Clinical Chemistry: Test Report Messages, Quality Control, and Specimen Collection

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/35

flashcard set

Earn XP

Description and Tags

Comprehensive vocabulary flashcards covering clinical laboratory error messages, Westgard quality control rules, calibration criteria, specimen rejection criteria, and tube additive requirements.

36 Terms

1
New cards

ABSORBANCE (Dimension Error Message)

An error indicating that the result is above the method assay range and cannot be calculated. The technician must check the method insert sheet to determine if the sample can be diluted, manually dilute at the lowest possible ratio, and enter the dilution factor into the Dimension. Do not report to the physician immediately.

2
New cards

HI (Dimension Error Message)

An error indicating that the test result is higher than the normal reference interval. No dilution is required if this is the only error, and it must be reported to the physician immediately if it is the sole error.

3
New cards

hp (Dimension Error Message)

An error indicating that the result is higher than the panic interval (critical value). The technician must follow laboratory critical value procedures, call the physician, document the date/time and contact person, obtain read-back verification, and provide immediate notification.

4
New cards

ASSAY RANGE (Dimension Error Message)

An error indicating that the result is above or below the method assay range. For low/negative results, perform a 50%50\% recovery of a known standard or QC material to confirm absence of activity; for high results, dilute the sample. Do not report to the physician immediately.

5
New cards

PROCESS ERROR

An instrument error indicating that the analyzer could not determine a result. The technician must press Alt/M and follow on-screen instructions, or inspect the error log screen to assess the specific failure. It is not reported to the physician because no valid result is available.

6
New cards

Abnl Assay (Abnormal Assay)

An error appearing in the reference interval column requiring further action. The operator runs a QC sample for the method: if no error occurs, call technical assistance; if the error recurs, change the reagent; if the error recurs twice, call technical assistance. Do not report to the physician immediately.

7
New cards

Hemolysis (HIL Interference)

Interference caused by red blood cell (RBC) lysis, which releases intracellular potassium into the extracellular fluid and falsely elevates potassium and other analyte levels. Requires following lab procedure and requesting specimen recollection.

8
New cards
<p>Hemolyzed vs. Normal/Icteric Specimen</p>

Hemolyzed vs. Normal/Icteric Specimen

A visual comparison of blood collection tubes showing a hemolyzed sample (left tube with red serum caused by RBC lysis and intracellular potassium release) versus an unhemolyzed/icteric sample (right tube with yellow serum).

9
New cards

Icterus (HIL Interference)

Interference characterized by dark yellow or amber-brown serum due to hyperbilirubinemia. Requires following the lab HIL procedure and typically appending a comment regarding icteric serum prior to reporting.

10
New cards

Lipemia (HIL Interference)

Interference characterized by a turbid or milky serum/plasma appearance caused by elevated lipoproteins or contamination. Addressed by following HIL procedure: recollect if contaminated, or airfuge/ultracentrifuge if due to a patient condition.

11
New cards

Linear Calibration Limits

An acceptable slope range of 0.97−1.030.97 - 1.03 and an acceptable intercept of 00 or close enough to 00 to be clinically insignificant (e.g., an intercept of 2.32.3 is acceptable for cholesterol with reference range <200 mg/dL< 200\,\text{mg/dL}, but unacceptable for calcium with reference range 8.6−10.2 mmol/L8.6 - 10.2\,\text{mmol/L}).

12
New cards

Non-linear Calibration Limits

An acceptable slope range of 0.95−1.050.95 - 1.05 with an intercept of 00 or clinically insignificant. Calibration curve adjustments are governed by coefficients C2C_2, C3C_3, and C4C_4.

13
New cards

Enzyme Calibration Limits

An acceptable slope range of 0.90−1.100.90 - 1.10 with an acceptable intercept of 00 or clinically insignificant.

14
New cards

C0C_0 Coefficient

The calibration coefficient responsible for adjusting the y-intercept of the calibration curve.

15
New cards

C1C_1 Coefficient

The calibration coefficient responsible for adjusting the slope of the calibration curve (mnemonic: the digit '1' resembles a straight line).

16
New cards

12s1_{2\text{s}} Westgard Rule

A quality control rule where one control value exceeds the mean by more than 2 SD2\,\text{SD}. It serves as a warning rule only and does not identify an active error type.

17
New cards

13s1_{3\text{s}} Westgard Rule

A quality control rejection rule triggered when one control observation exceeds 3 SD3\,\text{SD} from the mean; indicates random error.

18
New cards

22s2_{2\text{s}} Westgard Rule

A quality control rejection rule triggered when two consecutive control observations exceed 2 SD2\,\text{SD} on the same side of the mean; indicates systematic error.

19
New cards

R4s\text{R}_{4\text{s}} Westgard Rule

A quality control rejection rule triggered when the difference between two controls within the same run exceeds 4 SD4\,\text{SD}; indicates random error.

20
New cards

41s4_{1\text{s}} Westgard Rule

A quality control rejection rule triggered when four consecutive control observations exceed 1 SD1\,\text{SD} on the same side of the mean; indicates systematic error.

21
New cards

10x10\text{x} Westgard Rule

A quality control rejection rule triggered when ten consecutive control observations fall on the same side of the mean; indicates systematic error.

22
New cards

Shift (QC)

A quality control failure pattern defined by an abrupt or sudden movement of control values to a new mean; indicates systematic error.

23
New cards

Trend (QC)

A quality control failure pattern defined by a gradual, progressive drift of control values away from the mean; indicates systematic error.

24
New cards

Dilution Factor Programming

The procedure of inputting the manual dilution factor into the Dimension analyzer during patient entry, allowing the analyzer to automatically calculate and report the true final concentration.

25
New cards

QNS (Quantity Not Sufficient)

A specimen rejection condition caused by an insufficient volume of blood or serum to perform the requested test, requiring specimen recollection.

26
New cards

Short Blue Top Tube

An underfilled light blue collection tube rejected because an improper blood-to-citrate ratio alters coagulation tests (such as PT/INR and PTT), requiring recollection.

27
New cards

Unlabeled or Mislabeled Specimen

The most serious reason for laboratory specimen rejection due to patient identification and safety concerns, requiring specimen recollection.

28
New cards

Lavender (Purple) Top Tube

A blood collection tube containing EDTA additive, utilized for complete blood counts (CBC) and general hematology tests.

29
New cards

Light Blue Top Tube

A blood collection tube containing sodium citrate additive, utilized for coagulation testing including PT/INR and PTT.

30
New cards

SST Gold / Tiger Top Tube

A serum separator tube containing clot activator and gel, utilized for most routine chemistry tests.

31
New cards

Light Green Top Tube

A plasma separator tube containing lithium heparin additive, utilized for STAT chemistry tests.

32
New cards

Green or Light Green Top Tube

A collection tube containing heparin additive, utilized for measuring electrolytes.

33
New cards

Pink Top Tube

A collection tube containing EDTA additive, designated for blood bank testing, including type and screen.

34
New cards

Gray Top Tube

A collection tube containing sodium fluoride and potassium oxalate additives, utilized for glucose testing.

35
New cards

Red or SST Tube (Toxicology)

A collection tube containing either no additive or a clot activator, utilized for drug levels and toxicology screens.

36
New cards

Blood Culture Bottles

Specialized collection containers formulated with special culture media, used for the detection and isolation of pathogens from blood cultures.