Common terms/concepts related to microscope cameras
Linear Full-Well Capacity (LFWC)
Maximum number of electrons that can be collected and held by a single pixel on the camera’s sensor linearly BEFORE saturation
Pixel stepping
Camera/sensor moves by a very small, slight, precise distance (usually a fraction of a pixel) between successive exposures, increasing image resolution & color accuracy while reducing sensor artifacts
Peltier cooling
Solid-state cooling method
Peltier effect
When there’s electric current, one side becomes cold and the other becomes hot
Readout process
Step where camera converts captured light into digital image that can be displayed on computer
Triggering I/O
Hardware signals that lets camera synchronize image acquisition with external devices
Quantum efficiency
(e- generated/photons received) x 100%80%
What does 80% quantum efficiency mean?
For every 100 photons that hit the sensor, 80 e- are generated
Rolling shutter
Sensor exposed and read out row by row
CMOS
Complementary metal oxide semiconductor
CCD
Charge-coupled device
EMCCD
Electron multiplying charge-coupled device
Bit depth
How many depth levels a pixel can store
(Bit depth)2
Number of gray levels the pixel can measure
Conversion gain ratio
Measure of sensitivity; high gain = high contrast, low gain = low contrast
How to improve Quantum Efficiency?
Microlensing (putting tiny lenses on top of each pixel) or Back illumination
Bayer mask
Colored filter array placed on top of color camera sensors, allows monochrome sensor to capture color
Impact ionization
Electron loses kinetic energy by collision with another bound electron, knocking it into a higher energy state and creating secondary electrons → drives electron multiplier in EMCCD
Clock phase
Control electrode used to move charge through CCD
Adaptive optics
Technology that corrects optical aberrations in real time
Fluorescence correlation spectroscopy