Experiment 2 - Amino Acid Buffer Capacity and Identification

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/15

encourage image

There's no tags or description

Looks like no tags are added yet.

16 Terms

1
New cards

equivalence volume

volume of titrant needed to reach the end-point of a titration

  • point of inflection on titration curve


2
New cards

how to go from equivalence volume to identity of the unknown amino acid?

  1. (equivalence volume) x (titrant molarity) = moles of titrant

  2. (measured mass of unknown amino acid) / (moles of titrant) = molar mass of unknown amino acid


3
New cards

spectrophotometer

measures absorbance (or transmission) of light in both visible and/or UV regions of light spectrum

4
New cards

what does a spectrophotometer actually measure first? (and how is it converted to absorbance?)

measures transmitted light intensity (I) relative to incident light (IO)

  • converted to absorbance using A = log (IO / I)


5
New cards

why does absorbance lose its linear relationship with concentration at high concentrations?

because of instrument limitations and fundamental physics at high concentration

6
New cards

why are absorbance readings that are > 2 considered unreliable?

because A = 2 means only 1% of light reaches the detector

  • too little for reliable measurement


7
New cards

Beer-Lambert Law equation (and variables)

A = εcl

  • A = absorbance

  • ε = molar absorptivity / extinction coefficient (M^-1 cm^-1)

  • l = pathlength (cm)

  • c = analyte concentration


  • does not work at ALL concentrations, especially when at high sample concentrations

    • because of instrument limitattions


8
New cards

why do biochemists usually use a standard curve rather than calculating concentration directly by rearranging Beer-Lambert Law equation (c = A / (εl))

ε may be unknown or vary under assay conditions

  • and other molecules in the mixture can contribute to absorbance


9
New cards

when blanking a spectrophotometer and measuring the reading, where should the best fit line go through?

through the origin

10
New cards

what is the difference between 1/10 dilution and 1:10 ratio?

  • 1/10 dilution - stock is diluted tenfold to 1/10 its original concentration

    • ex. 1mL diluted to 10mL total volume

  • 1:10 ratio - compares parts of each component

    • ex. 1mL stock to 90mL water


11
New cards

serial dilution

making sequences of dilutions

  • diluting the stock, then diluting that diluted solution, and so on


12
New cards

when should you use a serial dilution instead of a single step dilution?

  • when a single-step calculation needs a volume too tiny to measure accurately

  • when increasing final volume isn’t practical

  • when stock compound is rare/expensive


13
New cards

why must cuvettes be clean, unscratched, and free of debris, condensation, or fingerprints

because anything that blocks or scatters light can reduce transmitted/absorbed light and cause an error in measurements.

14
New cards

3 approaches to solving dilution problems

  1. C1V1 = C2V2

  2. dimensional analysis

  3. serial dilution


15
New cards

is it better to use interpolation (estimating data between data points) or using the slope of a best-fit line

slope of a best-fit line since it smooths out random measurement error across multiple data points

  • also because slope (m) = εl under the conditions of the experiment


16
New cards

Based on our results from Experiment 2, Which is more accurate for reading absorbance, the spectrophotometer or the plate reader?

the plate reader showed higher % accuracy BASED ON OUR RESULTS