Vocabulary practice flashcards covering radiographic artifacts, technical exposure errors, darkroom troubleshooting, manual processing defects, and automatic processor maintenance.
Increased Film Density (Technical Errors)
A radiographic fault characterized by excessive darkness, caused by mAs or kVp settings that are too high, an FFD that is too close (too short), an anatomical part that is too thick or measured incorrectly, equipment malfunction, or intensifying screen speed that is too fast.
Decreased Film Density (Technical Errors)
A radiographic fault resulting in an overly light image, caused by mAs or kVp settings that are too low, an FFD that is too far away, an anatomical area measurement that is too little, intensifying screen speed that is too slow, or a faulty control panel button.
Black Marks or Artifacts
Dark film artifacts caused by film scratches before processing, crescent marks from rough handling, static electricity (linear dots or tree patterns), top-of-film exposure while inside the box, or a defective cassette.

Static Electricity Artifact
A black radiographic artifact appearing as linear dots or a branching tree pattern, caused by static electrical discharge during film handling.
Crescent Marks
Curved, black half-moon artifacts on a radiograph resulting from rough handling or bending of the film prior to processing.
White Marks Artifacts
Clear or radiopaque defects on a finished film caused by dirt or debris between the film and screen, defects or cracks in the intensifying screen, or contrast medium spilled on the tabletop, patient, or cassette.

Damaged or Dirty Intensifying Screen
Pitted surfaces, scratches, dirt, or debris on the screen face that block light emission and produce corresponding white spots or artifacts on radiographs.
Grey Film
A lack of image contrast and uniform dull gray appearance caused by accidental exposure to scatter, secondary, or direct radiation, lack of a grid, expired film, improper storage temperature (too hot or too cold), or incompatibility between the film and screen.
Distorted or Blurred Radiographs
Lack of radiographic sharpness or anatomical distortion caused by motion (patient, cassette, or machine), increased object-film distance (OFD), poor film/screen contact, or improper centering of the primary beam.
Linear Artifacts
Lines appearing on a radiograph caused by grid cut-off, an out-of-focus grid, an uncentered primary beam, an upside-down focused grid, or physical grid damage causing distorted grid lines.
Increased Film Density (Manual Processing)
Excessive darkness produced during manual tank processing when a film is overdeveloped (left in developer solution too long) or the temperature of the solution is too high.
Decreased Film Density (Manual Processing)
Underdeveloped or pale film caused by exhausted developer, contaminated developer, over-diluted developer, a defective thermometer, or failure to add replenisher solution.
Fogged Film
A generalized gray artifact caused by darkroom light leaks, improper safelight type for the film emulsion, or film exposure to scatter radiation when a cassette is left in the room during exposures.

Yellow Radiographs
A yellow-brown radiographic discoloration caused by an exhausted fixer solution or a fixation time that is too short.
White Spots (Chemical / Processing Artifacts)
Defects appearing as white areas on a radiograph, caused by a defective screen (pitted or scratched), debris on the film surface, drops of fixer on the film before processing, or emulsion scratches occurring after processing.
Black Spots (Chemical / Processing Artifacts)
Defects appearing as dark marks on a radiograph, caused by drops of developer solution on the film before processing, films stacked together, static electricity, or scratches occurring before processing.
Air Bells
Round, clear processing artifacts formed when a film is not agitated upon immersion in developer solution, allowing air bubbles to cling to the emulsion and prevent development.
Reticulation
A net-like or mottled artifact in the film emulsion resulting from uneven solution temperatures between the bottom and the top of the chemical processing tanks.
Brittle Radiographs
Dried films that become fragile and prone to cracking due to an excessively high drying temperature or prolonged drying time.
Technical Errors Causing Increased Density (Automatic Processor)
Automatic processor issues leading to excessive film darkness, including developer temperature being too high, over-replenishment, light leaks from the machine, roller speed being too slow, or a faulty thermostat.
Technical Errors Causing Decreased Density (Automatic Processor)
Automatic processor malfunctions resulting in under-density, including developer temperature being too low, under-replenishment, exhausted developer, or a faulty thermostat.

Processing Streaks (Automatic Processor)
Streaking along the path of film travel caused by dirty crossover rollers, dirty wash water, or air tubes requiring cleaning.
Guide Shoe Scratches
Parallel scratch marks across a radiograph caused by mal-aligned guide shoes or mal-positioned dryer air tubes in an automatic processor.
Wet or Damp Film (Automatic Processor)
Incompletely dried film emerging from an automatic processor caused by thermostat malfunction, a dryer temperature that is too low, insufficient air ventilation, or inadequate chemical hardening of the film.
Film Overlap
An artifact occurring in an automatic processor when films stick together or overlap because they were fed into the tray too rapidly or roller tension was set too high.

Hair in Cassette Artifact
A radiopaque artifact resembling slender, curved white lines caused by loose strands of hair trapped between the intensifying screen and the film emulsion.

Chemicals Not Stirred Artifact
An uneven film density gradient across a radiograph that occurs when manual processing tank solutions stratify by temperature or concentration and are not stirred before development.