1
New cards

Average photon energy of a bremsstrahlung beam

Approximately one-third of the maximum/nominal energy.

2
New cards

Co-60 photon energies

1.17 and 1.33 MeV.

3
New cards

Average Co-60 photon energy

~1.25 MeV.

4
New cards

Co-60 dmax

~0.5 cm.

5
New cards

6 MV dmax for 10×10 cm²

~1.5 cm.

6
New cards

10 MV dmax for 10×10 cm²

~2.4 cm.

7
New cards

18 MV dmax for 10×10 cm²

~3.3 cm.

8
New cards

PDD at 10 cm for Co-60, 10×10 cm²

~55%.

9
New cards

PDD at 10 cm for 6 MV, 10×10 cm²

~65%.

10
New cards

PDD at 10 cm for 10 MV, 10×10 cm²

~75%.

11
New cards

PDD at 10 cm for 18 MV, 10×10 cm²

~80%.

12
New cards

TMR at 10 cm for 6 MV

~0.78.

13
New cards

TMR at 10 cm for 10 MV

~0.85.

14
New cards

TMR at 10 cm for 18 MV

~0.90.

15
New cards

Approximate attenuation rate for 6 MV

~3% per cm.

16
New cards

Approximate attenuation rate for 10 MV

~2.5% per cm.

17
New cards

Approximate attenuation rate for 18 MV

~2% per cm.

18
New cards

Surface dose for Co-60

~50%.

19
New cards

Surface dose for 6 MV

~25%.

20
New cards

Surface dose for 10 MV

~23%.

21
New cards

Surface dose for 18 MV

~20%.

22
New cards

Scatter at 1 m from phantom

~1/1000 of primary beam.

23
New cards

MV f-factor in air, conversion for Roentgent to cGy

0.876 cGy/R.

24
New cards

MV f-factor in tissue, conversion for Roentgen to cGy

~0.97 cGy/R.

25
New cards

I-125 exposure rate constant

1.45 R cm2/mCi-hr

26
New cards

Ir-192 exposure rate constant

4.69 R cm2/ mCi-hr

27
New cards

Physical wedge out-of-field scatter

~2.5% of central axis dose.

28
New cards

Dynamic wedge out-of-field scatter

~1.0% of central axis dose.

29
New cards

Dose 10 cm outside field at dmax

~1% of central axis dose.

30
New cards

Dose 30 cm outside field at dmax

~0.2% of central axis dose.

31
New cards

Intraleaf MLC leakage

~2%.

32
New cards

Interleaf MLC leakage

~3%.

33
New cards

MLC end leakage

~15%.

34
New cards

LINAC head leakage limit

0.1% of central axis dose at 1 m.

35
New cards

Neutron head leakage 10 MV

~0.01%.

36
New cards

Neutron head leakage 18 MV

~0.15%.

37
New cards

6 MeV electron dmax

~1.2 cm.

38
New cards

9 MeV electron dmax

~1.9 cm.

39
New cards

12 MeV electron dmax

~2.1 cm.

40
New cards

20 MeV electron dmax

~2.5 cm.

41
New cards

6 MeV electron surface dose

~80%.

42
New cards

12 MeV electron surface dose

~90%.

43
New cards

20 MeV electron surface dose

~95%.

44
New cards

6 MeV electron x-ray contamination

~1%.

45
New cards

12 MeV electron x-ray contamination

~2%.

46
New cards

20 MeV electron x-ray contamination

~5%.

47
New cards

Lead thickness to stop electrons

Approximately E0(MeV)/2 in mm of lead.

48
New cards

Electron practical range Rp

Approximately E/2 in cm of water.

49
New cards

Electron therapeutic range R80/R90

Approximately E/3 in cm of water.

50
New cards

Electron R50 relation to Energy

R50 ≈ E/2.33 in cm.

51
New cards

Electron penumbra trend with energy

Penumbra generally increases with energy.

52
New cards

Proton range uncertainty common estimate

~3.5% of range plus 1–3 mm.

53
New cards

70 MeV proton range in water

~4 cm.

54
New cards

100 MeV proton range in water

~7.7 cm.

55
New cards

130 MeV proton range in water

12.3 cm

56
New cards

150 MeV proton range in water

~17.7 cm.

57
New cards

190 MeV proton range in water

~23.8cm

58
New cards

200 MeV proton range in water

~26 cm.

59
New cards

230 MeV proton range in water

~33 cm.

60
New cards

250 MeV proton range in water

~38 cm.

61
New cards

Proton calibration protocol often referenced

IAEA TRS-398; AAPM TG-185 for proton reference dosimetry.

62
New cards

LDR dose rate at prescription point

0.5–2 Gy/hr.

63
New cards

HDR dose rate at prescription point


12Gy/hr

64
New cards

Ir-192 half-life

73.8 days.

65
New cards

Ir-192 daily decay rule

~1% per day (~0.94% exact).

66
New cards

HDR source replacement interval

About every 3 months.

67
New cards

approximate dose from Cs-137 to Ir-192

Ir-192 = D(Cs-137) * 0.65

68
New cards

Time 1 m from exposed HDR source to occupational limit

Order of ~1 hour, depending on source strength.

69
New cards

Definition of 1 U, equivalent units

1 cGy*cm2/hr

70
New cards

TG-43 geometry function purpose

Accounts for inverse-square and source geometry effects.

71
New cards

TG-43 radial dose function purpose

Accounts for attenuation and scatter along transverse plane.

72
New cards

TG-43 anisotropy function purpose

Accounts for angular dose variation around source.

73
New cards

I-125 half-life

~59.4 days.

74
New cards

Pd-103 half-life

~17 days.

75
New cards

Cs-131 half-life

~9.7 days.

76
New cards

Cs-137 half-life

~30.2 years.

77
New cards

Co-60 half-life

~5.27 years.

78
New cards

Ra-226 half-life

1600 yrs

79
New cards

Au-198 half-life

2.7 days

80
New cards

I-125 average photon energy

~28 keV.

81
New cards

Pd-103 average photon energy

~21 keV.

82
New cards

Cs-131 average photon energy

~30 keV.

83
New cards

Ir-192 average photon energy

~380 keV.

84
New cards

Au-198 energy

420 keV

85
New cards

Cs-137 energy

660 keV

86
New cards

Ra-226 energy

830 keV

87
New cards

Co-60 photon energies

1250 keV

88
New cards

I-125 patient release limit

less than 1 mR/hr at 1 m.

89
New cards

Pd-103 patient release limit

less than 3 mR/hr at 1 m.

90
New cards

I-131 patient release limit

less than 7 mR/hr at 1 m.

91
New cards

Sealed source leak test removal threshold

More than 185 Bq (0.005 μCi) leakage.

92
New cards

Brachy medical event

Wrong patient, site, or radionuclide; total dose 20%, single fraction 50%

93
New cards

Teletherapy medical event

wrong patient, site; total dose differs by 20% (3+fx); total dose differs by 10% (-3fx)

94
New cards

Point A vs Point B

Point A from tandem, point b from midline

95
New cards

Occupational dose limit

50 mSv/year.

96
New cards

Pregnancy occupational dose limit

5 mSv for entire pregnancy.

97
New cards

Public dose limit

1 mSv/year.

98
New cards

Average occupational exposure in U.S. radiation worker

~2 mSv/year.

99
New cards

Average annual background + medical exposure

~4–5 mSv/year.

100
New cards

Radon annual contribution

~1 mSv/year.