timing cycles

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/55

encourage image

There's no tags or description

Looks like no tags are added yet.

Last updated 10:43 PM on 10/15/26
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai
Chat

No analytics yet

Send a link to your students to track their progress

56 Terms

1
New cards

goal

mimic natural cardiac cycle

2
New cards

most common modes

DDDR, VVIR, AAIR, DOO, VOO, DDIR

3
New cards

VOO/AOO/DOO

asynchronous pacing

No sensing or refractory period

used TEMPORARILY for outpatient surgery or MRI

90 BPM

magnet application and MRI mode pace asynchronously

4
New cards

open reed switch

normal function

5
New cards

closed reed switch

asynchronous pacing; AOO, VOO, DOO

6
New cards

reed switch in ICD

magnet will turn off shocking mechanism (ATP or tachy)

7
New cards

magnet will ___ reed switch

close

8
New cards

VVI

paces and senses; single chamber ICD

9
New cards

VVI 40

single chamber ICD with NO pacing indication and NO afib

10
New cards

VVIR 60

single chamber pacemaker for AFib with 1 lead in RV; indication for ICD and pacing (VT/VF and AFib)

11
New cards

rate response

synthetic SA node

12
New cards

AAI

uncommon in US but done for young people to minimize hardware

AAIR 60; R beacause something wrong with SA node

13
New cards

alert period

see everything, do/don’t something

14
New cards

blanking period

see nothing, do nothing

prevent: oversensing, double counting, crosstalk

15
New cards

cross talk

a-channel talking to v-channel

16
New cards

refractory period

see something, do nothing

does not reset lower rate

allows detection of rapid events (Afib or flutter) (PVARP)

sensed events utilize other functions: detection for mode switch or response to noise

prevent pace maker mediated tachycardia (PMT)

17
New cards

absolute refractory

phase 1-2; blanking period, no amount of power will make impulse

18
New cards

relative refractory

phase 3; refractory strong impulse could cause depolarization

19
New cards

phase 4

alert period

20
New cards

double counting

counting pacing impulse as “second” signal (“M” shaped P-wave)

21
New cards

VOO timing cycle

1 timing cycle, lower rate interval only

22
New cards

VVI timing cycle

3 timing cycle; lower rate limit, ventricular blanking period, ventricular refractory period

23
New cards

ventricular blanking period VBP

mimic absolute refractory period of action potential; begin with paced or sensed event

24
New cards

Ventricular refractory period VRP

mimic relative refractory period of action potential, begin at same time of blanking period

25
New cards

AAI timing cycles

3 timing cycles; lower rate limit, atrial blanking period, atrial refractory period

26
New cards

AV delay

mimic PR interval to allow ventricle to fill from atrial kick

27
New cards

AV delay is ___ channel parameter but results in ___ pacing

atrial; ventricular

28
New cards

post ventricular atrial refractory period

far-field r-wave solution

29
New cards

post-atrial ventricular blanking

far-field p-wave solution

30
New cards

ventricular channel parameter

post-atrial ventricular blanking and crosstalk detection window

31
New cards

crosstalk detection window

shorten AV delay and force V pace; A activity forces V pace

protects patient from unintended pacing inhibition so patient does not skip beat

ventricular safety pacing (V lead only)

occur with atrial pacing

32
New cards

why CDW is necessary

calling AP R-wave inhibits pacing (oversensing), shorten AVD

33
New cards

why use CDW and not lengthen PAVB

PVC or PJC followed by V-pace from AV delay could pace on T-wave (R on T = torsades)

34
New cards

post ventricular atrial blanking (PVAB)

avoid v-a cross-talk and cover farfield v signals

35
New cards

post ventricular atrial refractory period

avoid PMT and V-pace after PAC

36
New cards

LRL

lower rate limit

37
New cards

ventriculoatrial interval

lower rate limit - AV dela; ventricular based timing to protect V

38
New cards

When does LRL time out

when intrinsic R-wave comes through

39
New cards

paced av delay

LONGER to allow for interatrial conduction time between artifical pacing and natural depolarization

conduct slower cell to cell

40
New cards

sensed av delay

shorter

conducts faster when SA node and bachmann’s bundle propogate signal from LA to RA

41
New cards

positive AV hysteresis

extend av delay to allow/search for intrinsic conduction

avoid Unnecessary RV pacing

42
New cards

negative AV hysteresis

shortens AV delay and forces v-pace

common in BiV devices

occurs during CDW safety pacing

43
New cards

DAVID trail

DDDR70 vs VVI40

DDDR70 resulted in higher mortality and HF complications

want positive AV hysteresis and DO NOT RV pace unnecessarily

44
New cards

MOST trial

SSS patient with dual chamber PPM in DDDR or VVIR

unnecessary V-pace associated with increased risk of AFib and HF

want positive AV hysteresis

45
New cards

medtronic and boston V-pace avoidance

MVP & RhythmIQ

change mode between AAIR and DDDR

46
New cards

Abbott V-pace avoidance

VIP

extend paced av delay for up tp 3 cycles to search for intrinsic conduction

47
New cards

biotronik v-pace avoidance

I-Opt

extend paced AV delay to 400ms

48
New cards

can you use V-pace avoidant algorithm on dependant patient

no

49
New cards

AP to AS

lower LRL

50
New cards

VP to VS

extend AV delay

51
New cards

2 ventricular events in a row will be labeled as

PVC

52
New cards

DDIR to DDDR

A fib/flutter ends

53
New cards

DDDR to DDIR

enter afib/flutter

54
New cards

why would someone be programed to DDIR

someone with frequent atrial tach (not afib)

55
New cards

DDIR

no av delay, pacing both chambers looks like DOO with intrinsic atrial beat looks like VVI

56
New cards

why would V see A

high A threshold or sensitive V