1/111
Looks like no tags are added yet.
What is the systematic approach to ventilator assessment?
Assess → Interpret → Intervene → Reassess.
What should be assessed first when approaching a ventilated patient?
Patient condition, oxygenation, airway, circuit, and ventilator settings.
What can a higher total RR than set RR indicate?
Increased respiratory drive, discomfort, or patient-ventilator asynchrony.
What does exhaled Vt significantly lower than delivered Vt suggest?
A leak, incomplete volume delivery, or circuit/airway problem.
PIP increases, Pplat stays the same. What does this suggest?
Increased airway resistance
What is patient-ventilator asynchrony?
Mismatch between patient breathing effort and ventilator delivery.
What can cause patient-ventilator asynchrony?
Inappropriate trigger, flow, timing, cycling, or respiratory drive.
What should be considered before changing ventilator settings?
Patient assessment, ABG, pressures, graphics, trends, and hemodynamics.
What is static compliance (Cstat)?
Vt ÷ (Pplat − PEEP).
What is dynamic compliance (Cdyn)?
Vt ÷ (PIP − PEEP).
What is airway resistance (Raw)?
(PIP − Pplat) ÷ flow in L/s.
What does decreased static compliance indicate?
A stiffer respiratory system requiring more pressure for the same volume.
What does increased airway resistance indicate?
Greater opposition to airflow through the airways or artificial airway.
PIP increases and Pplat increases. What does this suggest?
Decreased compliance
What does a widened PIP-Pplat difference indicate?
Increased resistive pressure during inspiration.
What conditions commonly increase airway resistance?
Bronchospasm, secretions, kinked ETT, biting, or small ETT.
What conditions commonly decrease compliance?
ARDS, pulmonary edema, atelectasis, pneumothorax, or chest-wall restriction.
What is peak inspiratory pressure (PIP)?
Highest airway pressure during inspiration; reflects resistance and elastic pressure.
What is plateau pressure (Pplat)?
Pressure measured during an inspiratory hold with no airflow in a passive patient.
What is the lung-protective plateau pressure target?
Pplat ≤30 cm H₂O.
What is driving pressure?
Pplat − PEEP
What driving pressure target is emphasized in the lung-protection PowerPoint?
Approximately ≤15 cm H₂O.
What is total PEEP?
External PEEP plus intrinsic PEEP.
How is auto-PEEP calculated?
Total PEEP − set external PEEP.
What is auto-PEEP?
Unintentional positive pressure remaining at end expiration from incomplete exhalation.
What waveform finding suggests auto-PEEP?
Expiratory flow fails to return to zero before the next breath.
What causes auto-PEEP?
High RR, large Vt, short expiratory time, or airflow obstruction.
What ventilator adjustment commonly reduces auto-PEEP?
Decrease RR to allow more expiratory time when appropriate.
How can expiratory time be increased?
Decrease RR, shorten inspiratory time, or increase inspiratory flow when appropriate.
Why can increasing RR worsen COPD ventilation?
It shortens expiration, increasing air trapping and auto-PEEP.
How can auto-PEEP affect blood pressure?
Increased intrathoracic pressure can reduce venous return and cause hypotension.
What is the primary benefit of external PEEP?
Maintains alveolar recruitment and improves oxygenation.
What are the potential harms of excessive PEEP?
Overdistention, hypotension, barotrauma, and increased dead space.
What findings help evaluate PEEP effectiveness?
Oxygenation, compliance, driving pressure, plateau pressure, and blood pressure.
What suggests beneficial alveolar recruitment after increasing PEEP?
Improved oxygenation and compliance without harmful pressure or hemodynamic changes.
What suggests excessive PEEP or overdistention?
Worsening compliance, higher driving pressure, hypotension, or impaired ventilation.
What is an important hemodynamic effect of PEEP?
Increased intrathoracic pressure can decrease venous return.
What are the major clinical features of ARDS?
Acute hypoxemia, bilateral opacities, and edema not fully explained by cardiac failure.
What is the recommended initial ARDS tidal volume?
Approximately 6 mL/kg predicted body weight.
What tidal volume range may be used in ARDSNet protocols?
Approximately 4-8 mL/kg PBW with appropriate adjustment.
Why is PBW used instead of actual body weight?
Lung size relates more closely to height and sex than actual weight.
What is the male PBW formula?
50 + 2.3 × (height in inches − 60)
What is the female PBW formula?
45.5 + 2.3 × (height in inches − 60)
Why is permissive hypercapnia used in ARDS?
To avoid excessive tidal volumes and injurious airway pressures.
What is volutrauma?
Lung injury from excessive alveolar stretch or volume.
What is barotrauma?
Lung injury associated with excessive pressure, potentially causing air leaks.
What is atelectrauma?
Injury caused by repetitive alveolar collapse and reopening.
What is biotrauma?
Inflammatory injury triggered by mechanical ventilation.
What is the purpose of prone positioning in ARDS?
Improve ventilation-perfusion matching, recruitment, and oxygenation.
What P/F ratio is highlighted for considering prone positioning?
PaO₂/FiO₂
What prone positioning duration is emphasized?
At least 16 hours per day.
What is the P/F ratio formula?
PaO₂ ÷ FiO₂
What does a lower P/F ratio indicate?
More severe oxygenation impairment.
What P/F ratio corresponds to mild ARDS under standard criteria?
>200 to ≤300 with required PEEP/CPAP support.
What P/F ratio corresponds to moderate ARDS?
>100 to ≤200 with required PEEP support.
What P/F ratio corresponds to severe ARDS?
≤100 with required PEEP support.
What is the estimated desired FiO₂ formula?
Current FiO₂ × desired PaO₂ ÷ current PaO₂.
What is minute ventilation?
Vt × RR
What is alveolar ventilation?
(Vt − dead-space volume) × RR.
What is the relationship between PaCO₂ and alveolar ventilation?
Inverse: increasing alveolar ventilation lowers PaCO₂.
What is the desired RR formula?
Current RR × current PaCO₂ ÷ desired PaCO₂.
What happens to PaCO₂ when alveolar ventilation decreases?
PaCO₂ rises.
What is the classic obstructive ventilator pattern?
High Raw, prolonged expiration, possible auto-PEEP, and increased PIP.
What is the classic restrictive ventilator pattern?
Low compliance, elevated PIP and Pplat, and reduced Vt at a given pressure.
Which diseases are typically obstructive?
COPD and asthma.
Which conditions commonly create restrictive mechanics?
ARDS, pulmonary edema, atelectasis, pneumonia, or pneumothorax.
What does a pressure-time scalar display?
Airway pressure changes over time.
What does a flow-time scalar display?
Inspiratory and expiratory flow over time.
What does a volume-time scalar display?
Delivered and exhaled volume over time.
What does failure of volume to return to baseline suggest?
Possible leak or incomplete exhalation; correlate with flow and exhaled Vt.
What does a pressure-volume loop show?
Relationship between airway pressure and delivered volume.
What does flattening at the upper pressure-volume loop suggest?
Possible alveolar overdistention.
What may the lower inflection region of a pressure-volume loop represent?
Recruitment characteristics of previously unstable lung units.
What happens to the pressure-volume loop when compliance decreases?
Its slope decreases; less volume is delivered for a given pressure change.
What is the first priority with a sudden high-pressure alarm?
Rapidly assess the patient, oxygenation, airway, and ventilator circuit.
What troubleshooting do you need to check for when you are getting a high PIP with stable Pplat?
Check for bronchospasm, secretions, biting, and ETT obstruction. (Increased Raw)
What troubleshooting do you need to check for when you are getting a high PIP with increased Pplat?
Assess for pneumothorax, edema, ARDS, and pressures. (Decreased compliance)
What can cause sudden low exhaled tidal volume?
Circuit leak/disconnection, cuff leak, ETT displacement, or incomplete delivery.
What findings raise concern for pneumothorax?
Sudden deterioration, unilateral decreased breath sounds, hypoxemia, and increased pressures.
What should be assessed after any ventilator intervention?
SpO₂, pressures, exhaled Vt, waveforms, ABG when indicated, and hemodynamics.
What is the first step before changing ventilator settings?
Assess the patient, airway, oxygenation, hemodynamics, and current ventilator data.
What does a high total RR compared with set RR require?
Assess respiratory drive, patient comfort, synchrony, and possible auto-PEEP.
What should be assessed when Pplat exceeds 30 cm H₂O?
Check Vt, PEEP, compliance, and patient effort; reduce Vt per ARDSNet when appropriate.
What is the ARDSNet response to Pplat above 30?
Decrease Vt in 1 mL/kg PBW steps, down to 4 mL/kg if necessary.
What must be checked before calculating airway resistance?
Use inspiratory flow in L/s and passive, constant-flow conditions.
What does falling Cstat indicate?
The respiratory system is becoming stiffer; investigate loss of recruitment or worsening disease.
What should be checked before increasing RR for hypercapnia?
Expiratory flow, auto-PEEP, obstruction, and whether sufficient exhalation time remains.
What changes can help reduce auto-PEEP?
Reduce RR, shorten inspiratory time if appropriate, and treat airflow obstruction.
What should be reassessed after reducing auto-PEEP?
Expiratory flow, total PEEP, blood pressure, PaCO₂, and overall ventilation.
What is the priority when bronchospasm increases resistance?
Assess airflow and treat bronchospasm; also check secretions and ETT patency.
What finding shows incomplete expiration on a flow-time scalar?
Expiratory flow remains below zero when the next inspiration begins.
What should be investigated if the patient is triggering excessive breaths?
Pain, anxiety, hypoxemia, hypercapnia, inadequate support, and asynchrony.
Why might FiO₂ alone fail to correct ARDS hypoxemia?
Collapsed or fluid-filled alveoli create shunt; recruitment may be needed.
What should be evaluated after increasing PEEP?
PaO₂/SpO₂, compliance, Pplat, driving pressure, cardiac output, and blood pressure.
What suggests PEEP is recruiting alveoli effectively?
Better oxygenation and compliance without excessive pressures or hemodynamic compromise.
What suggests PEEP has become excessive?
Compliance falls, driving pressure rises, or hypotension/cardiac output deterioration occurs.
What does the professor's PEEP table show at PEEP 18?
Peak Cstat of 45 mL/cm H₂O and highest cardiac output of 6.5 L/min in that table.
What happens at PEEP 22 in the professor's table?
PaO₂ rises to 85, but Cstat falls to 33 and cardiac output falls to 4.9 L/min.
Why is the highest PaO₂ not always the optimal PEEP?
Oxygenation improvement may occur alongside overdistention and reduced cardiac output.
What is the stated PEEP titration step in the lung-protection slides?
Increase PEEP by 2-4 cm H₂O and observe the response for 10 minutes.