quality assurance (QA)

general management program, scheduling, image analysis outcomes, continuous quality improvement

2
New cards

quality control (QC)

measurement & testing, ensuring optimal images with minimal dose

3
New cards

before a new radiography system is used

when is acceptance testing performed

4
New cards

medical physicist

who performs acceptance testing

5
New cards

verify that hardware meets manufacturer specs & federal safety standards

why is acceptance testing done

6
New cards

periodically

routine monitoring is performed _________ (daily, weekly, monthly, or quarterly)

7
New cards

technologist

who performs routine monitoring

8
New cards

catch immediate or gradual shifts in equipment performance over time

why is routine monitoring done

9
New cards

proactively

preventative maintenance (PM) is scheduled _________ on an annual or semi-annual basis

10
New cards

biomedical service engineers

preventative maintenance (PM) is handled by specialized __________

11
New cards

1 mGy/hour at 1 meter

leakage radiation from tube housing must be limited to _____

12
New cards

control booth

is a secondary barrier that permits observation of patient during exposure

13
New cards

control console

must indicate when tube is energized, make audible sound, and have an illuminated radiation sign during exposure; have visible mAs & kVp readouts

14
New cards

300-400 lbs

radiographic tables must be able to hold how much weight

15
New cards

positive beam limitation (PBL)

automatic collimation that restricts maximum size of beam to IR size

16
New cards

15” (38cm)

SSD (source to skin distance) for fixed diag/fluoro equipment

17
New cards

12” (30cm)

SSD (source to skin distance) for portable equipment

18
New cards

161 LUX

collimator light source must be what luminance

19
New cards

171 LUX

monitors must be what luminance

20
New cards

total tube filtration

manufactures must build x-ray tubes with a minimum amount of total filtration by law. Most important patient protection device for ESE

21
New cards

annually

how often is tube filtration tested

22
New cards

total filtration = inherent + added

total filtration equation

23
New cards

inherent filtration

glass envelope, tube window, oil, mirror in collimator

24
New cards

added filtration

Al or Cu sheets

25
New cards

2.5mm Al equivalent

min. filtration for >70kVp

26
New cards

1.5mm Al eq

min. filtration 51-70 kVp

27
New cards

0.5 mm Al equiv.

min. filtration for < 50 kVp

28
New cards

half value layer (HVL)

thickness of Al that reduces a beam’s intensity to 50%; a measurement of beam quality at a given kVp to ensure that the inherent filtration hasn’t degraded

29
New cards

kVp, generator type, & tube age

HVL changes with _________

30
New cards

exposure field/Light field congruence

x-ray field light must coincide (be congruent) with the light field within 2% of SID and should be tested semiannually

31
New cards

2% of SID

exposure field/ light field congruence must be within what percent

32
New cards

2% of SID

SID indicator (laser or tape) must accurately measure distance from focal spot to the IR within what percent _________

33
New cards

1° of perpendicularity; 2% of SID

beam alignment of CR (perpendicularity) must be within ______ (in inner circle) OR _______ of SID

34
New cards

effective focal spot size

size stated by the manufacturer for the size of the beam origin from a pt’s perspective due to the actual angle of the anode

35
New cards

50%

the EFSS can vary by ______ of the manufacturer’s stated size

36
New cards

actual focal spot size

size of the area on the anode where the electron streams hits the anode

37
New cards

annually

EFSS should be evaluated _______ by physicist and when tube is replaced

38
New cards

pinhole camera

used for EFSS; difficult to use and requires excessive exposure time (for Spatial Resolution)

39
New cards

star pattern

for EFSS; easy to use but limited for FSS less than 0.3mm (Spatial Resolution)

40
New cards

slit camera

for EFSS; **Gold Standard; requires a complex geometric alignment (usually placed 1/3 down from the tube to ensure magnification is enough to see the line (closer it is to the tube, the more magnified) (Spatial Resolution)

41
New cards

½ way btw source & IR

where is pinhole tool placed to get an image the size of the focal spot

42
New cards

½ way btw source & IR - UNLESS too small to read → closer to source to magnify

where is star tool placed

43
New cards

closer to source to resolve the narrow profile of the focal spot

where is tool slit placed

44
New cards

line pair test tool

compares the 3 images at different EFSS (Spatial Resolution)

45
New cards

5%

kVp accuracy needs to be w/ in what percent

46
New cards

oscilloscopes

electronic _______ are the modern way to determine kVp accuracy

47
New cards

annually

how often do exposure timers need to be checked

48
New cards

5%

exposure timer needs to be within what percent

49
New cards

spinning top test

old, but evaluates knowledge of generators & rectification; catches the pulses as individual dots on an image during an exposure (Exposure Timer)

50
New cards

half wave rectified

60 pulses/s, dots = time (s) x 60°

51
New cards

full wave rectified

120 pulses/s, dots = time(s) x 120°

52
New cards

synchronous top

used for 3 phase, high frequency generators that creates a solid arc representing time

53
New cards

exposing IR thru varying thicknesses of Al or acrylic phantoms

how is AEC tested

54
New cards

5%

AEC must be within what percent of baseline output intensity

55
New cards

inserting a heavy lead filter/block

backup timer is tested by

56
New cards

6 sec or 600 mAs

backup timer should terminate the exposure at

57
New cards

10%

exposure linearity must be within _____ for adjacent mA stations

58
New cards

annually

when is exposure linearity tested

59
New cards

5%

exposure & AEC reproducibility: when the same technique (kVp/mAs) is used (but returned to original settings btw exposures) mGy should be the same = they are reproducible within _____ of each other