Cell: How do you get the cell potential from the two electrode potentials? (L16 s9)
> Ecell = E_cathode − E_anode (cathode = reduction, anode = oxidation)
> Galvanic (spontaneous) → Ecell > 0
> Nernst for the cell: Ecell = E°cell − (RT/zF) · ln Q
∴ Ecell = cathode − anode, + means spontaneous
Cell: What is required in an electrochemical cell?
> Redox reaction: something gets oxidized, something gets reduced
> Anode: the electrode where oxidation happens
> Cathode: the electrode where reduction happens
> Electrolyte: the ion-conducting stuff between the two electrodes (ions move thru here)
> External circuit = e⁻-conducting wire between the two electrodes (e⁻ move thru here)

Cell: What does a voltammetry setup need, and what is the Luggin capillary for? (L17 s23–24)
> Potentiostat + WE + CE + RE (in a Luggin capillary) + electrolyte
> CE = Pt mesh / gauze (large area to carry current)
> Luggin capillary → RE tip right next to the WE
> Frit → separates catholyte (around cathode) + anolyte (around anode) → ions pass, CE products don't reach the WE ( my reasoning)
∴ 3 electrodes + potentiostat; RE measured close to the WE
Electrode: What is the electrochemical double layer, and how does it form? (L16 s5, Skoog 22A-5)
> Electrode = conductor in an electrolyte
> Metal (Fe) → some Fe²⁺ dissolve → e⁻ left in metal → surface (-)
> (-) surface attracts (+) ions → (-) layer (metal) + (+) layer (solution) = double layer
> → electric field forms at the interface
> IHP → oriented solvent dipoles + ions touching the surface
> OHP → closest approach of solvated ions
> Diffuse layer → ions spread out → bulk solution
> Double layer acts like a capacitor → ↑ surface area → ↑ capacitance → ↑ non-Faradaic current (review sheet)
∴ conductor in solution → charge separates at the interface → double layer + electric field

Electrode: How are electrode potentials measured, and what is the common reference? (L16 s6)
> Interfacial potential difference → can't be measured directly
> → measure vs. a reference electrode (voltmeter between the 2 electrodes)
> Common reference = NHE → H₂ → 2H⁺ + 2e⁻
> E° = 0 V @ pH 0, p(H₂) = 1 atm, 298 K
∴ electrode potentials are always relative → reported vs. a reference (NHE = 0 V)
Electrode: What is the standard electrode potential (E°)? (L16 s7)
> E° = reduction potential of the electrode (cathode) vs oxidation potential of the reference electrode (anode)
> Standard conditions → all species a = 1 (c = 1 M, p = 1 atm)
> T must be stated (normally given, usually 298 K)
∴ E° = reduction potential vs reference, all a = 1
Electrode: What are ideal polarized vs ideal non-polarized electrodes, and which one should a reference electrode be? (L16 s18, s20)
> Ideal polarized → potential changes, current stays 0 (flat line on the current vs E plot)
> Ideal non-polarized → current changes, potential stays fixed (vertical line)
> Neither exists in reality
> Reference electrode → ideal non-polarizable → stable E with little dependence on T, c, and current
> Purpose of reference → current-independent (stable) reference value for E measurements (midterm review Q)
∴ Reference = non-polarizable → E doesn't move
Electrode: How do the Ag/AgCl and SCE reference electrodes work? (L17 s3–4)
> Ag/AgCl: AgCl + e⁻ ⇌ Ag + Cl⁻ → Ag wire coated with AgCl in KCl, frit to the sample
> SCE: Hg₂Cl₂ + 2e⁻ → 2Hg + 2Cl⁻ → Hg / Hg₂Cl₂ in saturated KCl
> Both: E = E° − (RT/F) ln[Cl⁻] → E depends only on [Cl⁻]
> Fixed [Cl⁻] (e.g. 1.0 M or saturated KCl) → fixed, stable E
∴ Fixed [Cl⁻] = stable reference E
![<p>> <strong>Ag/AgCl: AgCl + e⁻ ⇌ Ag + Cl⁻</strong> → Ag wire coated with AgCl in KCl, frit to the sample</p><p>> <strong>SCE: Hg₂Cl₂ + 2e⁻ → 2Hg + 2Cl⁻</strong> → Hg / Hg₂Cl₂ in saturated KCl</p><p>> <strong>Both: E = E° − (RT/F) ln[Cl⁻]</strong> → E depends only on [Cl⁻]</p><p>> <strong>Fixed [Cl⁻] (e.g. 1.0 M or saturated KCl) → fixed, stable E</strong></p><p>∴ <strong>Fixed [Cl⁻] = stable reference E</strong></p>](https://assets.knowt.com/user-attachments/8d96576d-569e-43ac-ae44-1f4ec7893b5a.png)
*Electrode: What are the 4 types of metallic sensing electrodes? (L17 s6)
> 1st kind → its own cation (e.g. Cu²⁺ + 2e⁻ ⇌ Cu)
> 2nd kind → ion that forms a precipitate / complex with the metal (e.g. AgCl + e⁻ ⇌ Ag + Cl⁻ → senses Cl⁻)
> 3rd kind → competing cation in a complex / solubility equilibrium (e.g. Hg(EDTA)²⁻ + Ca²⁺ + 2e⁻ ⇌ Hg + Ca(EDTA)²⁻)
> Inert metal (Pt) → redox equilibrium in solution (e.g. Fe(CN)₆³⁻ / Fe(CN)₆⁴⁻)
∴ Type = what the electrode responds to
Electrode: How does a pH glass membrane work, and what else is needed to measure pH? (L17
s7–9, Rev)
> Ionically conductive glass membrane → known [H⁺] inside (filling), unknown [H⁺] outside (sample)
> Gl–O⁻ + H⁺(aq) ⇌ Gl–OH on both surfaces → different [H⁺] → membrane potential (E_b)
> E_b = (RT/zF) ln(a₁/a₂) → E_b = 0.0592 V − 0.0592 V · pH (inside a = 0.1)
> Also needed: 2 reference electrodes (stable, known E), one on each side
> Measured potentiometrically at zero current
∴ H⁺ binding on glass → membrane potential → pH
Electrolyte: Why must the electrolyte transport both charge and mass? (L16 s11)
> Charge: e⁻ flow anode → cathode through the wire → (+) charge builds up at the anode, (−) at the cathode
> Ions move to cancel the buildup (anions → anode, cations → cathode) → circuit closed
> Mass: reaction uses up X at the electrode surface
> Fresh X brought in + products moved away → reaction keeps going
> No transport → charge piles up / surface runs out of X → current stops
∴ Charge transport closes the circuit, mass transport feeds the reaction
Electrolyte: How does the electrolyte transport charge and mass? (L16 s11)
> J = −D∇c − (zF/RT)·D·c·∇φ + ∇(c·u)
> Diffusion → concentration gradient (∇c) → high c → low c
> Migration → potential gradient (∇φ) → only charged X (z ≠ 0)
> Convection → bulk flow of solution (u) (e.g. stirring)
∴ Diffusion + migration + convection = mass transport
Electrolyte: Why do we add a supporting electrolyte? (L16 s14)
> Supporting electrolyte = high c of non-reacting salt (e.g. KCl) → only carries charge
> Few ions → ↑ ionic resistance → potential (∇φ) lost in the solution
> ↑ ions → ↑ charge carriers → ↓ resistance → ↓ potential lost
> J = −z·u·c·∇φ → ↑ c → same J with a smaller ∇φ
∴ Supporting electrolyte ↓ resistance → ↓ potential lost
Electrolyte: Why is KCl the most common electrolyte salt, and what if you used NaCl? (L17 s5, Rev)
> Junction = 2 electrolytes of different c meet → ions diffuse at different speeds → charge separation → junction potential (adds error to Ecell)
> K⁺ and Cl⁻ have very similar diffusivities → minimal junction potential
> NaCl → Na⁺ ≠ Cl⁻ diffusivity → bigger junction potential → must be measured + adjusted for
∴ KCl = similar diffusivities = minimal junction potential

Method: What is the measurement principle of potentiometry, and what are its components? (L16 s16, s19)
> Current-less (zero current) measurement of electrode potential vs a reference → quantitative c of redox species (via Nernst)
> Sensing / indicator electrode = cathode → its E depends on X
> Reference electrode = anode → stable, known E
> Electrolyte between them + voltmeter to read E
∴ Potentiometry = E of the sensing electrode vs the reference at zero current → c of X
Method: Why must potentiometric measurements be current-less? (L16 s17)
> Current flows → system no longer at equilibrium → measured E changes
> e.g. ↑ potential → pulls e⁻ away from the electrode → Fe → Fe²⁺ + 2e⁻ runs → current flows to re-equilibrate
> Nernst only holds at equilibrium → E only reflects c of X at zero current
∴ Current disturbs equilibrium → wrong E → measure at zero current
Method: What steps does X go through to react at an electrode? (L17 s17)
> Mass transfer → bulk to surface (diffusion, migration, convection)
> Chemical reaction → X changes form if needed (Ox → Ox′)
> Physical state change → e.g. adsorption on the electrode
> Electron transfer → Ox + ne⁻ → Red
> Red leaves by the same steps in reverse
∴ Bulk → surface → electron transfer → back out
Method: How does a potentiostat work, and what does each electrode do? (L17 s21)
> WE → where the reaction of interest happens
> RE → stable reference; potentiostat measures E between WE and RE
> CE → completes the circuit; potentiostat drives + measures current between WE and CE
> Adjusts WE–CE power until WE/RE potential = the set value
> No current through the RE → RE potential stays stable
∴ E measured vs RE, current flows WE ⇌ CE

Method: What is the difference between Faradaic and non-Faradaic current? (L17 s22)
> Faradaic → charge transfer across the electrode interface → redox reaction (e.g. Fe → Fe²⁺ + 2e⁻) → the signal
> Non-Faradaic → polarization of the interface, no charge transfer → double layer charges like a capacitor → background
∴ Faradaic = e⁻ cross (reaction); non-Faradaic = interface charges (no reaction)
Method: How does cyclic voltammetry work and what do you read from it? (L17 s28–31)
> Apply E swept linearly forward + back → measure current
> ↑ E → electrode more (+) → electrode pulls e⁻ from X → X loses e⁻ → X oxidized (electrode = e⁻ taker)
> ↓ E → electrode more (−) → electrode pushes e⁻ into X → X gains e⁻ → X reduced
> Oxidation peak (ipa, Epa) on the forward sweep + reduction peak (ipc, Epc) on the back sweep
> E½ → identifies the redox species; ip → concentration
> Reversible → fast kinetics, peaks fixed vs scan rate, ΔEp = 59 mV / z (25 °C)
> Irreversible → slow kinetics, peaks shift outward at ↑ scan rate, ΔEp > 59 mV / z
∴ E½ = which X, ip = how much X, ΔEp = how fast

Method: What is anodic stripping voltammetry (ASV) and when is it used?
> For trace metals (10⁻⁹–10⁻¹⁰ M)
> Step 1: hold cathodic E for a set time → metals deposit on WE (preconcentration ↑)
> Step 2: scan E positive → deposited metals oxidize (strip) → measure i
> Each metal strips at its own E → Pb + Cd resolved; charge (area) ∝ amount
∴ Trace Pb/Cd in soil → ASV: deposition = ↑ sensitivity + leaves the matrix behind
Equation: How do you balance a redox half-reaction in acidic solution? (L16 s4 + gen chem)
> Balance all atoms except O + H
> Balance O → add H₂O
> Balance H → add H⁺
> Balance charge → add e⁻ to the more positive side
> Full reaction → multiply half-reactions so e⁻ match → add → cancel
> e.g. 2MnO₂ + 2H⁺ + 2e⁻ → Mn₂O₃ + H₂O
> No condition given → assume acidic
∴ atoms → O (H₂O) → H (H⁺) → charge (e⁻) → check atoms + charge
Equation: How do you balance a redox half-reaction in basic solution? (L16 s4 + gen chem)
> Balance as if acidic first
> Add OH⁻ to both sides = # of H⁺
> H⁺ + OH⁻ → H₂O
> Cancel the H₂O on both sides
> e.g. 2MnO₂ + H₂O + 2e⁻ → Mn₂O₃ + 2OH⁻
> Basic → Q says basic/alkaline or gives OH⁻ (NaOH, KOH)
∴ basic = acidic method + OH⁻ to kill the H⁺ + cancel the extra H₂O
Equation: How do you balance equations, generally?
> Put a letter in front of every species
> Write one equation per element (left count = right count)
> For redox, ↑ a charge equation too
> Set one letter = 1 → solve the rest → scale up to whole numbers
Equation: How do you get an electrode potential at non-standard conditions? (L16 s8)
> Nernst: E = E° − (RT/zF) · ln(∏ aredνred / ∏ aoxνox)
> z = number of e⁻ exchanged per reaction; F = 96,485 C mol⁻¹; R = 8.315 J K⁻¹ mol⁻¹; T in K
> ↑ ared → ↑ ratio → ↓ E, ↑ a_ox → ↑ E
> All a = 1 → ln 1 = 0 → E = E°
∴ Nernst corrects E° for the real activities
Equation: How is the cell's ΔE related to the free energy of the reaction? (L16 s10)
> ΔG = −z · F · ΔE (ΔE measured at zero current, reversible)
> ΔE = driving force of the reaction → same thing as ΔG, just written as a voltage
> ΔE + → ΔG − → spontaneous (galvanic)
> ΔE − → ΔG + → not spontaneous
> +E electrode = takes e⁻ (reduced, cathode); lower E = gives e⁻ (oxidized, anode)
∴ +ΔE = −ΔG = spontaneous
Equation: How do ions migrate, and how does migration give current? (L16 s13)
> J = −z · u · c · ∇φ (u = ion mobility, ∇φ = electric field gradient)
> Anode → (+) charges formed / (-) consumed; cathode → (-) charges formed / (+) consumed
> Cations (supporting) → cathode, anions (supporting) → anode → cancels the charge buildup
> i = J · z · F → ↑ J = ↑ current density
∴ Migration = ions pulled by the field → carries the current
*Equation: What does the Butler–Volmer equation describe? (L17 s18)
> i = i₀ · [e(αa zF(E − Eeq)/RT) − e(αc zF(Eeq − E)/RT)]
> 1st term = oxidation (E > Eeq), 2nd = reduction (E < Eeq)
> E = Eeq → i = 0 (equilibrium, no net current)
> i₀ = exchange current → ↑ i₀ = faster e⁻ transfer kinetics (between electrode and X @ its surface)
∴ Current ↑ exponentially as E moves away from Eeq
![<p>> <strong>i = i₀ · [e<sup>(α</sup><sub><sup>a</sup></sub><sup> zF(E − E</sup><sub><sup>eq</sup></sub><sup>)/RT)</sup> − e<sup>(α</sup><sub><sup>c</sup></sub><sup> zF(E</sup><sub><sup>eq</sup></sub><sup> − E)/RT)]</sup></strong></p><p>> <strong>1st term = oxidation (E > E<sub>eq</sub>), 2nd = reduction (E < E<sub>eq</sub>)</strong></p><p>> <strong>E = E<sub>eq</sub> → i = 0</strong> (equilibrium, no net current)</p><p>> <strong>i₀ = exchange current → ↑ i₀ = faster e⁻ transfer kinetics </strong>(between electrode and X @ its surface)</p><p>∴ <strong>Current ↑ exponentially as E moves away from E<sub>eq</sub></strong></p>](https://assets.knowt.com/user-attachments/92fd8f12-272c-46a7-956e-13837f972fda.png)
Equation: How do you get a diffusion coefficient (D) from voltammetry? (L17 s33–34, Final Q18)
> Randles–Sevcik: ip = 0.4463 · zFAc · √(zFvD / RT) (v = scan rate, A = area, c = concentration)
> Run CV at several scan rates → plot ip vs √v → straight line
> Slope → solve for D
> Tafel (s32): ln i = ln i₀ + (αa zF / RT) η → intercept = ln i₀ (kinetics)
∴ Vary scan rate → ip vs √v → D
Signal: What is S/N and how do you estimate it from data or a plot? (L3 s6–7)
> S/N = mean / standard deviation = 1 / RSD
> Discrete data → s = √(Σ(xᵢ − x̄)² / (n − 1))
> Plot → s ≈ (noise max − noise min) / 5
> ↑ S/N → ↑ signal (mean) or ↓ noise (s)
∴ S/N = mean / s; plot noise span ≈ 5s
Noise: How does ensemble averaging improve S/N, and how many scans do you need? (L3 s12, Rev)
> Average n repeat measurements → S/N ↑ by √n
> n = ((S/N)₂ / (S/N)₁)² (e.g. 5 → 25 needs n = 25)
∴ S/N ∝ √n
Noise: What causes noise, and which hardware method fixes each? (L3 s8–11, s14, Rev)
> Instrumental: thermal → ↓ T, ↓ resistance; shot → fewer junctions; flicker → good connections
> Environmental (power lines, EM) → grounding + Faraday cage (grounded metal housing → induced currents drain to ground)
> Sample: vibrations → isolation table; T / humidity → environmental chamber
∴ Match the noise source to its fix
Circuit: How do voltage and current dividers work? (L3 s21–24, L4 s10–13, Rev)
> Series (voltage divider) → same I through all, V splits: (Vin − V2) / R1 = V2 / R2
> Parallel (current divider) → same V across all, I splits: 1/R = 1/R1 + 1/R2, I = V / R per branch
> Always V = I · R
∴ Series splits V, parallel splits I
Circuit: How do you minimize loading error from meters? (L4 s7, s14, Final Q2)
> Ammeter → in series → needs LOW impedance
> Voltmeter → in parallel → needs HIGH impedance
> Meters draw or block current → they change the circuit they measure
∴ Ammeter low Z in series, voltmeter high Z in parallel
Signal: How do you find ADC resolution or the bits needed? (L4 s23, s25, Final Q3)
> 2ⁿ bins (n = bits) → resolution = range / 2ⁿ
> e.g. 0–6 V, 8-bit → 6 / 256 = 23 mV
> Bits needed → smallest n with 2ⁿ ≥ range / resolution (0–12 V at 0.01 V → 1200 → n = 11)
∴ ↑ bits → smaller bins → better resolution
Calibration: Blank vs dummy vs standard vs control vs internal standard vs standard addition? (L2 s12–20)
> Blank → sample matrix without X (e.g. solvent in cuvette)
> Dummy → instrument run with nothing in it (e.g. empty IR holder)
> Standard → artificial X to calibrate; control → artificial sample close to real
> Internal standard → different signal, added equally to all → corrects stimulus / intake / dilution fluctuations
> Standard addition → same X added to the sample → extrapolate to 0 signal
∴ All of them reduce systematic (non-random) errors