Average photon energy of a bremsstrahlung beam
Approximately one-third of the maximum/nominal energy.
Co-60 photon energies
1.17 and 1.33 MeV.
Average Co-60 photon energy
~1.25 MeV.
Co-60 dmax
~0.5 cm.
6 MV dmax for 10×10 cm²
~1.5 cm.
10 MV dmax for 10×10 cm²
~2.4 cm.
18 MV dmax for 10×10 cm²
~3.3 cm.
PDD at 10 cm for Co-60, 10×10 cm²
~55%.
PDD at 10 cm for 6 MV, 10×10 cm²
~65%.
PDD at 10 cm for 10 MV, 10×10 cm²
~75%.
PDD at 10 cm for 18 MV, 10×10 cm²
~80%.
TMR at 10 cm for 6 MV
~0.78.
TMR at 10 cm for 10 MV
~0.85.
TMR at 10 cm for 18 MV
~0.90.
Approximate attenuation rate for 6 MV
~3% per cm.
Approximate attenuation rate for 10 MV
~2.5% per cm.
Approximate attenuation rate for 18 MV
~2% per cm.
Surface dose for Co-60
~50%.
Surface dose for 6 MV
~25%.
Surface dose for 10 MV
~23%.
Surface dose for 18 MV
~20%.
Scatter at 1 m from phantom
~1/1000 of primary beam.
MV f-factor in air, conversion for Roentgent to cGy
0.876 cGy/R.
MV f-factor in tissue, conversion for Roentgen to cGy
~0.97 cGy/R.
I-125 exposure rate constant
1.45 R cm2/mCi-hr
Ir-192 exposure rate constant
4.69 R cm2/ mCi-hr
Physical wedge out-of-field scatter
~2.5% of central axis dose.
Dynamic wedge out-of-field scatter
~1.0% of central axis dose.
Dose 10 cm outside field at dmax
~1% of central axis dose.
Dose 30 cm outside field at dmax
~0.2% of central axis dose.
Intraleaf MLC leakage
~2%.
Interleaf MLC leakage
~3%.
MLC end leakage
~15%.
LINAC head leakage limit
0.1% of central axis dose at 1 m.
Neutron head leakage 10 MV
~0.01%.
Neutron head leakage 18 MV
~0.15%.
6 MeV electron dmax
~1.2 cm.
9 MeV electron dmax
~1.9 cm.
12 MeV electron dmax
~2.1 cm.
20 MeV electron dmax
~2.5 cm.
6 MeV electron surface dose
~80%.
12 MeV electron surface dose
~90%.
20 MeV electron surface dose
~95%.
6 MeV electron x-ray contamination
~1%.
12 MeV electron x-ray contamination
~2%.
20 MeV electron x-ray contamination
~5%.
Lead thickness to stop electrons
Approximately E0(MeV)/2 in mm of lead.
Electron practical range Rp
Approximately E/2 in cm of water.
Electron therapeutic range R80/R90
Approximately E/3 in cm of water.
Electron R50 relation to Energy
R50 ≈ E/2.33 in cm.
Electron penumbra trend with energy
Penumbra generally increases with energy.
Proton range uncertainty common estimate
~3.5% of range plus 1–3 mm.
70 MeV proton range in water
~4 cm.
100 MeV proton range in water
~7.7 cm.
130 MeV proton range in water
12.3 cm
150 MeV proton range in water
~17.7 cm.
190 MeV proton range in water
~23.8cm
200 MeV proton range in water
~26 cm.
230 MeV proton range in water
~33 cm.
250 MeV proton range in water
~38 cm.
Proton calibration protocol often referenced
IAEA TRS-398; AAPM TG-185 for proton reference dosimetry.
LDR dose rate at prescription point
0.5–2 Gy/hr.
HDR dose rate at prescription point
12Gy/hr
Ir-192 half-life
73.8 days.
Ir-192 daily decay rule
~1% per day (~0.94% exact).
HDR source replacement interval
About every 3 months.
approximate dose from Cs-137 to Ir-192
Ir-192 = D(Cs-137) * 0.65
Time 1 m from exposed HDR source to occupational limit
Order of ~1 hour, depending on source strength.
Definition of 1 U, equivalent units
1 cGy*cm2/hr
TG-43 geometry function purpose
Accounts for inverse-square and source geometry effects.
TG-43 radial dose function purpose
Accounts for attenuation and scatter along transverse plane.
TG-43 anisotropy function purpose
Accounts for angular dose variation around source.
I-125 half-life
~59.4 days.
Pd-103 half-life
~17 days.
Cs-131 half-life
~9.7 days.
Cs-137 half-life
~30.2 years.
Co-60 half-life
~5.27 years.
Ra-226 half-life
1600 yrs
Au-198 half-life
2.7 days
I-125 average photon energy
~28 keV.
Pd-103 average photon energy
~21 keV.
Cs-131 average photon energy
~30 keV.
Ir-192 average photon energy
~380 keV.
Au-198 energy
420 keV
Cs-137 energy
660 keV
Ra-226 energy
830 keV
Co-60 photon energies
1250 keV
I-125 patient release limit
less than 1 mR/hr at 1 m.
Pd-103 patient release limit
less than 3 mR/hr at 1 m.
I-131 patient release limit
less than 7 mR/hr at 1 m.
Sealed source leak test removal threshold
More than 185 Bq (0.005 μCi) leakage.
Brachy medical event
Wrong patient, site, or radionuclide; total dose 20%, single fraction 50%
Teletherapy medical event
wrong patient, site; total dose differs by 20% (3+fx); total dose differs by 10% (-3fx)
Point A vs Point B
Point A from tandem, point b from midline
Occupational dose limit
50 mSv/year.
Pregnancy occupational dose limit
5 mSv for entire pregnancy.
Public dose limit
1 mSv/year.
Average occupational exposure in U.S. radiation worker
~2 mSv/year.
Average annual background + medical exposure
~4–5 mSv/year.
Radon annual contribution
~1 mSv/year.