instrumentation chat part 3 collimators

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/31

encourage image

There's no tags or description

Looks like no tags are added yet.

32 Terms

1
New cards

two functions of collimator

  1. defines field of view

  2. rejects unwanted photons


2
New cards

what is collimator made of

  • lead

  • because of high atomic number and density


3
New cards

septa

  • walls of lead between the holes

  • absorb off angle photons

  • only photons traveling in acceptable directions pass through the holes


4
New cards

spacial resolution

ability to distinguish two nearby radioactive points

5
New cards

sensitivitiy

number of phtons detected per unit of time for a given activity

6
New cards

how are sensitivity and spatial resolution related

inversely

7
New cards

high resolution collimator

  • smaller holes

  • better resolution

  • lower sensitivity

  • longer acquisition time to obtain same counts


8
New cards

high sensititvity collimator

  • larger holes

  • lower resolution

  • higher sensitivity

  • shorter acquisition time to obtian the same counts


9
New cards

what does smaller hole diameter to

  • improve spatial resolution

  • decrease sensitivity


10
New cards

what does larger hole diameter do

  • decrease spatial resolution

  • increase sensitivity


11
New cards

what does longer holes do

  • improves spatial resolution

  • decreases sensitivity slightly


12
New cards

what does shorter holes do

  • decreases spatial resolution

  • generally increases sensitivity


13
New cards

what does thicker septa do

  • helps prevent prenetration

  • decreases open area


14
New cards

septal penetration

  • occurs when gamma photons pass through the lead septa instead of bieng absorbed

  • happens when photons have enough energy to penetrate the lead


15
New cards

consequences of septal penetration

  • poor spatial resolution

  • blurring

  • star artifacts

  • poor image definition


16
New cards

higher energy photons require…

thicker septa to absorb them

17
New cards

example for high energy collimator isotope

I 131

18
New cards

example for medium energy collimator isotope

Ga - 67 and In 111

19
New cards

example for low energy collimator isotope

Xe 133 and Tc 99

20
New cards

what is septal thickness for low energy collimator

thin

21
New cards

what is septal thickness for medium energy collimator

intermediuate

22
New cards

what is septal thickness for high energy collimator

thick

23
New cards

parallel hole collimator

  • holes run parallel

  • no magnification

  • image remains approximately the same size as the object

  • most commonly used


24
New cards

converging collimator

  • holes angle toward focal point

  • produces a magnified image

  • used for small organs

  • increased geometric efficiency


25
New cards

diverging collimator

  • channels diverge outward

  • produces smallar/minified image

  • increases field of view

  • useful for organs larger than detectors normal FOV (kidney)


26
New cards

fan beam collimator

  • type of convering collimator

  • channels converge to a line parallel to cameras axis of roation

  • channels form a wedge like geometry


27
New cards

pinhole collimator

  • cone shaped with aperature facing patient

  • magnifies the image

  • image is inverted

  • used for small organs such as the thyroid


28
New cards

why is pinhole collimator useful for imaging small organs

  • because they produce magnified images with excellent resolution


29
New cards

characteristics of pinhole colimator

  • cone shpaed

  • apex faces patient

  • thick lead sides

  • removable tungsten aperature

  • poor sensitivity

  • image is reverse

  • magnification changes with distance from aperature


30
New cards

parallax errors

  • occur in pinhole collimators

  • occur because the apparent location of an object changes with the viewing angle


31
New cards

How does increasing the distance between a patient and parallel-hole collimator affect resolution and sensitivity?

  • spatial resolution decreases

  • image becomes blurrier

  • senstivity/count rate changes very slowly with distance


32
New cards

Predict how changing from an all-purpose to a high-resolution collimator affects the time needed to reach the same counts.

imaging time increases

  • high resolution has smaller holes so fewer photons reach crystal and decreases sensitivity