quality assurance (QA)
general management program, scheduling, image analysis outcomes, continuous quality improvement
quality control (QC)
measurement & testing, ensuring optimal images with minimal dose
before a new radiography system is used
when is acceptance testing performed
medical physicist
who performs acceptance testing
verify that hardware meets manufacturer specs & federal safety standards
why is acceptance testing done
periodically
routine monitoring is performed _________ (daily, weekly, monthly, or quarterly)
technologist
who performs routine monitoring
catch immediate or gradual shifts in equipment performance over time
why is routine monitoring done
proactively
preventative maintenance (PM) is scheduled _________ on an annual or semi-annual basis
biomedical service engineers
preventative maintenance (PM) is handled by specialized __________
1 mGy/hour at 1 meter
leakage radiation from tube housing must be limited to _____
control booth
is a secondary barrier that permits observation of patient during exposure
control console
must indicate when tube is energized, make audible sound, and have an illuminated radiation sign during exposure; have visible mAs & kVp readouts
300-400 lbs
radiographic tables must be able to hold how much weight
positive beam limitation (PBL)
automatic collimation that restricts maximum size of beam to IR size
15” (38cm)
SSD (source to skin distance) for fixed diag/fluoro equipment
12” (30cm)
SSD (source to skin distance) for portable equipment
161 LUX
collimator light source must be what luminance
171 LUX
monitors must be what luminance
total tube filtration
manufactures must build x-ray tubes with a minimum amount of total filtration by law. Most important patient protection device for ESE
annually
how often is tube filtration tested
total filtration = inherent + added
total filtration equation
inherent filtration
glass envelope, tube window, oil, mirror in collimator
added filtration
Al or Cu sheets
2.5mm Al equivalent
min. filtration for >70kVp
1.5mm Al eq
min. filtration 51-70 kVp
0.5 mm Al equiv.
min. filtration for < 50 kVp
half value layer (HVL)
thickness of Al that reduces a beam’s intensity to 50%; a measurement of beam quality at a given kVp to ensure that the inherent filtration hasn’t degraded
kVp, generator type, & tube age
HVL changes with _________
exposure field/Light field congruence
x-ray field light must coincide (be congruent) with the light field within 2% of SID and should be tested semiannually
2% of SID
exposure field/ light field congruence must be within what percent
2% of SID
SID indicator (laser or tape) must accurately measure distance from focal spot to the IR within what percent _________
1° of perpendicularity; 2% of SID
beam alignment of CR (perpendicularity) must be within ______ (in inner circle) OR _______ of SID
effective focal spot size
size stated by the manufacturer for the size of the beam origin from a pt’s perspective due to the actual angle of the anode
50%
the EFSS can vary by ______ of the manufacturer’s stated size
actual focal spot size
size of the area on the anode where the electron streams hits the anode
annually
EFSS should be evaluated _______ by physicist and when tube is replaced
pinhole camera
used for EFSS; difficult to use and requires excessive exposure time (for Spatial Resolution)
star pattern
for EFSS; easy to use but limited for FSS less than 0.3mm (Spatial Resolution)
slit camera
for EFSS; **Gold Standard; requires a complex geometric alignment (usually placed 1/3 down from the tube to ensure magnification is enough to see the line (closer it is to the tube, the more magnified) (Spatial Resolution)
½ way btw source & IR
where is pinhole tool placed to get an image the size of the focal spot
½ way btw source & IR - UNLESS too small to read → closer to source to magnify
where is star tool placed
closer to source to resolve the narrow profile of the focal spot
where is tool slit placed
line pair test tool
compares the 3 images at different EFSS (Spatial Resolution)
5%
kVp accuracy needs to be w/ in what percent
oscilloscopes
electronic _______ are the modern way to determine kVp accuracy
annually
how often do exposure timers need to be checked
5%
exposure timer needs to be within what percent
spinning top test
old, but evaluates knowledge of generators & rectification; catches the pulses as individual dots on an image during an exposure (Exposure Timer)
half wave rectified
60 pulses/s, dots = time (s) x 60°
full wave rectified
120 pulses/s, dots = time(s) x 120°
synchronous top
used for 3 phase, high frequency generators that creates a solid arc representing time
exposing IR thru varying thicknesses of Al or acrylic phantoms
how is AEC tested
5%
AEC must be within what percent of baseline output intensity
inserting a heavy lead filter/block
backup timer is tested by
6 sec or 600 mAs
backup timer should terminate the exposure at
10%
exposure linearity must be within _____ for adjacent mA stations
annually
when is exposure linearity tested
5%
exposure & AEC reproducibility: when the same technique (kVp/mAs) is used (but returned to original settings btw exposures) mGy should be the same = they are reproducible within _____ of each other