Faults Observed During Section Cutting and Tissue Processing

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/31

flashcard set

Earn XP

Description and Tags

Flashcards covering causes and remedies for various faults observed during section cutting and tissue processing in histology (References: Book by Gregorios, Lo and Bancroft's; Compiled by Ma. Cristina SJ Liwanag).

32 Terms

1
New cards

Disintegration or Expansion of Sections on Water Bath

A section cutting fault caused by poor tissue impregnation or water temperature being too high in the flotation bath. Remedied by re-processing tissue or turning down the flotation bath temperature.

2
New cards

Sections Rolling into a Coil

A fault during section cutting caused by a dull blade, a rake angle that is too small, or a section that is too thick. Remedied by using a new blade, reducing blade tilt if clearance angle is excessive, or reducing section thickness.

3
New cards

Grooved, Scored, Smeared, and Deformed Sections

A fault caused by a dull knife or blade edge, or an overly warm room. Remedied by re-sharpening/replacing the blade or cooling the room/block in ice water.

4
New cards

Sections Failing to Form Ribbons

Occurs when block edges are not parallel, horizontal block surface is not parallel to knife, paraffin wax is too hard, knife tilt is excessive, sections are too thick, or knife is dull. Remedied by re-trimming/re-adjusting the block, coating edges with lower melting point wax, reducing tilt, or honing and stropping.

5
New cards

Sections Rolling Up and Adhering to Knife Edge

A section cutting fault caused by a blunt knife, excessive knife tilt, or a dirty knife edge. Remedied by sharpening the knife, reducing tilt, and cleaning the knife edge.

6
New cards

Curved, Crooked, or Uneven Ribbon

A fault caused by a dull spot on the knife, non-parallel/round/wedge-shaped block edges, knife not parallel to block, or impure paraffin. Remedied by sharpening/adjusting the knife, re-trimming the block, readjusting alignment, or repeating impregnation with pure wax.

7
New cards

Compressed, Wrinkled, or Jammed Sections

A fault caused by a blunt knife, warm/soft block, paraffin-coated knife edge, thin sections, loose set screw, vertical tilt, rapid cutting, warm room, or non-parallel block edges. Remedied by sharpening knife, cooling block on ice water, tightening screws, reducing tilt, and decreasing cutting speed.

8
New cards

Squashed Sections

A fault where section width is less than block width, caused by loss of knife bevel due to incorrect sharpening. Remedied by re-sharpening using a knife back or automatic knife sharpener.

9
New cards

Hole Formation in Section

A section cutting fault caused by bubbles/dirt in embedding medium, calcium hard spots, or excessive dehydration/improper processing. Remedied by re-embedding in filtered wax, using a base sledge microtome with wedge knife, or soaking exposed tissue in ice water/cotton before sectioning.

10
New cards

Sections of Unequal Thickness

A fault caused by excessive knife tilt, loose clamp set screws, or blocks that are too large or too hard. Remedied by reducing tilt, tightening screws, cutting blocks into smaller fragments, or softening blocks in detergent or phenol.

11
New cards

Sections Adhering to Knife or Microtome Parts

A fault caused by static electricity from low atmospheric humidity, dirty or dull knife edge, or excessive knife tilt. Remedied by blowing gently on the block, boiling water in the room, cleaning/sharpening the knife, or reducing tilt.

12
New cards

Split Ribbon or Lengthwise Vertical Scratches

A fault caused by nicks or damage on the knife edge, dirty embedding material, dirty knife edge, or excessive knife tilt. Remedied by sharpening knife, re-embedding in filtered wax, cleaning knife edge with xylene, or reducing tilt.

13
New cards

Sections Lifted from Knife on Upstrokes

A fault caused by excessive knife tilt, a dull knife, or soft paraffin/warm room temperature. Remedied by reducing tilt, sharpening knife, or cooling paraffin wax in ice water.

14
New cards

Resistance Felt on Lower Part of Section

A section cutting fault caused when knife tilt is too small, compressing the paraffin block against the base of the knife towards the end of the stroke. Remedied by increasing knife tilt.

15
New cards

Chatters (Horizontal Furrows Across Section)

Parallel lines or furrows creating thin and thick zones, caused by knife edge vibration from hard tissue, excessive tilt, improper tissue moisture, dull knife, or rapid cutting. Remedied by treating tissue with phenol/collodion, reducing tilt, restoring moisture in ice water, sharpening knife, or slowing cutting speed.

16
New cards

Alternating Thin and Thick Section Cuts

A fault caused by a blunt knife, improperly clamped knife, excessive tilt, loose holder, or knife tilt being too small. Remedied by sharpening/adjusting knife, adjusting tilt, or tightening locking screws.

17
New cards

Metallic Scraping or Ringing Sound on Backstroke

A sound produced during cutting when knife tilt is too slanted or large, tissue is too hard, or knife blade is too thin. Remedied by readjusting knife angulation, using a fresh block treated with phenol, or changing the knife.

18
New cards

Crumbling Frozen Tissue

A fault where frozen tissue crumbles and comes off the block holder when cut due to inadequate freezing. Remedied by refreezing the tissue block.

19
New cards

Fragmenting/Chipping Frozen Tissue

A fault where frozen tissue chips into fragments when cut because tissue is frozen too hard. Remedied by warming tissue with fingers.

20
New cards

Pleated, Folded, and Crooked Ribbons

A fault caused by imperfections in the blade edge or non-parallel block edges relative to the knife edge. Remedied by trimming block parallel, re-embedding, or trying another part of the knife edge.

21
New cards

Brittle or Hard Tissues

A tissue processing fault resulting from prolonged fixation, dehydration, clearing, or paraffin infiltration, overheated oven, or drying before fixation. Remedied by soaking in water with detergent, phenol, or molliflex.

22
New cards

Clearing Agent Turning Milky

A processing fault occurring immediately upon tissue placement due to incomplete dehydration leaving water in tissue. Remedied by repeating dehydration with absolute alcohol before clearing again.

23
New cards

Tissue Smelling of Clearing Agent on Trimming

A processing fault caused by insufficient paraffin impregnation failing to remove clearing agent completely. Remedied by trimming near tissue, melting wax, and repeating impregnation with paraffin changes.

24
New cards

Opaque Tissue with Difficult Sectioning

A fault caused by presence of alcohol due to insufficient clearing. Remedied by repeating clearing (or prolonging up to 12 hours if already embedded) and re-embedding.

25
New cards

Tissue Shrinking Away from Wax

A tissue processing fault caused by insufficient dehydration leading to incomplete clearing and impregnation. Remedied by repeating the whole processing procedure.

26
New cards

Soft Tissue During Block Trimming

A tissue processing fault caused by incomplete fixation. Remedied by repeating the whole processing procedure.

27
New cards

Airholes in Tissue During Trimming

A tissue processing fault caused by incomplete paraffin impregnation. Remedied by repeating impregnation.

28
New cards

Crystalline Appearance of Wax on Trimming

A processing fault caused by contaminated wax or failure to cool the block rapidly enough. Remedied by re-embedding in freshly filtered wax.

29
New cards

Moist and Crumbling Paraffin Block After Cooling

A fault caused by insufficient paraffin impregnation. Remedied by repeating paraffin impregnation and re-embedding.

30
New cards

Inadequate Processing of Fatty Tissues

A fault occurring when routine processing fails to clear fat adequately because ethanol is a poor fat solvent. Remedied by thin sampling and inserting a lipid solvent (acetone or isopropanol) before final absolute ethanol.

31
New cards

Hard, Dense Tissues due to Keratin

A tissue condition caused by tough keratin components in certain tissues. Remedied by treating fixed tissues with 4%4\% aqueous phenol for 24-72 hours to soften.

32
New cards

Soft and Mushy Tissue Center on Embedding

A fault caused by inadequate processing due to very thick or fatty sections. Remedied by cutting thin sections at gross and ensuring complete fixation.