1/207
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Gas-filled detector regions in order of increasing voltage
Recombination, ionization, proportional, limited proportional, GM, continuous discharge
Mnemonic for the gas ionization curve regions
Really Intelligent People Like Good Coffee
Two useful flat plateaus of the gas ionization curve
Ionization chamber and Geiger-Mueller
Gas-detector region in which the dose calibrator operates
Ionization chamber region
Gas-detector region with full Townsend avalanche and identical pulse sizes
Geiger-Mueller region
Gas-detector region that can distinguish alpha from beta particles
Proportional region
Why a GM meter cannot measure photon energy
Avalanche makes every pulse identical
Instrument of choice to FIND low-level contamination
GM survey meter
Instrument of choice to MEASURE accurate high exposure rates
Ionization chamber (cutie pie)
GM meter behavior at very high dose rates
Saturates and reads falsely low
Survey instrument used for therapy patient release readings
Cutie pie (ionization chamber)
Distances for patient release survey readings
Contact, 30 cm and 1 meter
NRC-required survey meter detection range
0.1 mR/hr to 100 mR/hr
GM probe that must be pointed directly at the source
End-window probe
GM probe used to survey a larger area
Pancake probe
Survey meter daily check-source tolerance per Bolus
Within 20% of calibrated reading
Survey meter annual calibration tolerance (10 CFR 35.61)
20%
Regulation governing survey meter calibration
10 CFR 35.61
When survey meters must be calibrated
Before first use, annually, after repair
Retention period for survey meter calibration records
3 years
Three daily survey meter checks before use
Battery, check source, calibration sticker date
Dose calibrator detector type
Pressurized well ionization chamber
Dose calibrator activity range
0.01 µCi to curies
Dose calibrator constancy frequency
Daily before use
Dose calibrator linearity frequency
Quarterly
Dose calibrator accuracy frequency
Annually
Dose calibrator geometry frequency
At installation and when changing syringe/vial size
Acceptable limit for all dose calibrator QC tests
±10%
Mnemonic for dose calibrator test frequencies
Can Linda Assay Good (C daily, L quarterly, A annually, G install)
Action when dose calibrator accuracy or constancy fails
Repair or replace
Action when dose calibrator linearity or geometry fails
Apply a correction factor
Dose calibrator test that measures precision (reproducibility)
Constancy
Dose calibrator test of accuracy across a wide range of activities
Linearity
Dose calibrator percent error formula
(Expected − Actual) ÷ Expected × 100
Correction factor formula
Expected ÷ Actual
True activity using a correction factor
Reading × CF
Minimum number of sources for dose calibrator accuracy
Two
Minimum activity of dose calibrator accuracy sources
50 µCi (preferably >200 µCi)
Energy range one accuracy source must fall within
100–500 keV
Half-life and energy of Cs-137
30 years, 662 keV
Half-life and energy of Ba-133
10.74 years, 356 keV
Half-life and energy of Co-57
271 days, 122 keV
Settings on which constancy is measured
Every commonly used radionuclide setting
Constancy expected 78.5 µCi, reads 70.1 µCi. Result?
10.7% error, fails; repair
Last week's constancy average 142.2 µCi. Acceptable range?
128.0 to 156.4 µCi
Lowest activity for linearity decay testing (Bolus)
30 µCi or less
Lead-sleeve method for same-day linearity
Shield method (Cali-check or Lineator)
Short-lived isotope usable for same-day linearity decay test
F-18 (110 min)
5 mL syringe should read 2.95 mCi but reads 2.45. Correction factor?
1.2
Dose in that syringe reads 12.6 mCi with CF 1.2. True activity?
15.1 mCi
Retention period for dose calibrator QC records
3 years
Geometry of a well counter
Near-4π
Most sensitive sample counter in nuclear medicine
Well counter
Activity above which a well counter shows dead-time losses
About 2 µCi
Lead shielding around a well counter detector
About 5 cm
Minimum counts for well counter or probe samples
10,000 counts
Percent error for 10,000 counts
1%
Percent uncertainty formula
100 ÷ √N
Unusually HIGH well-counter constancy reading suggests
Contamination or a nearby source
Counting efficiency formula
DPM = CPM ÷ efficiency
Well counter efficiency 85%, wipe 120 cpm net. DPM?
141 dpm
Collimator type on a thyroid uptake probe
Flat-field collimator
%RAIU formula
(Neck − Thigh) ÷ (Standard − Background) × 100
Purpose of the thigh count in RAIU
Corrects for non-thyroid body background
Neck 12,000, thigh 2,000, standard 42,000, bkg 2,000 cpm. RAIU?
25%
Is the uptake probe standard capsule diluted?
No
Cs-137 10% window LLD and ULD
629 and 695 keV
Energy resolution formula
FWHM ÷ photopeak energy × 100
Smaller energy resolution percentage means
Better resolution (narrower peak)
Typical Cs-137 energy resolution for a probe
8–12%
Cs-137 FWHM 535 to 605 keV, peak 570. Energy resolution?
12.3%
Scintillation spectrometer energy resolution frequency
Annually
Well counter/probe voltage calibration (peaking) frequency
Daily
Well counter/probe constancy tolerance
±10%
Correct operating voltage during peaking
Setting with the highest count rate
Chi-square formula
Σ(N − N̄)² ÷ N̄
What chi-square tests
Whether count spread matches random decay
Bolus chi-square protocol
10 counts, 9 degrees of freedom
Acceptable chi-square P value range
0.1 to 0.9
Chi-square 46,998 ÷ 10,016 = 4.69 with 9 df. Result?
Acceptable (P 0.5–0.9)
Counting statistics: % within ±1, 2, 3 SD
68%, 95%, 99.7%
Room pressure required for Xe-133 and aerosol use
Negative pressure
Photon energy of Xe-133
81 keV
Half-life of Xe-133
5.24 days
Regulation known as the xenon gas rule
10 CFR 35.205
Order of ventilation and perfusion in same-day Xe-133 V/Q
Ventilation before perfusion
Common aerosol ventilation agent
Tc-99m DTPA
Fraction of loaded aerosol activity reaching the lungs
About 1–3%
Ultrafine carbon pseudogas preferred in COPD
Technegas
Bruce protocol stage length
3 minutes
Bruce stage 1 speed and grade
1.7 mph, 10%
1 MET equals
3.5 mL O₂/kg/min
Instrument required for H-3 wipe tests
Liquid scintillation counter
Maximum beta energy of H-3
18.6 keV
Quench type where light is never produced
Chemical quenching
Most common LSC quench correction method
External standard ratio (Ba-133)
Half-life of Rb-82
75 seconds
Only acceptable Rb-82 generator eluent
Additive-free 0.9% saline
Volume of Rb-82 eluate discarded each day
First 50 mL
When Sr breakthrough must be measured
Before the first patient dose each day