4 major componenet of gamma camera

  1. collimator

  2. scintillation crystal (NaI)

  3. photomultiplier tubes (PMTs)

  4. Position and energy circuits


2
New cards

collimator function

allows photos traveling in the correct direction to reach the crystal and absorbs unwanted off-angle photons

3
New cards

NaI crystl function

cinverts gamma ray energy into visible light

4
New cards

PMTs function

convert light into electrical signals and amplify those signals

5
New cards

positioning and energy cicuit function

determin the X,Y and Z energy of each event, then determine whether the event should be accepted

6
New cards

sequence of gamma photon from patient to camera

  1. gamma photon leaves patient

  2. passes through the collimator if its direction is acceptable

  3. interacts with the NaI crystal

  4. produces scintillation light

  5. light reaches multiple PMTs

  6. PMTs convert light into electrical signals and amplify them

  7. positioning circuit calculate X and Y

  8. the sun circuit calculates Z, representing energy

  9. PHA checks whether the event falls withing the energy window

  10. accepted event is assigned to a pixel in the image matrix and stores


7
New cards

how does sodium iodide crystal convert gamma ray energy into light

  • gamma ray deposit energy in the crystal

  • electrons in crystal become excited

  • as electrons return to lower energy states they release energy as visible ligt

  • process is called scintillation


8
New cards

why must crystal be protects

  • hygroscopic = absorbs moisture

  • moistrure can dama crystal and make it cloudy or opaque

  • rapid temp changes cause thermal stress and cracking

  • 9 degree fereheintet per hour cracking risk


9
New cards

thicker crystal means _ sensitivity

higher

  • more photons absorbed

  • lower intrinsic spatial resolution


10
New cards

thinner crystal means __ sensitivty

lower

  • more photons may pass through

  • higher intrinsic spatial resolution


11
New cards

3 functions of PMTs

  1. convert scintillation light into electrical signals

  2. amplify the electric signals

  3. help determine photon energy and position


12
New cards

X ?

horiozontal position

13
New cards

Y ?

vertical positon

14
New cards

Z ?

total energy of the event

15
New cards

Pulse height analyzer function

  • determines whether a phtoon has an acceptable energy

  • has LLD, UUD, and energy window


16
New cards

energy window

  • range between upper and lower level discriminator

  • typically 20%


17
New cards

upper level discriminator

rejects events above selected energy range

18
New cards

lower level discriminaotr

rejects events below selected energy range

19
New cards

define peaking

  • process of adjusting the gamma cameras energy setting so that the photopeak is cenetered within the selected energy window


20
New cards

why does peaking (energy window alignment) matter

  • ensures the correct energy photons are accepted

  • prevents excessive rejection of desired photons

  • helps reduce scatter contamination

  • prevents nonuniformity caused by incorrect energy settings


21
New cards

what can shift a photopeak or change energy response

  • PMT gain changes or drift

  • changes in PMT high voltage

  • temperature changes

  • changes in detector electronics

  • changes in scintillation light output

  • incorrect energy calibration


22
New cards

what is a change in PMT gain

  • changes the size of the electrical signal and therefore can shift the measured photopeak


23
New cards

what does energy correct

z pulse/photopeak dfferences

24
New cards

what does linearity correct

X/Y positioning errors

25
New cards

what does unifomity/sensitivity correct

regional count sensitivity differences

26
New cards

energy and linearity corrections are applied…

event by event

27
New cards

PMT gain drift occurs when…

the amplification of a PMT changes over time

28
New cards

autotuning

automatically adjusts PMT voltage and preamplifier gain to maintain stable output

29
New cards

what are the three methods for autotuning

  1. photopeak monitoring

  2. split photopeak monitoring

  3. LED monitoring


30
New cards

what is photopeak monitoring

  • method of autotuning

  • monitors the photopeak at each PMT and adjusts gain relative to a reference


31
New cards

what is split photopeak monitoring

  • method of autotuning

  • uses two narrow energy windows on the high side of the photopeak and adjusts high voltage to maintain a predetermined count ratio


32
New cards

what is LED monitoring

  • method of autotuning

  • uses flashes of light from LEDs and compares each pMTs voltage output to a reference


33
New cards

what is the typical adjustment range for PMT gain drift

plus or minus 3keV

  • PMTs outside of range require engineer service


34
New cards

what can a poorly tuned PMT cause

  • nonuniformity and positioning errors even when correction maps are being used


35
New cards

Edge packing

  • occurs when photos are incorrectly positioned near the outer edge of the detector

    • PMTS at edge have different light detection characteristics

    • positioning calculations become less accurate

    • counts may become artificially concentrated near the detector edges

    • this profuces an abnormal flood pattern

  • a positional problem