1. What is the primary purpose of dialysis?
A. Treat liver failure
B. Filter blood externally for renal failure
C. Increase cardiac output
D. Reduce blood pressure
B. Filter blood externally for renal failure
2. Dialysis access requires which of the following hemodynamic conditions?
A. Low flow, high resistance
B. Low flow, low resistance
C. High flow, low resistance
D. High flow, high resistance
C. High flow, low resistance
3. A synthetic dialysis graft is most commonly made of which material?
A. Dacron
B. Silicone
C. PTFE (Gore-Tex)
D. Polyethylene
C. PTFE (Gore-Tex)
4. A loop graft differs from a straight graft primarily by its:
A. Diameter
B. Flow direction
C. Configuration
D. Anastomosis type
C. Configuration
5. A native autogenous fistula is best described as:
A. Artery to synthetic graft
B. Vein to synthetic graft
C. Artery directly connected to vein
D. Vein connected to vein
C. Artery directly connected to vein
6. A Brescia-Cimino fistula connects which vessels?
A. Brachial artery to basilic vein
B. Radial artery to cephalic vein
C. Ulnar artery to cephalic vein
D. Brachial artery to cephalic vein
B. Radial artery to cephalic vein
7. Which flow pattern is normal in a functioning dialysis access?
A. High resistance, low velocity
B. Triphasic
C. Low resistance, high velocity
D. No diastolic flow
C. Low resistance, high velocity
8. A bruit or thrill in a dialysis access is considered:
A. Abnormal
B. A sign of infection
C. Normal
D. A sign of occlusion
C. Normal
9. The most common location for stenosis in a dialysis access is the:
A. Inflow artery
B. Arterial anastomosis
C. Venous anastomosis / outflow vein
D. Distal artery
C. Venous anastomosis / outflow vein
10. Which transducer frequency is typically used for dialysis access evaluation?
A. 1-3 MHz
B. 2-4 MHz
C. 4-10 MHz
D. 10-15 MHz
C. 4-10 MHz
11. When evaluating a graft, which segments should be assessed?
A. Only the arterial side
B. Entire graft length only
C. Inflow artery, anastomoses, graft, and outflow vein
D. Only areas with symptoms
C. Inflow artery, anastomoses, graft, and outflow vein
12. Volume flow calculation requires which velocity measurement?
A. PSV
B. EDV
C. TAMV
D. Peak mean velocity
C. TAMV
13. Proper volume flow measurement requires the sample volume to:
A. Be placed near the vessel wall
B. Sample part of the lumen
C. Encompass the entire vessel lumen
D. Be placed distal to stenosis
C. Encompass the entire vessel lumen
14. Accepted normal volume flow in a functioning access is generally:
A. >300 mL/min
B. >500 mL/min
C. >800 mL/min
D. >1200 mL/min
C. >800 mL/min
15. A volume flow <500 mL/min suggests:
A. Normal access
B. High-output failure
C. Failing access
D. Infection
C. Failing access
16. General velocity criteria for >50% stenosis at the anastomosis include:
A. PSV ≥200 cm/s B. PSV ≥300 cm/s C. PSV ≥400 cm/s D. PSV ≥500 cm/s
17. A velocity ratio >3 at the anastomosis suggests:
A. Normal flow
B. Mild stenosis
C. Significant stenosis
D. Occlusion
C. Significant stenosis
18. Along the venous outflow, >50% stenosis is suggested by PSV ≥:
A. 150 cm/s
B. 200 cm/s
C. 250 cm/s
D. 300 cm/s
D. 300 cm/s
19. Which waveform suggests inflow artery disease?
A. High velocity, low resistance
B. Triphasic
C. Dampened waveform with decreased systolic upstroke
D. Turbulent waveform
C. Dampened waveform with decreased systolic upstroke
20. Which finding is most consistent with outflow stenosis?
A. Low velocity, high resistance
B. High resistance waveform
C. Decreased diastolic flow
D. High velocity with low resistance
B. High resistance waveform
21. Volume flow measurements are best used to:
A. Diagnose stenosis
B. Replace velocity measurements
C. Follow access performance over time
D. Identify occlusion location
C. Follow access performance over time
22. Fistula occlusion is most easily identified by:
A. Elevated velocities
B. No flow despite imaging the vessel
C. Increased diastolic flow
D. Triphasic waveform
B. No flow despite imaging the vessel
23. In fistula occlusion, which vessel often remains patent?
A. Outflow vein
B. Graft
C. Artery
D. Pseudoaneurysm
C. Artery
24. An extravascular anechoic mass with no flow may represent all EXCEPT:
A. Seroma
B. Hematoma
C. Thrombosed pseudoaneurysm
D. Active pseudoaneurysm
D. Active pseudoaneurysm
25. A pseudoaneurysm is defined as:
A. Dilatation of the vessel wall
B. Flow within all vessel layers
C. Active blood flow outside the vessel
D. Thrombosed access
C. Active blood flow outside the vessel
26. The classic Doppler finding of a pseudoaneurysm neck is:
A. Triphasic flow
B. High resistance flow
C. To-and-fro flow
D. Continuous flow
C. To-and-fro flow
27. Which factor may preclude thrombin injection in a pseudoaneurysm?
A. Large sac size
B. Small neck size
C. Large neck / defect
D. Low flow
C. Large neck / defect
28. Treatment options for pseudoaneurysm include all EXCEPT:
A. Surveillance
B. Surgical repair
C. Endovascular stent graft
D. Thrombolysis
D. Thrombolysis
29. Access aneurysm formation is expected because of:
A. Infection
B. Low pressure
C. High venous pressure
D. Low flow
C. High venous pressure
30. Steal syndrome occurs when:
A. Blood flows normally through the fistula
B. Blood is redirected away from the hand
C. Venous pressure decreases
D. Arterial flow increases distally
B. Blood is redirected away from the hand
31. Symptoms of steal syndrome may include all EXCEPT:
A. Pain
B. Pallor
C. Ulceration
D. Hypertension
D. Hypertension
32. Retrograde arterial flow distal to the fistula suggests:
A. Normal physiology
B. Access occlusion
C. Steal physiology
D. Infection
C. Steal physiology
33. The preferred method to assess steal syndrome is:
A. Duplex only
B. CTA
C. PPG tracings with compression
D. ABI
C. PPG tracings with compression
34. If PPG tracings improve when the fistula is compressed, this indicates:
A. No steal
B. Normal circulation
C. Steal syndrome
D. Venous occlusion
C. Steal syndrome
35. If PPG tracings do NOT change with fistula compression, this suggests:
A. Steal syndrome
B. No steal
C. Severe occlusion
D. Pseudoaneurysm
B. No steal
36. Which pathology is associated with repeated cannulation trauma?
A. Stenosis
B. Pseudoaneurysm
C. Scarring
D. All of the above
D. All of the above
37. High-output cardiac failure in dialysis patients may result from:
A. Low access flow
B. Steal syndrome
C. Excessive access flow volume
D. Occlusion
C. Excessive access flow volume
38. The most reliable comparison in dialysis access evaluation is:
A. Single exam values
B. PSV alone
C. Serial follow-up studies
D. EDV measurements
C. Serial follow-up studies
39. Which finding best suggests access patency?
A. Triphasic waveform
B. Continuous low resistance flow
C. No diastolic flow
D. High resistance flow
B. Continuous low resistance flow
40. Velocity measurements are primarily used to:
A. Assess access performance
B. Diagnose stenosis
C. Calculate volume flow
D. Replace volume measurements
B. Diagnose stenosis