Immune Modulation, Monitoring and Metabolism

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/60

encourage image

There's no tags or description

Looks like no tags are added yet.

61 Terms

1
New cards

What are the 2 golden rules of cytometry?

  1. What you see is what you get

  2. Garbage in, garbage out


2
New cards

Name some important factors that influence sample preparation

  • Sampling methods: surgery, freezing and thawing/cell culture harvest

  • Sample logistics/storage: duration/temperature/storage solution

  • Preparation protocol: buffers/reagents/enzymes/mechanical disruption


3
New cards

Name some controls used for flow cytometry

  • Beads control —> machine optimization

  • Autofluorescence controls —> unstained ctrl

  • Compensation controls —> single stains

  • Isotype control —> Irrelevant target, to test NSB

  • Gating controls —> FMO’s

  • Experimental controls —> untreated ctrl / positive ctrl

  • Reference control —> batch to batch variation


4
New cards

What are some non-antibody based applications in flow cytometry?

  • Viability stainings (e.g. LiveDead)

  • Proliferation assays (e.g. CFSE or CellTrace Violet)

  • Apoptosis detection (e.g. Annexin V staining)


5
New cards

What is excitation and emission?

Emission spectra are probabilistic EM wavelengths that will be emitted upon excitation of a given fluorescent molecule

<p><span style="background-color: inherit; color: windowtext;">Emission spectra are probabilistic EM wavelengths that will be emitted upon excitation of a given fluorescent molecule</span></p>
6
New cards

What are the 4 flow cytometer components?

  1. Fluidics

  2. Optics

  3. Electronics

  4. Data


7
New cards

What does a laser do?


Add energy to fluorescent molecules. This excites their outer shell electrons into a higher energy state which emit (predictable) wavelengths of light.
 

8
New cards

What is ‘time correction‘ in flow cytometry?

When multiple lasers are used, the cytometer will fix the time delay between the 2 pulses so you know which pulse belongs to which cell. How much time the particle need to pass all detectors: distance form laser to laser. 

9
New cards

What is sheath fluid used for in flow cytometry?

To narrow the sample to one small line and create a laminar flow.

<p>To narrow the sample to one small line and create a laminar flow.</p>
10
New cards

What is acoustic focussing in flow cytometry?

With laminar flow cells will not be perfectly in the middle, with this method the cell can be perfectly aligned in the centre using an acoustic field.

11
New cards

What are optic filters in flow cytometry?

Optic filters can let certain wavelengths of lasers pass. You have short pass, band pass and long pass filters.

<p>Optic filters can let certain wavelengths of lasers pass. You have short pass, band pass and long pass filters.</p>
12
New cards

What do detectors do in flow cytometry?

They convert the fluorescent flashes (photons) into measurable electronic signals (electrons).

13
New cards

What is an Avalance Photon Diode (APDs)?

  • Light hits a semiconductor material and generates electrons. 

  • These electrons are accelerated, causing an avalanche effect that multiplies the electrical signal. 

  • Generally more compact and can offer better quantum efficiency than PMTs at certain wavelengths. 


14
New cards

What are Photomultilpiers Tubes (PMTs)

  • Light hits a photocathode, generating electrons. 

  • These electrons are amplified through a series of dynodes. 

  • This produces a measurable electrical signal. 

  • Extremely sensitive, making PMTs particularly useful for detecting weak fluorescence signals


15
New cards

What is the difference between PMTs and APDs

Both convert photons into electrical signals, but PMTs amplify electrons using dynodes, whereas APDs amplify them through an avalanche effect in a semiconductor. 

16
New cards

How does signal generation work in flow cytometry?

The more fluorescence, the higher the intensity of the emission peak 

<p><span style="background-color: inherit; color: windowtext;">The more fluorescence, the higher the intensity of the emission peak</span><span style="line-height: 20.7px; color: windowtext;">&nbsp;</span></p>
17
New cards

What is the difference between spectral and conventional cytometry?

Conventional cytometry relies on one detector per fluorochrome which could need compensation, while spectral cytometry uses a spectral signature from many detectors combined with unmixing algorithms.

18
New cards

What is imaging flow cytometry?

  • Each cell is imaged as it flows through the machine 

  • This way allows for both of best worlds, with speed and visual details 


19
New cards

What is FACS?

  • Fluorescence-Activated Cell Sorting 

  • Allows researchers to separate up to 6 different populations at high speed based on their fluorescence characteristics.  


20
New cards

What is the refractive index?

The refractive index is a measure of how much light bends, or refracts, as it passes through a material. It is defined as the ratio of the speed of light in vacuum to the speed of light in the material. (n = c / v)

21
New cards

What is chromatic aberration?

A distortion in an optical image that occurs when different wavelengths of light do not converge at the same point after passing through a lens, resulting in coloured fringes around objects.

22
New cards

What is spherical aberration?

An optical distortion that occurs when light rays striking a lens near its edge focus at different points than those striking near the center, leading to a blurred or distorted image.

23
New cards

What is aperture?

Aperture refers to the opening in a lens through which light passes. It controls the amount of light entering the camera or optical system, affecting resolution, contrast and depth of field. A higher magnification (low NA) is not always better.

24
New cards

What 3 light sources are used most in microscopes?

  1. Lamp —> broad spectrum, no collimation (rays of light are made parallel)

  2. LEDs —> narrower spectrum, difficult to collimate and focus

  3. Lasers —> very narrow spectrum, very specific collimated beam


25
New cards

What is a Charge Coupled Device (CCD)?

  • Light hits the pixels and generates electrons. 

  • The electrons are transferred from pixel to pixel across the detector. 

  • Eventually, they reach an output node, where the charge is converted into voltage. 

  • The signal is then used to construct an image. 


26
New cards

What is a Complementary Metal-Oxide Conductor (CMOC)?

  • Light also generates electrons in each pixel. 

  • However, each pixel converts its own charge into voltage. 

  • This means the signal doesn't need to travel across the detector before conversion. 


27
New cards

What is Stoke’s Shift?


The difference between the wavelength a fluorochrome is excited with (shorter wavelength, higher energy) and the emission wavelength (longer wavelength, lower energy)
. Allows separation of emission and excitation

28
New cards

What is widefield microscopy


Illuminates the entire sample at once, uses cameras (CCD or CMOS) for detection, images can be blurry and outside light can get in
 

29
New cards

What is confocal microscopy 


Uses a focused laser to illuminate one small point at a time, uses PMTs or APDs for detection, sharper images with better contrast and outside light is blocked out
 

30
New cards

What is Numerical Aperture (NA)?

Numerical Aperture (NA) is a measure of a lens's ability to gather light and resolve fine detail. A higher NA means better resolution and brighter images.

31
New cards

What is fluorophore crosstalk, and which 2 types are there?

Fluorophore crosstalk is the unwanted signal from one fluorophore being detected in the channel meant for another fluorophore. You have;

  1. Excitation crosstalk: multiple fluorophores are excited by a single excitation wavelength 

  2. Emission crosstalk: emission wavelengths of multiple fluorophores are detected in the same channel 


32
New cards

What is multispectral imaging?

It captures images at several specific wavelength bands to record spectral information about a sample 

33
New cards

What is compensation?

It refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome from all detectors except the one devoted to measuring that dye . (Calculate what % of correction is needed to make sure the median, of the dye not present, of both negative and positive populations is the same)

<p><span style="background-color: inherit; color: windowtext;">It&nbsp;refers to the process of correcting&nbsp;for fluorescence spillover, i.e.,&nbsp;removing the signal of any given&nbsp;fluorochrome from all detectors&nbsp;except the one devoted to&nbsp;measuring that dye</span><span style="color: windowtext;">&nbsp;. (</span><span style="background-color: inherit; color: windowtext;"><em>Calculate what % of correction is needed to make sure the median, of the dye not present, of both negative and positive populations is the same</em></span><span style="color: windowtext;">)</span></p>
34
New cards

What types of spillover are there?

  1. Adjacent spectra —> emissions are close in the spectrum

  2. Tandems and their bases —> fluorochromes that result from multiple fluorochromes together

  3. Cross-laser excitation —> similar excitation but there are different emissions 


35
New cards

What is a tandem dye?

A tandem dye is a fluorochrome made by linking two different fluorophores together, a donor and an acceptor, so that energy absorbed by the donor is transferred to the acceptor, which then emits light at a longer wavelength.

36
New cards

What are the 3 rules for compensation? (ON EXAM)

  1. Use the same fluorophores as in your experiment --> The fluorochrome in the single stain control must have exactly the same spectral characteristics as the one used in the experiment 

    • Especially for tandem dyes

  2. Controls must be sufficiently bright --> The positive population in the single stain control must be as bright or brighter than the positive control cells in the experiment 

    • If the compensation is calculated on a dim population, brightly stained cells in the mix can be easily under- or overcompensated. E.g. if the values fall 'outside', they become unreliable 

  3. Use single stain controls --> Within a single stain control, the negative population should have the same autofluorescence as the positive control 

    • The compensation uses the slope of the line between the positive and negative for calculation, so they must match. When you compensate, the medians must line up. If one has higher autofluorescence the results become unreliable 

    • The autofluorescence in the beads are almost always identical, that is why you can use beads as SS instead of cells e.g. 


37
New cards

What is an important rule about single stain controls?


Whatever you do to your SS controls (whether cells or beads), you should do to the experimental samples and the other way around
 

38
New cards

What is an ‘N x N‘ plot used for?

Troubleshooting, it is a rapid mean to identify problems. You look for abnormal compensation in the gatings.

<p>Troubleshooting, it is a<span style="background-color: inherit; color: windowtext;"> rapid mean to identify problems</span><span style="line-height: 20.7px; color: windowtext;">. </span><span style="background-color: inherit; color: windowtext;">You look for abnormal compensation in the gatings</span><span style="line-height: 20.7px; color: windowtext;">.</span></p>
39
New cards

What is negative and super negative in flow cytometry?

  • Negative: Cells that do not express the marker of interest, so they show only background/autofluorescence.

  • Super-negative: Cells with an even lower fluorescence signal than the normal negative population.


40
New cards

Which antibody is able to activate/recruit neutrophils?

IgA

41
New cards

What is trogocytosis?

Cell death by neutrophils that tear the tumour cell apart by pulling on the tumour cells

42
New cards
43
New cards
44
New cards
45
New cards
46
New cards
47
New cards
48
New cards
49
New cards
50
New cards
51
New cards
52
New cards
53
New cards
54
New cards
55
New cards
56
New cards
57
New cards
58
New cards
59
New cards
60
New cards
61
New cards