Types of information provided by Doppler ultrasound
Presence of flow
Direction of flow
Speed of flow
Character of flow
Doppler effect
The change in frequency (and wavelength) caused by motion of a sound source, receiver of reflector
Doppler shift (fD)
The difference between the frequency of the received sound and the frequency of the emitted sound
Doppler shift (fD) =
fR - fT
Doppler shift can determine
Direction and velocity
Doppler shift units
kHz
Doppler equation
fD = 2/c x v x fT x cos(angle)
Doppler modes
Color, power, tissue
Doppler modes cw or pw?
Pulsed wave so location Information can be displayed
Provides qualitative info
Spectral Doppler modes
Pulsed wave, pulsed wave tissue Doppler, continuous wave
Quantitative information
“Spenctral” refers to
Graphical representation of velocity (y axis) over time (x axis)
Doppler angle
Angle between the direction of sound propagation and the direction of flow
Should always be less than 60°
What happens when flow is perpendicular to the sound beam (90°)
No shift would be detected so no flow would be displayed
How to avoid perpendicular incidence
Steer the beam using sequencing or phasing
True or false:steering the color box changes the Doppler angle
True
Sample volume/gate
Region selected where Doppler information will be displayed
CW sample gate
Region of overlap between emitting and receiving beams
Color sample gate
Region of color box
PW spectral Doppler sample gate
Position and size of sample gate
Wall filter/reject
Filter out unwanted motion from the vessel walls
Wall thump/clutter
Movement of vessel walls
Tend to be low velocity, but create very high amplitude reflections
True or false: wall filter is only applicable for color Doppler
False
It is applicable to all Doppler modes
Spectral Doppler flow direction
Above baseline- toward transducer
Below baseline- away from transducer
Invert
Flips color map and spectrum
Autocorrelation
A mathematical process that rapidly determines shift information for each sample time (location) along the scan line
Autocorrelation is associated with what Doppler
Color Doppler
Mean
Average
Variance
The spread around the mean
How many Doppler samples (locations) are used per scan line
100-400
Hue
Color
Saturation
Intensity of color
Expressed as it differs from white
luminance
Brightness
Color map
Gives information about sign, magnitude and variance of the shifts
Color map hue
Determines sign of the shift (direction)
Color map saturation or luminance
shows shift magnitudes (velocity)
Variance map
Represents a change in hue from left to right across the map
Ensemble length/ packet size
Number of pulses used for each color scan line
Common ensemble length
10-20
Higher ensemble length provides
More accurate estimates of mean shift
Improved detecting of slow flow
More complete display of flow information
Relationship between ensemble length and FR
Inverse
Color Doppler persistence
Frame averaging technique that reduces noise similarly to gray scale
Relationship between color box width and FR
Inverse
Do we want to use a bigger or smaller color box
Smallest color box possible
Color priority
Priority selects the gray scale echo strength below which color (instead of gray) will be displayed at each pixel location
Low color priority
Weak
Little color displayed
High color priority
More color displayed
PRF/scale sets the limit at the ______
Color bar extremes
For color Doppler, is PRF controlled by operator/sonographer
Yes, sonographer can adjust PRF scale
Lower PRF scale color Doppler
Allows for display of slow flow
Higher PRF scale color Doppler
Shows high velocity without aliasing
Nyquist limit =
PRF / 2
Color baseline
Black area on color map where no shifts detected
Change in direction
Can color a baseline be adjusted
Yes, can be shifted up or down
Color inversion
Switched color assignment to other side of baseline on color map
Color Doppler gain
Amplification of Doppler echoes
Color Doppler limitations
Angle dependent
Low FR
Cannot use multiple foci
Displays mean shifts, not peak
Gives qualitative information, cannot give quantitative measurements
Color interpretation
Allows us to determine flow abnormalities
(Stenosis, blood clot, plaque)
Power Doppler
Color encodes the strength (power intensity) of the shifts
What is the power of the shifts determined by
The concentration of moving scatterers (blood cells)
What is power Doppler helpful for
Showing presence of flow
Power Doppler advantages
Angle independence
No aliasing
Improved sensitivity
Power Doppler disadvantages
No directional information
No flow speed information
No flow character information
Tissue Doppler imaging (TDI)
Color encodes movement of tissue
Angle dependent
Can you use pulsed wave spectral Doppler to measure velocities of tissue motion?
Yes you can
Is continuous wave damped
No, air is the backing material
Does pulsed wave detect complete Doppler shift
No, obtains samples of it
Continuous wave advantages
No aliasing- shows high velocity
Can use high frequencies
Sensitive to low velocities
continuous wave disadvantages
No depth selectivity
Potential overlap of Doppler signals
Pulsed wave advantages
Depth selectivity
Adjustable sample volume size
Pulsed wave disadvantages
Aliasing limits max velocity
Doppler shift detector aka
Quadrate phase detector
Doppler shift detector
Identifies forward and reverse components of the Doppler signal
Determines direction of motion
Fast Fourier Transform (FFT)
Mathematical process to present Doppler shifts in frequency order in a Doppler frequency spectrum
Breaks down the Doppler signals into individual frequencies
Displays frequencies as vertical lines
Spectral display brightness
Intensity if the echo
Spectral Doppler scale -PRF
Adjusts the velocity displayed in the y axis
When do you increase spectral Doppler scale
For higher velocity or to eliminate aliasing
When do you decrease spectral Doppler scale
For low velocity flow
What does adjusting spectral Doppler baseline correct?
Corrects for aliasing
Spectral Doppler gain
A,polities the Doppler shifts and makes the waveform appear brighter
When to increase spectral Doppler gain
to show low amplitude velocities
When to decrease spectral Doppler gain
To eliminate noise
Evaluation of spectral Doppler waveforms include
Direction of flow
Velocity of flow
Type and character of flow
Various measurements (RI, AT, PI)
Spectral window
Black area under a waveform
Narrow range of velocities
Spectral broadening
Widening of the Doppler shift spectrum
Filling in of the spectral window
Broad range of velocities
causes of spectral broadening
Disturbed or turbulent flow
Large sample volume

Type of flow
Laminar

Type of flow
Disturbed

Type of flow
Turbulent

Type of flow
Laminar

Type of flow
Disturbed

type of flow
Turbulent
Peak systolic velocity (PSV)
Highest peak of a waveform corresponding to cardiac systole
End diastolic velocity (EDV)
Velocity at the end of diastole, just before systole
Acceleration time (AT)
Time between beginning of systolic flow and its peak velocity, useful for predicting significant stenosis
Doppler indices
Provides info about vascular resistance and compliance
Resistive index (RI) =
PSV - EDV / PSV
Pulsatility index (PI) =
PSV - EDV / mean velocity
Antegrade
Forward flow
Flows in normal direction for the artery being evaluated
Retrograde
Reverse flow
Flow opposite to the normal direction for the artery being evaluated
Bidirectional
“To-fro”
Blood flow enters and
Eaves a contained space via the same orifice