1/40
Comprehensive practice flashcards covering Malaysian Medical Device Act 737, electrical safety standards (IEC 60601 and IEC 62353), testing and commissioning, acceptance, calibration, and maintenance management according to MS 2058.
What is the definition of earth leakage current in a Class I medical device according to IEC 60601?
Earth leakage current is the current that normally flows from the mains part through insulation into the protective earth conductor of a Class I protectively earthed device.
What are the IEC 60601 maximum allowable limits for earth leakage current under Normal Condition (NC) and Single Fault Condition (SFC)?
Under Normal Condition (NC), the limit is 500μA (0.5mA). Under Single Fault Condition (SFC), the limit is 1000μA (1mA).
What are the four classifications of medical devices based on risk under the Medical Device Regulation 2012?
How are a ventilator and a phototherapy unit classified according to the Medical Device Regulation 2012?
A ventilator is classified as Class C (High-Moderate risk), while a phototherapy unit is classified as Class B (Low-Moderate risk).
What are the four types of establishment parties recognized under the Malaysian Medical Device Act 2012 (Act 737)?

Which IEC 60601 applied part classification is represented by this symbol?
Type B Applied Part (a non-isolated applied part connected to protective earth, depicted as a standing human figure without an enclosing square box).
How are Class II equipment and Type CF applied parts graphically symbolized under IEC 60601?
Class II Equipment is represented by two concentric squares (a smaller square inside a larger square, signifying double or reinforced insulation). A Type CF Applied Part is represented by a heart symbol enclosed inside a square box.
What are four physiological effects of alternating current (AC) on the human body across increasing current thresholds?
How do IEC 60601 and IEC 62353 differ in terms of purpose, ground bond test current, and test stress?
What is enclosure (chassis) leakage current, and what are its IEC 60601 allowable limits?
It is unintentional current flowing from live mains or internal electronics through insulation to an exposed, non-earthed conductive part of the chassis, and through a person to earth. Under IEC 60601, it must not exceed 0.1mA (100μA) in Normal Condition and 0.5mA (500μA) in Single Fault Condition.
What are the sequential steps in the MDA Medical Device Registration Process under Act 737?
What is the full name and primary function of the IEC?
IEC stands for International Electrotechnical Commission. It is an international organization that prepares and publishes consensus standards for all electrical, electronic, and related technologies, including medical electrical equipment safety standards.
What is equipment leakage current under IEC 62353, and what are the pass/fail limits for Class I devices using the Direct Method?
Equipment leakage current is the total leakage current flowing from the mains parts to earth via accessible conductive parts of the enclosure and applied parts. For Class I equipment, both normal and reversed polarities must not exceed 500μA (0.5mA).
What is the core definition of Medical Electrical (ME) Equipment under IEC 60601?
ME Equipment is electrical equipment having an applied part or transferring energy to or from the patient (or detecting such energy transfer), provided with not more than one connection to a particular supply mains, and intended for diagnosis, treatment, or monitoring of a patient.
What is the Healthcare Technology Life Cycle, and what are its three operational stages?
It refers to the continuous management activities relating to medical devices from initial conception to final decommissioning. Its three stages are:
Who are the three designated responsible parties on a Testing & Commissioning (T&C) Certificate?
What are the three main professional interaction groups of a Biomedical or Clinical Engineer?
What are the three overall phases of the Medical Device Life Cycle?
What are the two primary objectives of medical device regulation in Malaysia?
What are four reasons for conducting Electrical Safety Testing (EST) on medical equipment?
What factors determine the physiological severity of an electric shock?
The amount of current (I=RE), the path taken through the body (e.g., across the heart/chest), the duration of shock exposure, and the frequency (f) of the electrical current.
What is the purpose of the Medical Device Acceptance Procedure?
To ensure all staff involved in the purchase, loan, or trial of medical devices follow a standardized process verifying that equipment meets technical specifications and clinical needs before handover and patient use.
What are the calibration requirements for an ECG machine during Planned Preventive Maintenance (PPM)?
It falls under active medical device functionality calibration. It is performed quantitatively by technical staff using a calibrated ECG/patient simulator or built-in self-calibration. An external ISO 17025 certificate is not required, but quantitative results must be documented in the PPM report.
Name six active medical devices that do NOT require installation.
What conditions classify an active medical device as requiring installation?
What are the functions of the ISOTECH TTI-10, 935-14-82/TTI PRT, and 989 Blackbody Source in an infrared thermometer calibration setup?
What are the two pre-requirement checks for active medical devices that do NOT require installation?
What is the definition of Medical Device Equipment Calibration (MDEC)?
A procedure for detecting and fixing measurement uncertainties by comparing a device of unknown uncertainty against a reference standard of known uncertainty to ensure acceptable measurement accuracy.

What radiation shielding specifications are designated in this X-ray room layout?
Lead shielding of 2.0mmPb thickness across the brick walls, lead-lined doors, lead glass control window, and wall Bucky stand backing, accompanied by an external door warning light and a 3-phase 415V isolator switch.
What are two mandatory calibration criteria for measuring or test equipment like Electrical Safety Analyzers?

What are the specified workbench dimensions in this MS 2058 Annex M Biomedical Engineering Services workshop floor plan?
2′-6′′×10′-0′′ (with dedicated admin workstation at 2′-0′′, cleaning area at 6′-0′′, equipment storage at 3′-0′′, and consumables storage at 4′-0′′).
What are the three critical planning factors when designing a medical device maintenance program?
What are the five sequential stages of the Biomedical Equipment Life Cycle in Malaysian hospitals?
What documents must be included in a Hospital Specific Implementation Plan (HSIP)?
Organization Chart, Quality Assurance Program (QAP), list of subcontractors & equipment, layout of BES facilities, standard operating procedures/forms, Scheduled Maintenance Planner, User Training Planner, and list of BER equipment.
What are two main purposes of maintaining an equipment inventory in a biomedical maintenance program?
What is the procedure for Decommissioning & Disposal (Stage 5) of biomedical equipment?
BES prepares Beyond Economic Repair (BER) documentation, the Hospital/User endorses and issues the BER certificate, and disposal is executed following MS 2650 (and Act 304 for ionizing radiation equipment) after complete decontamination.
How are physical and human resources classified under Initial Costs versus Operating Costs in maintenance planning?
What are the four core technical support functions (Stage 4) of Biomedical Engineering Services?
What are the response time limits for Unscheduled Maintenance (Corrective Maintenance) under MS 2058?
Emergency response must occur within <15mins, while normal routine calls require a response within <120mins.
What are the three maintenance delivery models for hospital medical equipment?
What are the three common work prioritization models under operational management?