two functions of collimator
defines field of view
rejects unwanted photons
what is collimator made of
lead
because of high atomic number and density
septa
walls of lead between the holes
absorb off angle photons
only photons traveling in acceptable directions pass through the holes
spacial resolution
ability to distinguish two nearby radioactive points
sensitivitiy
number of phtons detected per unit of time for a given activity
how are sensitivity and spatial resolution related
inversely
high resolution collimator
smaller holes
better resolution
lower sensitivity
longer acquisition time to obtain same counts
high sensititvity collimator
larger holes
lower resolution
higher sensitivity
shorter acquisition time to obtian the same counts
what does smaller hole diameter to
improve spatial resolution
decrease sensitivity
what does larger hole diameter do
decrease spatial resolution
increase sensitivity
what does longer holes do
improves spatial resolution
decreases sensitivity slightly
what does shorter holes do
decreases spatial resolution
generally increases sensitivity
what does thicker septa do
helps prevent prenetration
decreases open area
septal penetration
occurs when gamma photons pass through the lead septa instead of bieng absorbed
happens when photons have enough energy to penetrate the lead
consequences of septal penetration
poor spatial resolution
blurring
star artifacts
poor image definition
higher energy photons requireā¦
thicker septa to absorb them
example for high energy collimator isotope
I 131
example for medium energy collimator isotope
Ga - 67 and In 111
example for low energy collimator isotope
Xe 133 and Tc 99
what is septal thickness for low energy collimator
thin
what is septal thickness for medium energy collimator
intermediuate
what is septal thickness for high energy collimator
thick
parallel hole collimator
holes run parallel
no magnification
image remains approximately the same size as the object
most commonly used
converging collimator
holes angle toward focal point
produces a magnified image
used for small organs
increased geometric efficiency
diverging collimator
channels diverge outward
produces smallar/minified image
increases field of view
useful for organs larger than detectors normal FOV (kidney)
fan beam collimator
type of convering collimator
channels converge to a line parallel to cameras axis of roation
channels form a wedge like geometry
pinhole collimator
cone shaped with aperature facing patient
magnifies the image
image is inverted
used for small organs such as the thyroid
why is pinhole collimator useful for imaging small organs
because they produce magnified images with excellent resolution
characteristics of pinhole colimator
cone shpaed
apex faces patient
thick lead sides
removable tungsten aperature
poor sensitivity
image is reverse
magnification changes with distance from aperature
parallax errors
occur in pinhole collimators
occur because the apparent location of an object changes with the viewing angle
How does increasing the distance between a patient and parallel-hole collimator affect resolution and sensitivity?
spatial resolution decreases
image becomes blurrier
senstivity/count rate changes very slowly with distance
Predict how changing from an all-purpose to a high-resolution collimator affects the time needed to reach the same counts.
imaging time increases
high resolution has smaller holes so fewer photons reach crystal and decreases sensitivity