radiation detector requires a _____________
medium
a radiation detector converts?
energy into a signal
a transducer converts?
one form of energy into another
radiation causes ______________ or _____________
excitation or ionization
ionizations create _________ ____________
ion pairs
an ion pair is a free ______________ and _________ __________
free electron and positive atom
gas filled detector contains __________ between two _____________
gas; electrodes
incident radiation creates ________ _________
ion pairs
electrons go towards the ___________
anode
positive atoms go towards the _____________
cathode
electronics can measure the amount of _____________ as either _____________ or _____________
electrons as current or voltage
current is ______________ mode
current
voltage is ______________ mode
pulse
dose calibrators operate in ___________ mode
current
ionization chambers operate in ___________ mode
current
GM survey meters operate in ____________ mode
pulse
the number of electrons measured depends on the number of ________, the ____________ of radiation, _______________ of detector
photos, energy, geometry
the ____________ is the driving force of the movement of electrons through the gas
voltage
the ____________ determines the number of ion pairs that reach the anode and cathode
magnitude
what measures the signal vs. applied voltage at a given level of incident radiation?
voltage response curve

recombination region

saturation region

proportional region

GM region

spontaneous discharge
in the recombination region ___________ _____________ bc voltage is so low
ions recombine
in the saturation region, voltage is ____________ enough to collect all charges produced in the detector, but increasing the voltage _____________ give more signal
sufficient, does not
in the proportional region, electrons are __________, which means more _____________! increasing the voltage ________________ give more signal
accelerated, energy, does!
in the GM region, gas amplification is at _____________, _____ ionization is all it takes, the signal remains ____________
maximum, one, constant
in the spontaneous discharge region, the voltage is super __________, this is not useful for?
high, radiation detection
current mode is ______________ ________
continuous flow
current mode is a ___________ ___________ of ______ ___________
constant average
current mode cannot detect ___________ _____________
individual ionizations
pulse mode is a signal seen as a series of _____________ corresponding to interactions
pulses
voltage can be measured across a ____________ and a ____________ circuit which alters or __________ the pulses
resistor, capacitor, shapes

short time constant

fast time constant
dose calibrators operate in the _______________ region
saturation
dose calibrators measure ______________
activity
dose calibrators are filled with pressurized __________ or _____________ gas
air, argon gas
dose calibrators are under __________ pressure
high
ionization chambers measure __________ ____________
exposure rate
ionization chambers are not as _____________
sensitive
ionization chambers are operated in the _____________ region
saturation
ionization chambers are filled with __________
air
ionization chambers are used for _____________ radiation levels in ______________
monitoring, patients
GM suvery meters detects ___________
contamination
GM survey meters operate in the ____________ region
GM
in GM meters, each radiation events produces a _____________ of ions
avalanche
the avalanche stops when ___________ ions slowly reach the ____________, which decreases the _____________ field
positive, cathode, electric
GM counters have ______________!
deadtime
GM counters use _____________ gas
quenching
quenching gas is used to stop UV protons produced from recombination which would restart the ______________
avalanche
what is quenching gas?
organic or halogen gas
GM meters are filled with ____________ or ____________ gas
helium or argon
PRO of gas filled detectors
reliable, easy to measure signals
CON of gas filled detectors
empty space, low efficiency for high E photos, cannot tell indv energies
geometry is perfomed ______________
initially
geometry meausres the activity in a _________ and ____________
vial and syringe
constancy is perfomed _____________
daily
constancy uses _____________________
Cs-137
linearity is perfomed _______________
quarterly
linearity ensures proper measurement over a …
wide range of activities
linearity uses ___________
Tc-99m
accuracy is perfomed ____________
yearly
accuracy uses a ________ _________ soruce
long lived