Absorbed dose
Amount of energy deposited by radiation in a mass; total energy from ionizing radiation absorbed per unit mass
What can the "mass" be in absorbed dose?
Anything: water, rock, air, people, etc.
Absorbed dose units
Expressed in milligrays (mGy); SI unit = Gray (Gy); traditional unit = rad; 1 Gy = 100 rad
Equivalent dose
Calculated for individual organs; based on the absorbed dose to an organ, adjusted to account for the effectiveness of the type of radiation
Equivalent dose units
Expressed in millisieverts (mSv) to an organ
Effective dose
Calculated for the whole body; the addition of equivalent doses to all organs, each adjusted for the sensitivity of the organ to radiation
Effective dose units
Expressed in millisieverts (mSv); SI unit = Sievert (Sv)
Effective dose: what it is used for and what it considers
Sum of equivalent doses weighted by radiosensitivity of exposed tissue/organ; used to estimate risk in humans; considers relative biologic effectiveness of different radiation types and radiosensitivity of different tissues in terms of risk for cancer formation or heritable effect
Total radiation exposure (US population, NCRP Report 160)
Grand total 6.2 mSv (natural background 3.1 + medical 3.0 + consumer products and other 0.1)
Natural background radiation total
3.1 mSv (radon 2.3, space 0.3, internal radionuclides 0.3, terrestrial 0.2)
Artificial exposure total
3.1 mSv (medical imaging 3.0 mSv; consumer products and other 0.1 mSv)
Radon exposure dose
2.3 mSv (largest natural source)
Space (cosmic) exposure dose
0.3 mSv
Internal radionuclides exposure dose
0.3 mSv
Terrestrial sources exposure dose
0.2 mSv
Medical exposure breakdown (mSv)
CT 1.5; nuclear medicine 0.8; interventional fluoroscopy 0.4; conventional radiography and fluoroscopy 0.3; dental 0.007; subtotal medical 3.0
Consumer products and other exposure dose
0.1 mSv
Source of the exposure data tables
National Council on Radiation Protection and Measurements (NCRP) Report 160, Ionizing Radiation Exposure of the Population of the United States, Bethesda, MD, 2009
Radon source of radiation
Radon-222 decays (polonium-218); emits alpha particles
Space (cosmic) radiation depends on what?
Altitude; sea level = 0.33 mSv/year; at 1600 m = 0.50 mSv/year
Internal radionuclides source
Foods containing uranium and thorium
Terrestrial radiation source
Potassium-40, uranium and others; found in the top 20 cm of soil
Radon Zone 1 (red)
Highest potential; predicted average indoor radon screening level greater than 4 pCi/L
Radon Zone 2 (orange)
Moderate potential; predicted average indoor radon screening level between 2 and 4 pCi/L
Radon Zone 3 (yellow)
Low potential; predicted average indoor radon screening level less than 2 pCi/L
Pennsylvania radon zones (Zone 1 and Zone 3)
Most PA counties are Zone 1 (red, highest potential); Philadelphia County is Zone 3 (yellow, low potential)
Pennsylvania Zone 2 (orange) counties
Erie, Crawford, Warren, McKean, Potter, Forest, Elk, Jefferson, Mercer, Lawrence, Cambria, Somerset, Fayette, Greene, Washington, Wayne, Pike
Natural background radiation pie chart (3.1 mSv/yr)
Radon 73%; space 11%; internal radionuclides 9%; terrestrial 7%
Artificial sources pie chart (3.1 mSv/yr)
CT 47%; nuclear medicine 25%; interventional radiography and fluoroscopy 14%; conventional radiography and fluoroscopy 10%; consumer products and other 4%; dental 0.26%
Medical x-ray exams in the US
400 million x-ray exams; 25% are dental
Largest medical contributors to artificial exposure
CT and cardiac nuclear medicine
Dental share of total medical exposure
0.26%
Consumer product sources of radiation
Cigarette smoking, building materials, air travel, mining agriculture, fossil fuels
Other artificial sources of radiation
Caregivers of nuclear medicine treatment patients, radiation workers
ALARA
As Low As Reasonably Achievable; guiding principle for exposure and dose in radiography
Deterministic (nonstochastic) effects: goal
Prevent their occurrence
Stochastic effects: goal
Reduce the likelihood of occurrence
Two ways to estimate risk from exposure
Using equivalent natural (background) exposure, or the probability of stochastic effect
Intraoral dose with rectangular collimation
Posterior bitewings (PSP or F-speed): 5 µSv = 0.6 day; full-mouth PSP or F-speed: 40 µSv = 5 days; full-mouth CCD sensor (estimated): 20 µSv = 2.5 days
Intraoral dose with round collimation
Full-mouth D-speed film: 400 µSv = 48 days; PSP or F-speed: 200 µSv = 24 days; CCD sensor (estimated): 100 µSv = 12 days
Extraoral exam doses
Panoramic 20 µSv = 2.5 days; cephalometric 5 µSv = 0.6 day; chest 100 µSv = 12 days
Cone beam CT doses
Small FOV (<6 cm) 50 µSv = 6 days; medium FOV (dentoalveolar, full arch) 100 µSv = 12 days; large FOV (craniofacial) 120 µSv = 15 days
Multidetector CT doses
Maxillofacial 650 µSv = 2 months; head 2 mSv = 8 months; chest 7 mSv = 2 years; abdomen and pelvis with and without contrast 20 mSv = 7 years
Organizations that recommend annual effective dose limits
NCRP (National Council on Radiation Protection and Measurements) and ICRP (International Commission on Radiation Protection)
Purpose of annual effective dose limits (stochastic effects)
Ensure the probability of stochastic effects is as low as reasonably and economically feasible
Purpose of annual effective dose limits (occupational risk)
Ensure the risk of radiation workers being afflicted with fatal cancer is ≤ the risk of fatal accidents in non-radiation occupations
Occupational dose limits
50 mSv per year (NCRP); 20 mSv per year averaged over 5 years (ICRP)
Non-occupational (public) dose limits
10% of occupational limits; 5 mSv per year for infrequent exposure; 1 mSv per year for continuous exposure
Operator protection: barrier materials
Lead; gypsum drywall, brick, or concrete (lead lining rarely needed); 3/4 inch glass
Operator protection: distance and position
At least 2 m (about 6.5 ft) from the x-ray source, at 90° to 135° to the x-ray beam
Who may assist in an exposure?
Patient's relative, with protection; never use a staff member
Three properties of x-rays relevant to operator protection
Travel in straight lines from the source; intensity diminishes rapidly with distance; can be scattered or deflected
Guiding principle for diagnostic radiology in dentistry (JADA 2001; reaffirmed ADA 2024)
Enhance the diagnostic benefits of dental radiographs and minimize the associated radiation risks to patients and staff
Patient protection: significance in diagnosis
Consider signs, symptoms, radiographic history, and positive findings
Intraoral receptor selection: film speed
Use the fastest speed film; F-speed requires only 40% of the exposure of D-speed film
Intraoral receptor selection: digital
Digital sensors (CCD/CMOS) give about 50% patient dose reduction vs F- or E-speed film; PSP plates are similar to F-speed film
Extraoral receptor selection: screens
Use intensifying screens; rare earth screens decrease patient exposure by 55% compared with older calcium tungstate screens
Digital receptors in panoramic and cephalometric radiography
No significant dose reduction
Collimation: exposure area size
Exposure area less than 7 cm (2.75 inches) in diameter
Collimation: PID type
Open end PID (position-indicating device)
Rectangular collimation benefits vs round
~60% less skin area exposed; ~50% lower thyroid dose; up to 5× lower effective dose
Focal-film distance: federal regulation
X-ray source to skin distance must not be less than 7 inches
Focal-film distance: recommendation
Use longer FFD (8, 12, or 16 inch PIDs)
Filtration requirement
2.5-mm equivalents of aluminum
Source-to-skin distance and collimator diagram (A, B, C)
A: short source-to-skin distance; B: longer source-to-skin distance; C: rectangular collimator, giving a smaller, less divergent beam and a smaller volume of tissue irradiated than A and B
Kilovoltage for intraoral radiographs
60-100 kVp; 70 kVp is recommended
Thyroid collars and lead aprons: national guidance (AAOMR 2023; ADA 2024)
No longer routinely recommended; no routine thyroid or gonadal shielding for any dental imaging
Rectangular collimation vs thyroid collar
Rectangular collimation lowers thyroid dose more than a collar does; collars can block the beam and cause retakes
Our clinic's thyroid collar policy
Still use a thyroid collar for intraoral imaging, because we use round collimation and learners are still building technique
Thyroid collar placement rule
Keep the collar out of the primary beam; if it shows on the image, it caused the retake
Receptor holders: benefit
Reduce retakes and radiation exposure
Patient protection: exposure and processing
Proper exposure and process time; darkroom practices (light tight, ventilation); infection control practice; no skin contact with chemicals
Patient protection: quality assurance, image viewing, education
Monitor x-ray generators, processors, and chemicals at regular intervals; ideal image viewing conditions; continuing education
Communicating radiation risk: first steps
Allow patients to fully express concerns (don't interrupt or dismiss); explain the specific diagnostic need for radiographs for their condition; describe protective measures (selection criteria, fast digital receptors, collimation)
Communicating radiation risk: background comparisons
Bitewings: less than 1 day with rectangular collimation, a few days with round; full-mouth series: about 12-24 days with round (2.5-5 days rectangular); panoramic: 2-3 days of natural background; 5-hour flight = 25 µSv
What to emphasize when talking with patients about dose
"Dental radiographs deliver negligible to minimal doses relative to other medical imaging"
5 steps for talking with patients about radiographs
1 Listen (let them voice concern fully); 2 Justify (why this image, for this patient, now); 3 Protect (name what we do to keep dose low); 4 Compare (days of background, or a flight); 5 Respect (if they decline, explain what we could miss, and document it)
Patient communication: what to say vs avoid
Say: "This x-ray shows me decay between your teeth that I can't see by looking." and "It's about the same as a few days of the natural radiation we all get." Avoid: "There's zero radiation. It's totally safe." (not accurate) and "Everyone gets these every year." (images are prescribed by need, not by the calendar)
ACS lifetime cancer probability (2008-2010)
Developing any cancer: males 43.92% (1 in 2), females 38.00% (1 in 3); dying from cancer: males 22.94% (1 in 4), females 19.34% (1 in 5); prostate 15.33% (1 in 7) in males
Risk perspective: baseline, dental share, occupational dose
~40% of people develop cancer in their lifetime; dental radiography is 0.26% of total medical radiation exposure; average occupational dose in dentistry is about 0.06 mSv per year (ADA 2024), far below the 20-50 mSv annual limits
Risk perspective: pregnancy, everyday risk, benefit vs risk
Fetal doses from dental radiography are 42,000-fold lower than the threshold for deterministic effects; dental x-rays carry less risk than many everyday activities; diagnostic benefits far outweigh minimal radiation risks