digital image capture (3 Cs)
Capture element, coupling element, collection element
Capture element (1/3 C’s)
absorbs exit radiation
Coupling element (2/3 C’s)
transfers signal to collection element
collection element (3/3 C’s)
collects and temporarily stores signal
Computed Radiography (CR)
two separate steps in image acquisition: image capture and readout
Direct radiography
combines image capture and image readout into one device
CR system components
imagine plate (IP) in a cassette, reader, workstation
The imaging Plate: Basic construction
support, protective & phosphor layers
The imaging plate: phosphor layer
barium fluorohalide doped in europium
CR Reader
drive mechanism, optical system, photodetector (PMT), ADC
CR Image production step (1/4)
Remnant radiation interacts with Image plate
CR image production step (2/4)
latent image stored in electron traps
CR image production step (3/4)
energy released in the reader in the form of light
CR image production step (4/4)
computer processing makes the latent (invisible) image visible (manifest)
direct radiography uses
flat panel detectors (FPD)
DR system component (1/3)
solid-state IRs using a large active-matrix array (AMA)
DR system components (2/3)
combines image capture & readout
DR system components (3/3)
“self-scanning”
Flat panel detector (FPD) Design: basic layers
X-ray converter, TFT array, substrate (glass)
Flat panel detector (FPD) Design: detector elements (DEL)
array is divided into detector elements (microns)
decreased DEL size = increased _____
spatial resolution
3 Parts of the DEL
detection surface, capacitor, TFT
DEL design
solid-state diode
sensing area = signal collection (sensitive to light or x-rays)
dead space = no signal collection
fill factor
% of DEL dedicated to sensing area
greater fill factor increases sensitivity
DEL: efficiency is measured as
DQE
From patient → IR → Monitor: Voxel
a volume element (or volume of tissue) within the patient
From patient → IR → Monitor: Dexel
a detector element (DEL) in the image receptor
From patient → IR → Monitor: Pixel
a picture element in the final image
information from different voxels (in patient) is collected by the _____
DEL’s (of IR) and computer-processed to become the pixels (of final images)
The TFT requires a ____
small electrical charge before exposure (hardwired, tether, connection, rechargeable battery)
Generator and ____ must communicate to synchronize image acquisition
FPD
Pre-exposure: ____ elements are charged to prepare to receive signal
TFT
during exposure: signal ____ and ____
conversion, collection
post-exposure: signal storage, ______, and prep for next exposure
readout
FPD Categories
indirect & direct conversion
All indirect detectors
use a scintillator, use a coupling element, two-step process
3 types of indirect detectors
TFT, CCD, or CMOS
scintillators
produce light when stimulated with radiation
2 types of scintillators
unstructured/turbid, structured
scintillator: turbid
gadolinium oxysulfide
scintillator: structured
cesium iodide
indirect conversion image production: scintillator
converts exit radiation into visible light
indirect conversion image production: photodetector/photodiode
converts light to electrical charges
indirect conversion image production: electrical signals are
digitized by the ADC
Indirect conversion
Scintillation material (gadolinium oxysulfide, cesium iodide)
“contact layer”
Amorphous silicon TFT array
glass substrate
indirect TFT image production step (1/5)
exit radiation causes scintillator to fluoresce
indirect TFT image production step (2/5)
TFT detectors convert light into electrical charges
indirect TFT image production step (3/5)
electrical charges stored in the capacitor
indirect TFT image production (4/5)
TFT acts like a switch to release stored signals
indirect TFT image production step (5/5)
signal converted by ADC (indirect)
Charge-coupled-device (CCD) components
very light sensitive & wide dynamic range
uses Csl scintillator
tiling used to make larger IR
CCD image production steps (1/3)
exit radiation causes scintillator to fluoresce
CCD image production steps (2/3)
optical lens or fiber optics channel light to CCD
CCD image production steps (3/3)
CCDs convert light into electrical charges
CMOS components
scintillator (crystalline silicon matrix)
Amplifier
photodiode
storage capacitor
ADC
CMOS image production: (1/5)
exit radiation causes scintillator to fluoresce
CMOS image production: (2/5)
Optical lens or fiber optics channel
light to CMOS
CMOS image production: (3/5)
CMOS converts light into electrical
charges
CMOS image production: (4/5)
Electrical charges stored in a
capacitor
CMOS image production: (5/5)
signal digitized in each CMOS DEL
Direct TFT components
one step/single stage conversion process
does not need a coupling element
amorphous selenium TFT array
Direct TFT image production: (1/5)
amorphous selenium converts radiation into electrical charges
Direct TFT image production: (2/5)
signal stored in a capacitor (DR)
Direct TFT image production: (3/5)
electrical signals are sent to the ADC for digitization
Direct TFT image production: (4/5)
TFT acts like a switch to release stored signals
Direct TFT image production: (5/5)
signals converted by ADC
converting the signal
signal sent to ADC
digitized information arranged to form matrix
each pixel has a brightness level
ALWAYS signal loss
ADC & quantization: pixel
numeric value assigned to each sampling point
ADC & quantization: shades of gray
pixel bit depth - contrast resolution
Sampling the analog signal: X-axis of graph
sampling points per unit time (frequency)
Sampling the analog signal: Y-axis of graph
sampling pitch
Aliasing
too few samples will create and aliased image
preventing aliasing: nyquist frequency
ensures accurate representation
sampling must be at least 2x the highest spatial frequency
DR panel care
manufacturer-approved disinfectants
avoid drops, excessive pressure, improper handling
protective cover to protect panel from contamination
damaged panels often returned for detailed failure analysis
quality control: physicists & vendors
equipment testing, calibration, & maintenance
quality control: radiographers
QC tasks
visual checks (artifacts) & routine cleaning
quality control: performance monitoring
identify quality changes that may indicate malfunction
Quality control key takeaway
Quality control is a shared responsibility
Regular QC by radiographers
testing by physicists and vendors, helps ensure optimal
image quality and system reliability