  1. Which of the following is likely to result in the highest efficacy of medical instrument cleaning:

a) Use of a central reprocessing area for all instrument cleaning b) Local instrument reprocessing within the area of care c) Use of an acidic pH cleaner d) Allowing the instrument to dry after use and before cleaning

a) Use of a central reprocessing area for all instrument cleaning

2
New cards
  1. Inadequate sterilization of surgical instruments and tubing may result in which of the following adverse events after cataract surgery?

a) Toxic Shock Syndrome b) Posterior Segment Syndrome c) Anterior Syndrome d) Toxic Anterior Segment Syndrome

d) Toxic Anterior Segment Syndrome

3
New cards
  1. One disadvantage of liquid sterilization is:

a) Liquid sterilants are highly toxic and items must be aerated before use b) It is a high heat process so it may not be used on heat-labile items c) It is not an appropriate process for critical items d) Items cannot be wrapped during the sterilization process so sterility cannot be maintained during storage

d) Items cannot be wrapped during the sterilization process so sterility cannot be maintained during storage

4
New cards
  1. During infection control rounds in the emergency department, the manager tells you that they will be doing endovaginal ultrasounds soon and would like to use probe covers to speed up the cleaning process. You inform the manager that:

a) Using a probe cover will eliminate the need to high level disinfect the probe b) Using a probe cover will reduce the amount of time needed to clean the probe c) Using a probe cover is acceptable, but high level disinfection is still required d) Using a probe cover is acceptable, but this device requires steam sterilization

c) Using a probe cover is acceptable, but high level disinfection is still required

5
New cards
  1. Which of the following practices have studies shown is most likely to result in improved infection prevention?

a) Strictly adhering to the EPA-registered product label contact time for low level disinfection of environmental surfaces b) Utilizing a one minute contact time for low level disinfection of environmental surfaces regardless of the EPA-registered product label contact time c) Thoroughly cleaning and disinfecting all surfaces in a room that have potentially come into contact with hands d) Thoroughly cleaning and disinfecting all high touch surfaces in a room

c) Thoroughly cleaning and disinfecting all surfaces in a room that have potentially come into contact with hands

6
New cards
  1. The director of surgical services has received a call from a neurosurgeon who would like to schedule a brain biopsy on a person suspected of having Creutzfeldt-Jakob Disease (CJD). The staff have expressed concern cleaning and sterilizing the surgical instruments. You response is:

a) Surgical instruments are very expensive and should be cleaned and sterilized after the procedure b) The recommendations are unclear as to how to clean and sterilize instruments c) There are no special requirements for cleaning and sterilization of surgical instruments d) The instrument used in these cases require special processing

d) The instrument used in these cases require special processing

7
New cards
  1. You have been asked to do an in-service for Environmental Services on cleaning procedures. Which of the following is the best practice for cleaning a patient room?

(1) Clean the patient zone first, and then the perimeter of the room (2) Clean the perimeter of the room first, and then the patient zone (3) Clean items that are low to the floor first and then work your way up to higher items (4) Clean items that are higher up first, and then work your way down to lower items

a) 1, 3 b) 2, 3 c) 1, 4 d) 2, 4

d) 2, 4

8
New cards
  1. The purpose of cleaning medical devices before sterilization or high level disinfection is to:

a) Reduce bio-burden b) Add an additional step to the process c) Replace the sterilization process d) Increase the amount of time it takes to clean an endoscope

a) Reduce bio-burden

9
New cards
  1. The purpose of including a biological indicator (BI) in an autoclave load is:

a) To determine whether the items being autoclaved are properly sterilized, which is indicated by a positive BI result b) To determine whether the items being autoclaved are properly sterilized, which is indicated by a negative BI result c) To determine whether the items being autoclaved are properly cleaned, which is indicated by a positive BI result d) To determine whether the items being autoclaved are properly cleaned, which is indicated by a negative BI result

b) To determine whether the items being autoclaved are properly sterilized, which is indicated by a negative BI result

10
New cards
  1. In an effort to reduce cost and decrease turn around time, the endoscopy manager has asked if the enzymatic detergent can be used for multiple patients. Your response is:

a) It is acceptable to reuse the detergent b) The enzymatic detergent must be discarded after each use c) There are no recommendations on reuse of the detergent d) It is not necessary to use an enzymatic detergent for endoscopes

b) The enzymatic detergent must be discarded after each use

11
New cards
  1. An OR team was using a unique surgical instrument on a patient and the instrument was accidentally dropped on the floor. The team needs to reprocess it as quickly as possible to finish the surgery and would like to use Immediate-Use Steam Sterilization (IUSS). Which of the following is the best recommendation for this?

a) The team must clean and inspect the instrument before proceeding with IUSS b) The team can proceed with IUSS without prior cleaning and inspection of the instrument since it will be used on the same patient c) There is no need to use IUSS as the team can simply soak the instrument in ten percent bleach for 10 minutes d) There is no situation in which IUSS is acceptable

a) The team must clean and inspect the instrument before proceeding with IUSS

12
New cards
  1. Types of tissue considered at high risk for CJD include:

(1) Heart (2) Blood (3) Eye (4) Brain

a) 1, 4 b) 1, 3 c) 3, 4 d) 2, 4

c) 3, 4

13
New cards
  1. Which of the following describes the correct storage to maintain sterility of reprocessed items?

a) At least 12 inches below the ceiling, at least six inches above the floor, at least one inch from the outside wall, and on a rack with a solid bottom b) At least 18 inches below the ceiling, at least six inches above the floor, at least one inch from the outside wall, and on a rack with a solid bottom c) At least 18 inches below the ceiling, at least eight inches above the floor, at least two inches from the outside wall, and on a rack with a solid bottom d) At least 18 inches below the ceiling, at least eight inches above the floor, at least one inch from the outside wall, and on a rack with a solid bottom

c) At least 18 inches below the ceiling, at least eight inches above the floor, at least two inches from the outside wall, and on a rack with a solid bottom

14
New cards
  1. Prior to opening a sterile package, the end user should inspect the package for:

(1) Tears (2) Moisture (3) Date of manufacture (4) The name of the person who packaged the kit

a) 1, 2 b) 2, 4 c) 3, 4 d) 1, 3

a) 1, 2

15
New cards
  1. Ultrasonic cleaners are used for:

a) Removing gross soil from an instrument b) High-level disinfection of instruments c) Sterilization of instruments d) Fine cleaning of instruments

d) Fine cleaning of instruments

16
New cards
  1. Aseptic technique is defined as:

a) No touch technique b) A process used in the operating room c) An absence of organisms d) The process of keeping away disease producing organisms

d) The process of keeping away disease producing organisms

17
New cards
  1. Which of the following indicators is most accurate for assessing whether surgical instruments have been processed correctly in a steam sterilization cycle:

a) Autoclave tape b) Chemical indicator c) Biological indicator d) The cycle indicator on the steam sterilizer

c) Biological indicator

18
New cards
  1. You have recently accepted a position at a free standing endoscopy center. After reviewing the policy and procedures, you discover that the last time staff competency for cleaning and disinfecting endoscopes was completed was three years ago. Your next step is:

(1) Schedule a meeting with the facility administrator to discuss your findings (2) Do not worry about it because this happened before you worked there (3) Develop competencies for cleaning and disinfection of endoscopes (4) Assume that it is not a problem because there have been no outbreaks at the center

a) 1, 3 b) 2, 4 c) 3, 4 d) 2, 3

a) 1, 3

19
New cards
  1. A technician in the OR has reported an infestation of fruit flies in the sterile instrument storage room. The sterile packs were temporarily removed and pest management was called in to eradicate the flies. The OR staff want to know whether they can use the sterile packs. Which of the following is the best response?

a) The OR staff can use the packs without any reprocessing because there are no flies visible on the outside of the packs b) The OR staff can use the packs without any reprocessing if they open a test pack and don't find any flies inside it c) The OR staff don't need to repack the items but they do need to run the packs through the sterilizer again d) They need to completely reprocess all of the items that were in the affected room

d) They need to completely reprocess all of the items that were in the affected room

20
New cards
  1. Which of the following items can generally be reprocessed by only using low-level disinfectant:

a) Blood pressure cuffs b) Anesthesia equipment c) Bronchoscopes d) Surgical instruments

a) Blood pressure cuffs

21
New cards
  1. Your endoscopy unit has experienced an outbreak associated with bronchoscopes. The CDC recommends that outbreaks in this setting be reported to which agencies:

(1) The Joint Commission (2) Centers for Medicaid/Medicare Services (3) The Food and Drug Administration (4) The bronchoscope manufacturer

a) 1, 2 b) 2, 3 c) 3, 4 d) 2, 4

c) 3, 4

22
New cards
  1. According to AAMI ST79, which of the following are among the recommendations for testing of mechanical cleaning equipment in order to verify adequate cleaning?

(1) Verification should be carried out monthly (2) Verification should be carried out upon installation (3) Verification should be carried out after major repairs (4) Verification should be carried out when changing cleaning chemistry

a) 1, 2, 3, 4 b) 1, 2, 4 c) 1, 2, 3 d) 2, 3, 4

d) 2, 3, 4

23
New cards
  1. A biological indicator that was included in a load of surgical implantables has indicated a failure. You are able to deduce that this was due to operator error, as this one load was not run on the correct cycle. Which of the following actions should you take?

a) No action is needed as the implantables did go through a sterilizer cycle b) The load with the positive indicator must be reprocessed c) A recall must be initiated; any loads of implantables run by the technician who ran this load must be reprocessed d) A recall must be initiated; any loads of implantables run by any technician must be reprocessed

b) The load with the positive indicator must be reprocessed

24
New cards
  1. Which of the following are used for sterilization of medical instruments?

(1) Gravity-displacement steam sterilizer (2) Pasteurizer (3) Ethylene oxide sterilizer (4) Ultrasonic cleaner

a) 1, 2 b) 1, 3 c) 1, 2, 3 d) 1, 2, 3, 4

b) 1, 3

25
New cards
  1. The endoscopy manager has asked for your guidance in finding a location to store sterile supplies. Your response is:

a) The room must be positive pressure with a temperature of no greater than 65°F, with a relative humidity of less than 70 percent b) The room must be positive pressure with a temperature of approximately no greater than 75°F, with a relative humidity not to exceed 70 percent c) The room must be negative pressure with a temperature of 65°F, and a relative humidity of greater than 70 percent d) There are no specific requirements for storing sterile supplies

b) The room must be positive pressure with a temperature of approximately no greater than 75°F, with a relative humidity not to exceed 70 perc