stationary phase
the non-moving phase coated inside the column with which sample components interact as they travel through
mobile phase
the liquid/gas/superfluid that flows through the system, carrying sample molecules down the column
partition coefficient (equation)
the equilibrium constant representing the ratio of analyte concentration in the stationary phase to that in the mobile phase:

retention time
the total time elapsed between the sample injection and the point where the analyte peak reaches the maximum signal at the detector. It measures how long a specific component spends traveling through the column.
chromatogram x-axis
retention time
chromatogram y-axis
detector signal/intensity
calibration curve x-axis
concentration
calibration curve y-axis
peak area
In general, how would you determine which of several peaks in a chromatogram obtained for a mouthwash sample might correspond to ethanol? (Is your suggested method 100% reliable?) How would you use that peak to determine how much ethanol is present in the mouthwash?
Compare the mouthwash peak’s retention time to a pure ethanol standard run under the same conditions. This is not 100% reliable because different compounds can have the same retention time (co-elution). To determine ethanol concentration, measure the peak area, create a calibration curve of standard peak area vs. concentration, and use the line equation (y = mx + b) to calculate the concentration.
In general, will an increase in operating temperature cause a given peak to have a longer or a shorter retention time? Why?
An increase in temperature causes a shorter retention time because higher temperature increases the solute’s vapor pressure, favoring the mobile (gas) phase over the stationary phase, so the solute moves through the column faster.
In general, will an increase in mobile phase flow rate cause a given peak to have a longer or a shorter retention time? Why?
Increasing the mobile phase flow rate will cause shorter retention time. This is because A higher mobile-phase flow rate moves the vaporized sample through the column faster, reducing the time the molecules spend in the column.
Suppose that you need to “speed up” a chromatogram (i.e., decrease the retention times of all the peaks). Suggest three specific GC conditions that you could change in order to accomplish this.
increase column temperature, increase carrier-gas flow rate, and decrease column length
In a GC experiment, a student injects 2.0 μL of a 20% ethanol/water solution into the GC. Calculate the weight of ethanol that has been injected. (The density of ethanol is 0.789 g/mL at 20° C.)
3.2 × 10^-4 g or 0.32 mg of ethanol