Scatter radiation
A change in direction of the x-ray photon after interaction with the atoms of the patient
How does scatter affect an image?
Density
Contrast
Resolution
What is increase in density called
Fog
Primary radiation
radiation emitted from the x-ray tube
Remnant (exit) radiation
unabsorbed and scattered x-rays that interact with the IR
it carries the aerial image
Factors that influence the degree of scatter radiation
Field size, density, tube potential (kv) and thickness of the object
Field size
area of the x-ray beam. Field size is altered by the use of a beam limiting device
Thickness of the object
more atoms are present in larger objects which cause more interactions
What does thickness scatter increase
Increase object and increase scatter
What does field size increase
Increases field size and increases scatter
Density
the denser the object, the more scatter results
What does density increase
Increase density and increase scatter
Tube potential (kV)
as kV is increased, the energy of scattered radiation also increases. Thus, scatter radiation has a better chance of reaching the IR
What does tube potential (kv) increase
Increase kv and increase scatter
Beam limitation
Reduces patient exposure
Increase image quality by decreasing scatter to the IR
Increases visibility of detail
50-90% of density on the radiograph is the result of
scatter radiation
Penumbra greater at the
edges of the beam
Types of Beam Limiting Devices
Aperture Diaphragm
Cones (flared and cylinder)
Variable Aperture
***constructed of lead or lead-lined because of its characteristic to attenuate
Aperture Diaphragm
Lead plates with windows cut out in various sizes
Aperture Diaphragms
simplest, least expensive
change for each size
large amount of blur
dedicated units
Cones
Flared
Cylinder
External device placed in the path of the primary beam
Flared Cone
Matches the divergence of beam
Flare is usually larger than beam
Cylinder Cone
Blur is reduced because it truly limits the field size
Circular field
Flared cone
Flared portion of a flare cone does not contribute to beam restriction
Cylinder Cone
Increased beam restriction obtained with the use of a cylinder cone
Disadvantage of cones with circular field
Excessive radiation for larger body parts
VARIABLE APERTURE
Collimator
Automatic Collimator
Positive Beam Limiting Device (PBL) – feature of automatic collimator that assures the field size does not exceed the size of the IR
Positive Beam Limiting Device (PBL)
feature of automatic collimator that assures the field size does not exceed the size of the IR
Automatic Collimator (Housing)
Two sets of horizontal lead shutters
Adjustable- allows for infinite number of field sizes
Upper shutters control stem radiation (off-focus)
Lower shutters reduce blur
Contributes to “added” filtration
Provides a light field
Upper shutters control
stem radiation (off-focus)
Lower shutters reduce
blur
Stem Radiation AKA
off-focus, extra-focal
Stem Radiation
Refers to photons not produced at the focal spot (target) of the anode
tube housing
evaporate tungsten on envelope
Light field (light bulb and mirror) represents the
X-ray field
Cross-hairs represent the
CR
Poor field-to-light congruency may lead to
parts of the anatomy being cut off