dipstick steps

2
New cards

dipstick possible errors

  • Dipping too long.

  • Not removing excess urine.

  • Timing incorrectly.

  • Not mixing urine and missing formed elements that sink to the bottom.

  • Strips are protected from heat, light, and chemicals.

  • The bottle must have a desiccant to take out moisture.

  • Store dipsticks at room temperature.


3
New cards

why we mix urine

Formed elements, such as red blood cells, white blood cells, and casts, can settle at the bottom. If you do not mix the sample, you might miss them.

4
New cards

What should you do before testing urine with a reagent strip?

mix the urine

5
New cards

QC does what

Checks that the reagent strips are working correctly before patient results are reported.

6
New cards

QC done

uDipsticks are tested with control solutions

uControl Level – negative control

uControl Level – positive control

uBoth a positive and negative reaction should be seen.

uDone every 24 hours and with each new bottle/ lot of reagent strips

7
New cards

Which medication can interfere with urine dipstick reactions because it strongly colors the urine?

Pyridium- turns urine orange or reddish-orange= possible interference, must use confirmation test

8
New cards

What should be considered when urine pigmentation makes a dipstick result difficult to interpret?

Use an appropriate confirmation test

9
New cards

What can protein in urine indicate?

Possible kidney damage or renal disease

10
New cards

Proteinuria caused by a problem before blood reaches the kidneys is:

prerenal

11
New cards

Proteinuria caused by glomerular damage is:

renal

12
New cards

Protein entering urine from bladder inflammation is:

postrenal

13
New cards

Where is albumin produced?

liver

14
New cards

What is a major function of albumin?

maintains blood oncotic

15
New cards

what protiens screened in dip stick

only albumin. SSA screening test may require for other proteins detection

16
New cards

transferrin

transport iron in the blood

17
New cards

tamm-horsfall protein

normal urinary glycoprotein made by renal tubules

18
New cards

multistix

detects albumin best, less sensitive to other proteins, SSA test detect a wider range of proteins in the urine

19
New cards

Bence Jones proteinuria is classified as which type in the traditional classification of proteinuria?

Prerenal


Memory tip: Multiple myeloma → plasma cells → light chains → Bence Jones proteins → prerenal proteinuria.

20
New cards

At what temperature do Bence Jones proteins traditionally precipitate?

A. 0–10°C B. 40–60°C C. 100°C only

B. 40–60°C

21
New cards

A patient has Bence Jones proteins in the urine. Which result is possible?

A. Negative Multistix, positive SSA

B. Positive Multistix, negative SSA always

C. Both tests must be negativ

A. Negative Multistix, positive SSA

22
New cards

neg multistix protein result mean

proteinuria may still be there

23
New cards

limitations of dipstick proteins

  1. mainly detects albumin

  2. false-positive= highly alkaline urine

  3. false neg = other protein go undetected

  4. highly pigmented urine

  5. timing reading error


24
New cards

All protein methods measure the amount of protein or albumin in urine, but this does not take into consideration the patient's hydration state.

  • Dehydration: Urine becomes more concentrated, which can make protein concentration appear higher.

  • Overhydration: Urine becomes more diluted, which can make protein concentration appear lower.

  • 24-hour urine total protein test: Measures the total amount of protein excreted over 24 hours. It has been used to assess kidney damage, but collection can be time-consuming and inconvenient.


25
New cards

physical examination

evaluates color, clarity, odor, and specific gravity.

26
New cards

chemical examination

uses reagent strips and backup or confirmation tests when initial results are questionable or interfering substances may cause false-positive or false-negative results.

27
New cards

microscopic examination

examines cells, casts, crystals, microorganisms, and other formed elements.