Vocabulary flashcards covering X-ray tube shielding, secondary barriers, console controls, collimation, filtration, and radiation protection principles.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Leakage Radiation
Any X-rays that escape through the protective tube housing rather than the collimator port when operating at maximum voltage.
Primary Shielding
Radiation shielding positioned anywhere the X-ray tube can aim directly, engineered to withstand the full potential beam energy (maximum kVp).
Secondary Barrier
Protective barriers designed to attenuate scatter radiation rather than the primary beam, such as control booth windows or non-primary walls.
Radiolucent
The property of a material (such as a wood and carbon fiber composite examination table) that allows X-rays to pass through without appearing on the finished image or altering exposure.
Source-to-Image Distance (SID)
The distance from the X-ray tube target to the image receptor; standard practice requires maintaining at least 40inches (or 36inches where calibrated) to reduce patient entrance dose caused by beam divergence.
Detents
Mechanical stops that align the X-ray tube at fixed distances (such as 72inches), which regulatory standards often require to be calibrated within an error margin of 1% to 2%.
Aperture
The opening and closing mechanism of the collimator light field, which is round in a C-arm fluoroscopy unit and square in standard radiography systems.
Tight Collimation
The restriction of the exposed X-ray field size strictly to the anatomical area of clinical interest, primarily functioning to decrease undesirable scatter radiation and patient dose.
Inherent Filtration
Filtration naturally present in the construction of the X-ray tube assembly, consisting of the glass envelope and the surrounding insulating oil.
Added Filtration
Absorbing materials intentionally placed into the primary beam path (such as aluminum sheets mounted in the collimator box) to remove low-energy photons below 40kVp.
Aluminum Filter
A lightweight, low-density metallic filter whose molecular structure allows diagnostic photons to penetrate while selectively absorbing non-diagnostic photons under 40kVp.
Underexposure
An exposure error that compromises image quality and is considered as hazardous to cumulative patient dose as overexposure because it necessitates repeating the projection.