callibration laboratory)
spot checks
performed by therapist, equipment less precise
array detectors
multiple tests in 1 exposure (output, symmetry, constancy)
daily QA dosimetry
x-ray output constancy: 3%
electron output constancy: 3% (all energies at least 2x weekly)
daily QA mechanical
lasers: 2mm (non IMRT), 1.5 (IMRT), 1mm (SBRT)
ODI: 2mm
collimator size: 2mm (non IMRT/IMRT), 1mm (SBRT)
daily QA safety
door interlock & audiovisual monitor: functional
daily QA checks
water level/temp/pressure
gas pressure SF6!!
general conditions
tx devices
supplies
monthly QA check - dosimetry
x-ray output constancy: 2%
x-ray beam profile constancy: 1%
electron output constancy: 2%
electron beam profile constancy: 1%
backup monitor: 2%
dose rate constancy: 2%
monthly QA check safety
emergency off & wedge interlocks: functional
monthly QA check mechanical
lasers 2mm (non IMRT), 1mm (IMRT, SBRT, SRS)
light vs x-ray: 2mm/1mm (asym)
FS indicators: 2mm/1mm (asym)
tx couch position (2mm/1 degree
gantry/collimator: 1 degree
cross hair center: 1 mm
light field intensity & tray/wedge latches: functional
temperature and pressure corrections
most ion chambers unsealed
chamber sensitivity changes with temp and pressure
increased temp, reading goes up
increased pressure, reading goes down
monthly QA respiratory gating
beam output constancy: 2%
phase, amplitude beam control: functional
in room respiratory monitoring system & gaing interlock: functional
annual QA check dosimetry
x-ray flatness.symmetry: 1%
electron flatness/symmetry: 1%
annual QA check safety
per manufacturer
annual QA check mechanical
collimator rotation iso: 1mm
coincidence of radiation and mechanical iso: 2mm
table top sag: 2mm
table angle: 1 degree
electron applicator interlocks: functional
stability of isocenter
stable: convergence @ iso
wobble: non-convergence @ iso
split field test: used to QA POP fields
split field test
materials: film, plastic tray on 4 feet, 2 shielding blocks
split field technique
gantry @ 0, film on tray
shielding block on top of film
shielding block under tray, opposite side/jaw
expose film @ 0 and 180
couch isocenter
gantry @ 0, ref. marj w. table at iso height, rotate through various angles
not exceed 1mm
error: bad ref. mark, collimator, light crosshair alignment
machine dose rate constancy
constancy: machine output, cGy/MU
spot check confirms baseline parameters
often combined with flatness, symmetry, beam penetragtion
possible reasons: magnetron current, electron gun current, change in monitor ion chamber
instruments: ion chamber, diodes, film, TLD
flatness
maximum difference from CA dose over 80% field dimension
defined at a depth of 10cm
symmetry
difference in dose rate between any 2 points within 80% of field of dimension
FFF
ff removed in beam to attenuate photons
beam penetrative quality
PDD/HVL - established upon commissioning using precision instruments
spot check - standardize SSD and FS
materials: tissue equivalent, ion chamber/diods
technique: 2 measurements, compare ratio to established baseline usually at dmax and some depth <70%
imagers QA - kV and MV
daily: collision/door interlock, warning lights.sounds, laser/image.tx iso coincidence (2mm), phantom localization and couch shift
monthly: kV/MV/laser alignment (1mm), high contrast spatial resolution, CT # accuracy
annual: imaging dose and x-ray generator performance