1/36
Looks like no tags are added yet.
Radio-opaque blue line
Assists with identification on X-ray.
Murphy's Eye
Ensures ventilation occurs even if the primary opening is occluded.
Left-facing bevel tip
Improves the view of the vocal cords during laryngoscopy and offers a narrow point with which to pass the vocal cords.
Left-facing bevel tip diameter
Increases the diameter of primary opening, making it less likely that it will be occluded by secretions
High-volume, low-pressure cuffs
Used in ETT and Trachs to avoid barotrauma.
Increased volume in cuffs
More surface area = less pressure to seal tube against tracheal wall.
Tracheal mucosa protection
High-volume, low-pressure cuffs result in less pressure on tracheal mucosa.
Age cutoff for cuffs
8 yrs. (not absolute)
Cuff Pressure Measurement Techniques
Minimal occlusion volume or Minimal leak technique
Mechanical Manometer & Three-Way Stopcock
Older AG Cuffil (not used anymore).
Posey Cufflator
Cuff pressure manometer used sometimes.
DHD Cuff-Mate2
A type of cuff pressure manometer.
AG Cuffill
Digital cuff pressure manometer that is disposable with less equipment.
Pilot Balloon
Used to measure/fill cuff pressure manometer
Minimal Leak Volume (MLV) / Minimal Leak Technique (MLT)
Auscultate upper airway, add air to cuff until leak disappears, remove a small amount until leak is just heard again.
Minimal Occlusive Volume (MOV) / Minimal Occlusive Technique (MOT)
Auscultate upper airway, add air to cuff until leak just disappears.
Cuff Pressure Manometers
Measure pressure in cuff used to seal trachea.
Normal Cuff Pressure
20 to 30 cmH2O.
Cuff Pressure Manometers are calibrated to
Calibrated to atmospheric P.
Respironics PressureEasy
Helps with tubes having a "leaky" cuff.
Respironics PressureEasy Technique
Adds P to cuff during inspiration, when it is needed most, and reduces cuff pressure during exhalation.
Respironics PressureEasy Ventilator Use
Utilizes positive pressure from the ventilator to add pressure to the cuff during inspiration when leaks are most likely.
Combitube Airway Structure
Double lumen: Esophageal gastric airway & ETT
Combitube Airway Insertion
Designed for blind insertion
Combitube Airway Cuffs
Has 2 cuffs and you inflate both
Combitube Airway Ventilation
Ventilate through either port, confirming which produces chest rise & adequate ventilation
When is a combitube airway used?
When the ETT fails or is unavailable.
Where does the tip of the combitube airway sit?
In the esophagus, posing a high aspiration risk.
Endotracheal Tube Exchanger function
Facilitates the removal and replacement of an endotracheal tube.
What are reasons for an endotracheal tube change?
Blown cuff, upsizing, or being damaged.
How is an Endotracheal Tube Exchanger used?
Threaded through the old tube, remains in place while the old tube is removed, and a new tube is fed over it.
What guarantees an exchanger passes through the vocal cords?
If the old tube was properly in place.
Tracheostomy Tubes
Bypasses the entire upper airway.
Tracheostomy Tube Location
Usually made between the 2nd and 3rd tracheal rings.
Cuffed, Disposable Tracheostomy Tube
Contains a cuff and pilot balloon.
Cuffed, Disposable Tracheostomy Tube Inner Cannula
No disposable inner cannula; the entire trach is changed periodically.
Cuffed, Disposable Tracheostomy Tube Material
Usually made of PVC.