Vocabulary practice flashcards covering the physical principles, equations, pulse sequences, spatial encoding gradients, k-space mechanics, and contrast weighting of Magnetic Resonance Imaging (MRI).
Magnetic Resonance Imaging (MRI)
A major non-invasive medical imaging modality that uses a strong static magnetic field, magnetic field gradients, and radiofrequency (RF) pulses to create anatomical and physiological images by imaging hydrogen nuclei (1H protons) found in water and fat molecules.
Nuclear Spin
A fundamental physical property of subatomic particles and atomic nuclei, analogous to electric charge or mass, which exists in integer or half-integer multiples of 1/2 (or 21) and produces a magnetic dipole when associated with a moving electrical charge.
Gyromagnetic Ratio (γ)
A proportionality constant specific to a given atomic nucleus that relates its precession frequency to the strength of the external magnetic field. For a hydrogen proton (1H), γ=42.58MHz/T (commonly rounded to 42.6MHz/T).
Bulk Magnetization (M0)
The net macroscopic magnetic vector produced by the vector sum of individual nuclear magnetic moments per unit volume aligning with a strong static external magnetic field (B0) at thermal equilibrium.
Larmor Equation
The governing equation f=γB0, which states that the precession frequency (f, or Larmor frequency) of a nuclear spin is directly proportional to the gyromagnetic ratio (γ) and the strength of the external magnetic field (B0).
Larmor Frequency (f0)
The rate at which nuclear magnetic moments precess around the direction of an external static magnetic field (B0); for example, hydrogen protons precess at approximately 63.9MHz in a 1.5T magnetic field.
Resonance
The physical condition achieved when an applied radiofrequency (RF) pulse oscillates at the exact Larmor precession frequency of the nuclei, enabling the net magnetization vector to absorb energy and rotate away from the longitudinal axis.
Flip Angle (α)
The angle through which the net magnetization vector is tipped away from the longitudinal (z) axis toward the transverse (x-y) plane during RF excitation, mathematically defined by α=2πγB1τ, where B1 is the RF magnetic field strength and τ is pulse duration.
90∘ RF Pulse
An RF excitation pulse that rotates the net magnetization vector by exactly 90∘ from the longitudinal axis into the transverse (x-y) plane, setting the transverse component Mxy equal to M0.
180∘ RF Pulse
An RF pulse with twice the duration or power of a 90∘ pulse that inverts the longitudinal magnetization to −M0 or mirrors the dephased transverse spin vectors across an axis to initiate spin rephasing.
Free Induction Decay (FID)
The transient, oscillating, and exponentially decaying electrical voltage induced in a receiver coil immediately after an RF excitation pulse as transverse spins dephase.
Bloch Equations
A system of macroscopic differential equations that model the time-dependent evolution of the net magnetization vector components (Mx, My, Mz) in the presence of applied magnetic fields, incorporating T1 and T2 relaxation rates.

Longitudinal Relaxation (T1 Relaxation)
The thermal or spin-lattice relaxation process by which the longitudinal magnetization component (Mz) recovers back to its equilibrium value (M0) through energy exchange with the surrounding molecular lattice, following Mz(t)=M0(1−e−t/T1).
T1 Relaxation Time
The time constant required for the longitudinal magnetization (Mz) to regrow to approximately 63% of its final equilibrium value (M0) following RF excitation.
Transverse Relaxation (T2 Relaxation)
The spin-spin relaxation process by which transverse magnetization (Mxy) decays irreversibly due to random microscopic variations in local magnetic fields causing dephasing among neighboring spins, following Mxy(t)=M0e−t/T2.
T2 Relaxation Time
The time constant required for the transverse magnetization (Mxy) to decay to approximately 37% of its initial post-excitation amplitude.
T2∗ (T2-Star) Relaxation
The effective transverse relaxation decay caused by the combined effects of true spin-spin interactions (T2) and static magnetic field inhomogeneities (ΔB), governed by the relationship T2∗1=T21+γΔB.

Tissue Relaxation Parameters (T1 and T2)
Inherent tissue-specific time constants; for example, at 1.5T, skeletal muscle exhibits T1=870ms and T2=47ms, fat exhibits T1=260ms and T2=84ms, gray matter exhibits T1=920ms and T2=101ms, and white matter exhibits T1=790ms and T2=92ms.

Slice Selection Gradient (Gz)
A linear magnetic field gradient applied along the z-axis simultaneously with an RF pulse to vary Larmor frequencies along the patient's body axis, selectively exciting only a specific anatomical slice matching the RF bandwidth.
RF Bandwidth
The range of transmitted frequencies within an RF pulse that, together with the slope of the slice-selection gradient (Gz), defines the thickness of the excited anatomical slice.
Phase Encoding Gradient (Gy)
A magnetic field gradient applied momentarily along the y-axis prior to signal readout that imparts a position-dependent phase shift to protons across different rows within the selected slice.
Frequency Encoding Gradient (Gx)
A magnetic field gradient (also known as the readout gradient) applied along the x-axis during echo signal sampling, causing proton precession frequencies to vary linearly by column position across the slice.
Gradient Coils
Specialized electromagnetic coils (Gx, Gy, Gz) configured within an MRI bore to produce controlled linear variations in the main magnetic field along three orthogonal directions for spatial encoding.

Spin Echo (SE) Sequence
A standard MRI pulse sequence consisting of a 90∘ excitation pulse followed after time interval τ=TE/2 by a 180∘ refocusing pulse, producing a refocused spin echo signal at TE=2τ while eliminating static field inhomogeneity dephasing.

Repetition Time (TR)
The duration between successive 90∘ RF excitation pulses applied to the same slice, determining the amount of longitudinal (T1) recovery that occurs between cycles.
Echo Time (TE)
The time elapsed from the center of the initial 90∘ RF excitation pulse to the peak of the formed spin echo signal, controlling the degree of transverse (T2) decay.
k-Space
An array or matrix representing the 2D spatial frequency domain in which raw digitized MRI echo signals are stored; each line in the matrix corresponds to a separate phase-encoding step.
Central k-Space
The region near the center of the k-space matrix that holds low spatial frequency data, which determines the overall contrast, brightness, and general shapes within the reconstructed image.
Peripheral k-Space
The outer border regions of the k-space matrix that hold high spatial frequency data, which determines image edge sharpness, fine detail, and spatial resolution.
2D Fourier Transform (2DFT)
The mathematical algorithm applied to k-space raw data to transform spatial frequency information into a reconstructed two-dimensional spatial-domain anatomical image.
T1-Weighted Image
An MRI sequence acquired with a short TR (<600ms) and a short TE, where tissues with rapid longitudinal regrowth (short T1, such as fat) appear hyperintense (bright) and tissues with slow regrowth (long T1, such as CSF) appear hypointense (dark).
T2-Weighted Image
An MRI sequence acquired with a long TR (>600ms or >1500ms) and a long TE, where tissues with slow transverse dephasing (long T2, such as CSF, edema, and lesions) appear hyperintense (bright) and tissues with fast dephasing (short T2, such as muscle) appear dark.
Proton Density (PD)-Weighted Image
An MRI sequence acquired with a long TR (minimizing T1 effects) and a short TE (minimizing T2 effects), producing image contrast predominantly dictated by differences in tissue proton concentration (N(H)).
MRI Image Contrast Matrix
A operational framework matching sequence parameters to tissue weighting: short TR / short TE produces T1 contrast; long TR / long TE produces T2 contrast; long TR / short TE produces Proton Density contrast; and short TR / long TE yields poor diagnostic contrast.