C, lambda, and nu relation

C = λv

2
New cards

wavenumber units

inverse cm

3
New cards

what is v̄

wavenumbers, or 1/λ

4
New cards

Energy relations

E = hv = hc/λ = hcv̄

5
New cards

Lowest to highest energy EM spectrum

Radio waves, Microwave, Infrared, Visible, UV, X ray, Gamma rays

6
New cards

visible light spectrum

800nm (red, low energy) to 400nm (blue, high energy)

7
New cards

reflection


light bounces off hard surface

8
New cards

what are often used in optics of instruments

mirrors, prisms

9
New cards

attenuated total reflection can…

…be used as a spectroscopy technique (IR)

10
New cards

refraction


bending of light after passing between two different media

ex: prism monochromators

11
New cards

diffraction

bending of light after passing through a small opening

ex: x-ray diffraction and grating monochromators

12
New cards
13
New cards

describe absorption of light

transition from ground electronic state to higher energy level through added light energy

14
New cards

describe emission of light

transition from higher energy level to lower (or ground) electronic state through release of light energy

15
New cards

true or false: all transitions are allowed

false

16
New cards

What is the first selection rule (equation and words)

Ephoton = hv= ∆Etransition

the energy of the photon must be equal to the energy needed for the transition

17
New cards

what is the second selection rule

Spin Selection Rule: Electron spin cannot change during a transition.

  • The total spin change must be zero (ΔS=0\Delta S = 0)

  • Spin multiplicity (2S+12S + 1) must remain the same in the ground and excited states (e.g., singlet to singlet)


18
New cards

what is the third selection rule

change in dipole needed for IR

sigma to sigma star, pi to pi star for UV vis

19
New cards

atoms have only __ states

electronic

20
New cards

spacing between energy levels is ___ in atoms

large

21
New cards

solution spectra are..

…broadened due to collisions and solvent

22
New cards

excited states will __ __ ___ to the __ __

relax back down, ground state

23
New cards

radiative emission

atom emits photon with energy matching energy gap

24
New cards

non radiative emission

atom relaxes without emission of photon

  • collisional deactivation

  • release energy as heat to solvent


25
New cards

molecules can emit a photon ____ or with ___

spontaneously, stimulation

26
New cards

sources of atomic emission

  • line sources

  • excitation through electrical means


27
New cards

What is the primary difference in timing and mechanism between fluorescence and phosphorescence?

  • Fluorescence happens rapidly as a molecule drops from a singlet state (to the ground state

  • Phosphorescence is much slower (seconds to hours) because it involves a "forbidden" spin flip, dropping from a triplet state to the ground state



28
New cards

How do electron spins differ between a singlet state and a triplet state in a Jablonski diagram?

  • In a singlet state all electron spins are paired (opposite directions).

  • In a triplet state the excited electron undergoes a spin flip and becomes parallel (same direction) to its partner.



29
New cards

Identify which processes on a Jablonski diagram emit a photon (radiative) and which release heat (non-radiative).


  • Radiative (Emits light): Fluorescence and Phosphorescence.

  • Non-Radiative (Emits heat): Vibrational Relaxation, Internal Conversion (IC), Intersystem Crossing (ISC), and External Conversion (EC).



30
New cards

What dictates whether a non-radiative transition is classified as Internal Conversion (IC) or Intersystem Crossing (ISC)?


Internal Conversion (IC): Occurs between states of the same spin multiplicity

Intersystem Crossing (ISC): Occurs between states of different spin multiplicities

31
New cards

What are the five basic components of a spectrometer

radiation source, wavelength selection, sample containement, radiation detector, signal processing and reading

32
New cards

What do you want in a radiation source

high intensity output

flat intensity vs wavelength response

low source intensity variability

33
New cards

two main types of radiation sources

line and continuous

34
New cards

Characteristics of continuous radiation sources

  • emits over large range of wavelengths

  • many based on blackbody radiation

  • intensity and wavelength are a function of T

  • max wavelength is inversely proportional to T


35
New cards

examples of continuous radiation sources

Infrared - globar and nichrome wire

visible - tungsten lamp

ultraviolet - deuterium or xenon lamp, high pressure mercury arc lamp

36
New cards

examples of line sources

low pressure metal vapor lamps - Na and Hg

hollow cathode lamp - element specific

37
New cards

Lasers basics

  • stimulated emission of radiation

  • need 3-4 quantum states w appropriate lifetimes

  • give high intensity monochromatic coherent light


38
New cards

What is the problem with lasers

rate of emission/absoprtion is proportional to number of molecules in excited/ground state

  • need more population in excited state

  • no two level systems


39
New cards

what is a monochromator’s function

seperate light frequencies with dispersion to select a single wavelength from a broadband light source

40
New cards

what are the three types of monochromators

filter, prism, grating

41
New cards

Filters basics

  • polychromator

  • low pass: absorb wavelengths above cutoff

  • high pass: absorb wavelengths below cutoff

  • notch: high and low wavelength cutoffs


42
New cards

Prism basics

  • refraction of light → seperate light by dispersion

  • combined with an exit slit


43
New cards

Prism materials

UV - quartz

visible - glass

IR - NaCl or KBr

44
New cards

True or false: prisms disperse all wavelengths equally in space

false

45
New cards

seperation in dispersion is better at…

…short wavelengths

46
New cards

fixed slits cause different…

…bandwiths

47
New cards

grating basics

  • diffraction off of finely spaced grating

  • different wavelengths in phase at different points along plane

  • rotating grating → change wavelength

  • normal dispersion, not wavelength dependent


48
New cards

problem with grating..

Multiple lower-wavelength harmonics will diffract at the exact same angle as the fundamental wavelength, meaning they all get through together.

49
New cards

solution to overlapping harmonics in grating

  • Place optical filters in the path to block the shorter, lower-wavelength harmonics.



50
New cards

What is the Bragg’s Law formula

𝑛𝜆=2𝑑sin(𝜃)

51
New cards

what do n =1 and n = 2 to infinity represent in Bragg’s Law

n = 1 → fundamental wavlength

n = 2 to infinity → higher order harmonics

52
New cards

How do you find the line spacing (𝑑) and the 2nd harmonic wavelength (𝜆_2) for an 800 lines/mm grating at 𝜃=33

  1. find d (1/line spacing)

  2. find fundamental (1*line spacing * sin(33))

  3. find second harmonic (fundamental/2)


53
New cards

advantages of gratings over prisms

  • linear dispersion

  • larger usefull range

  • smaller


54
New cards

disadvantages of gratings over prisms

  • cost

  • more stray light

  • higher order harmonics


55
New cards

photovoltaic detector

range: 250-750nm

photon makes electrol-hole pair in semiconductor → creates current

intesity proportional to the number of photons

56
New cards

photovoltaic detector advantages

  • no external power supply

  • rugged


57
New cards

photovoltaic detector disadvantages

  • limited wavelength range

  • not very sensitive


58
New cards

phototube detector

  • similar to photovoltaic but powered

  • photon → ejection of electrons

  • electrons collected at anode


59
New cards

phototube detector advantages

  • more sensitive

  • can work in UV and Visible range


60
New cards

phototube detector disadvantages

  • dark current → spon. emission of electrons

  • expensive


61
New cards

photomultiplier tube detector

  • much more sensitive version of phototube

  • has anodes with amplification at each step

  • high gain of electrons


62
New cards

photomultiplier tube detector advantages

  • sensitivtiy

  • versatility


63
New cards

photomultiplier tube detector disadvantages

  • dark current

  • cost


64
New cards

photodiode array detector

  • current flow only upon light excitation

  • very small

  • put sample between source and grating

    • single PMT → sample after grating

  • replace exit slit w/ array of photodiodes

  • measures all wavelengths at once

  • and less sensitive


65
New cards

IR Heat detector

  • need to measure IR photons

  • need to measure very small change in temperature



66
New cards

two types of IR heat detectors

thermocouple

golay cell

67
New cards

Thermocouple IR heat detector

  • two unlike metals

  • voltage develops at junction

  • temperature and pressure dependent voltage


68
New cards

Golay cell IR heat detector

  • temperature change (absorption) cuases change in pressure

  • pressure change can be read using optical system


69
New cards

Lambert-Beer’s law

  • light power decreased through absorption

  • I = (I_0) e ^(-σ_λnl)


70
New cards

Calibration curve for Beer’s Law

  • use internal standards

  • only valid if beer’s law is linear in that concentration region

  • can be used to interpolate


71
New cards

Describe standard addition

adding increasing amounts of standards to samples at fixed volumes

make a plot of A vs total concentration of standard added

x intercept = negative of initial concentration

72
New cards

C_0 formula

(A0C1)/(A1-A0)

73
New cards

standard addition more in depth

  1. take several identical portions of unknown sample

  2. leave flask 0 alone (only unknown concentration)

  3. spike flasks 1,2,3 with increasing, known concentrations of a pure standard


Sample 0 (Unspiked):

𝐴0=𝑘⋅𝐶0

Sample 1:

𝐴1=𝑘(𝐶0+𝐶1)

Sample 2:

𝐴2=𝑘(𝐶0+𝐶1+𝐶2)


74
New cards

Most deviations from beer’s law are…

…negative deviations

75
New cards

to correct deviations in beer’s law, you CAN ___, but it is better to ___

interpolate, dilute

76
New cards

A is ___, but we usually measure ___

logarithmic, %T

77
New cards

error in %T is ___ and leads to ___________

constant, varying error in A

78
New cards

for less than 3.5% error, stay between….

… 15% and 65% or 0.82 and 0.18 (A)

79
New cards

Beer’s law is only accurate for what kind of solution?

Dilute solution

80
New cards

in beer’s law, molecules…

… should NOT be interacting with eachcother - this alters the electron cloud.

81
New cards

If a solution is too dilute or epsilon is too small..

.. you cannot accurately measure decrease in light intensity

82
New cards

what is bandwidth

range of light wavelengths getting through the monochromator.

  • usually constant

  • measure as close to maximum of epsilon vs lambda as possible

  • small bandwidth is good


83
New cards

stray radiation effect

hit detector → false signal

  • normally doesnt pass through signal

  • A will be smaller than it should be

  • can fix by cutting down on optics or painting it black


84
New cards

A_obs formula

-log(P+Ps/P0+Ps)

85
New cards

Describe why spectroscopy of atoms is simple

only sharp electronic transitions

86
New cards

What is the natural linewidth of atomic transitions?

Approximately 10^-5 nm.

87
New cards

What causes natural linewidth?

The energy–time uncertainty principle. The finite lifetime of an excited state creates uncertainty in its energy.

88
New cards

What causes Doppler broadening?

Random thermal motion of atoms or molecules causes Doppler shifts in transition wavelengths.

89
New cards

How does Doppler broadening affect a spectral peak?

It broadens the peak because particles moving at different velocities produce different frequency shifts.

90
New cards

What is a blue shift?

A shift to a shorter wavelength.

91
New cards

What is a red shift?

A shift to a longer wavelength.

92
New cards

What causes pressure broadening?

Interactions between particles perturb quantum energy levels, broadening spectral lines.

93
New cards

What four flame conditions are desirable in atomic absorption spectroscopy?

  • High degree of atomization

  • Low degree of excitation

  • Low degree of ionization

  • Minimal formation of oxides


94
New cards

Why should excitation be kept low in atomic absorption spectroscopy?

Atomic absorption primarily measures ground-state atoms. Excitation reduces the ground-state population.

95
New cards

What determines the excited-state population in a flame?

Flame temperature, according to the Boltzmann distribution.

96
New cards

What factors determine flame temperature?

  • Fuel and oxidant

  • Position within the flame

  • Burner construction


97
New cards

Why must samples be nebulized before flame analysis?

To convert a liquid sample into a fine aerosol that can be introduced into the flame.

98
New cards

What is atomization in atomic absorption spectroscopy?

The process of converting the analyte into free gas-phase atoms, separating it from its chemical compounds.

99
New cards

What two processes are required to prepare typical liquid samples for atomic absorption analysis?

  • Nebulization: Converts the liquid into an aerosol.

  • Atomization: Uses heat to produce free gas-phase atoms. This involves desolvation, vaporization, and atomization.


100
New cards

What are the two common atomization sources in atomic absorption spectroscopy?

  • Flame

  • Electrothermal furnace