equivalence volume
volume of titrant needed to reach the end-point of a titration
point of inflection on titration curve
how to go from equivalence volume to identity of the unknown amino acid?
(equivalence volume) x (titrant molarity) = moles of titrant
(measured mass of unknown amino acid) / (moles of titrant) = molar mass of unknown amino acid
spectrophotometer
measures absorbance (or transmission) of light in both visible and/or UV regions of light spectrum
what does a spectrophotometer actually measure first? (and how is it converted to absorbance?)
measures transmitted light intensity (I) relative to incident light (IO)
converted to absorbance using A = log (IO / I)
why does absorbance lose its linear relationship with concentration at high concentrations?
because of instrument limitations and fundamental physics at high concentration
why are absorbance readings that are > 2 considered unreliable?
because A = 2 means only 1% of light reaches the detector
too little for reliable measurement
Beer-Lambert Law equation (and variables)
A = εcl
A = absorbance
ε = molar absorptivity / extinction coefficient (M^-1 cm^-1)
l = pathlength (cm)
c = analyte concentration
does not work at ALL concentrations, especially when at high sample concentrations
because of instrument limitattions
why do biochemists usually use a standard curve rather than calculating concentration directly by rearranging Beer-Lambert Law equation (c = A / (εl))
ε may be unknown or vary under assay conditions
and other molecules in the mixture can contribute to absorbance
when blanking a spectrophotometer and measuring the reading, where should the best fit line go through?
through the origin
what is the difference between 1/10 dilution and 1:10 ratio?
1/10 dilution - stock is diluted tenfold to 1/10 its original concentration
ex. 1mL diluted to 10mL total volume
1:10 ratio - compares parts of each component
ex. 1mL stock to 90mL water
serial dilution
making sequences of dilutions
diluting the stock, then diluting that diluted solution, and so on
when should you use a serial dilution instead of a single step dilution?
when a single-step calculation needs a volume too tiny to measure accurately
when increasing final volume isn’t practical
when stock compound is rare/expensive
why must cuvettes be clean, unscratched, and free of debris, condensation, or fingerprints
because anything that blocks or scatters light can reduce transmitted/absorbed light and cause an error in measurements.
3 approaches to solving dilution problems
C1V1 = C2V2
dimensional analysis
serial dilution
is it better to use interpolation (estimating data between data points) or using the slope of a best-fit line
slope of a best-fit line since it smooths out random measurement error across multiple data points
also because slope (m) = εl under the conditions of the experiment
Based on our results from Experiment 2, Which is more accurate for reading absorbance, the spectrophotometer or the plate reader?
the plate reader showed higher % accuracy BASED ON OUR RESULTS