Sampling - what can go wrong?
Wrong location, depth, timing → sample never represented population
Preservation - what can go wrong?
Analyte degrades, volatilizes, or reacts before getting to lab
Sample prep - what can go wrong?
extraction, digestion, or dilution introduces loss/contamination
Separation - what can go wrong?
Poor resolution means overlapping peaks are misquantified
Detection - what can go wrong?
Even perfect instrument reports what was put in. Garbage in garbage out
Variance equation
Sigma²=sigma² sampling + sigma² prep + sigma² analysis
What are five sampling failures?
Representativeness (Real) Design (Dogs) Contamination (Can’t) Preservation (Play) Documentation (Dominoes)
Category of failure matters more than…
Mistake
Grab Sample
Single sample taken at one place at one time
Composite Sample
Multiple grab samples taken and combined before analysis
Incremental Design
Small increments collected and combined systematically
If I double the particle size how much does the sample mass multiply by?
d³… so 8x (Gy’s)
What does a larger sample mass do to the random sampling variability?
Lowers random sampling variability
What does a small sample size do to variance?
Increases it
Compositional heterogeneity
Particles are different to each other in composition
Distributional heterogeneity
Particles of a given composition are not randomly distributed
What could go wrong with preservation of volatile organics?
Volatilization, biodegradation
What could go wrong with preservation of trace metals?
Adsorption to container walls, precipitation
What could go wrong with preservation of microbial/biological?
Growth or die-off before analysis
What could go wrong with preservation of pesticides
Photodegradation, adsorption
Mu - population parameters
Population mean - mean of entire population
Sigma - population parameters
Population std dev - variability of target pop.
x bar - population parameters
Sample mean - best estimate of pop. mean
S - population parameters
Sample std. dev. - best estimate of population std. dev
Standard error of mean shrink with..
Square root of n
What to do other than run Q-test?
Document investigation, follow QC plan. Do not remove results due to inconvenience
What does compositing reduce?
Variance (and hides information)
What do more samples do to mean?
Improve confidence in it.
What does more samples NOT improve?
Uniformity among individual results
Steps of creating sampling plan?
State problem, identify decision, specify tolerable error, design the plan
Systematic sampling grid
Regular spacing across whole grid.
Stratified sampling grid
Subareas with 3 identical spots each
Judgemental sampling grid
All dots in hot spot area
Coning and quartering
Preserves representativeness by reducing mass
Flow-proportional not time-proportional
Flow → standard for wastewater
Time → over represents low-flow periods and under represents pollutant load
Active: sampling gases
Pulls known volume of air through filter over defined time
Passive: sampling gases
Diffusion, no pump, trades precision for implicity
Container choice for trace metals
Acid washes HDPE
Container choice for volatile organics
Sealed glass vial
Container choice for light sensitive organics
Amber glass
Solid - dominant heterogeneity
Compositional and distributional
Liquid - dominant heterogeneity
Mostly temporal, flow, discharge cycles
Gas - dominant heterogeneity
Mostly temporal, low spatial variation
Field blank
Water taken to field exposed to air. Catches contamination
Trip blank
Clean water goes with samples but unopened
Field duplicate
2 samples same location. Measures precision and combined sampling
Matrix spike
Known amount of analyte added to sample
Mistake - Convenience bias
Sampling where its easy
Mistake - undersized n
Skipping sample size calculation
Mistake - Ignoring particle size
Using too little mass for coarse sample
Mistake - Over-compositing
Compositing away hot spot signal
Mistake - Custody gaps
Missed documentation
Stationary phase
Particles or liquid film held inside column
Mobile phase
Gas or liquid thar flows through column and carries solutes
Distribution constant K formula
K=Cs/Cm (C concentration in ____ phase)
Low distribution constant means:
Less time spent in column
K constants can be only compared to K constants from…
samples from same instrument
tm
dead time → time for unretained solute to pass thru column
tR
retention time → time from injection to peak max of given solute
tR’
adjusted - tR’=tR-tM
Retention factor k’
k’=(tR-tM)/tM
Retention volume
VR=tR x F
Selectiity - comparing solutes with a (alpha)
alpha = k’2/k’1 (k’2>k’1)
Permanent gas case
Adsorption - separate gases on solid (porous)
Inorganic ions case
ion exchange, separate Na+ K+ and Ca2+
Target protein case
affinity, capture one tagged protein from cell lysate
Organic solute case
partition, separate neutal organic solutes by polarity
Packed column chromatography
Mobile phase flows thru particles held inside tube
TLC
mobile phase climbs coated plate by capillary action
Normal phase
Stationary: polar silica surface
Reversed phase
Stationary: non polar c18 surface
Gas v liquid chromatography
Carrier gas moves vaporized sample thru heated column. Liquid solvent pumped thru packed column
Column efficiency and band width
H=L/N → higher plate number means less band broadening
Peak width → calc. plate Number
Baseline width: w=4 sigma N=16 (tR/w)² . Half height width: w1/2=2.35 sigma
Effective plates → correcting N for dead time
N eff = 16 (tR’ / w)²
Fronting
Stationary phases become saturated, analytes skip steps and crowd up front
Multiple paths
different routes around packing particles
Longitudinal diffusion
band diffuses forwards and backwards
Mass transfer
molecules do not equilibrate instantly