Types of information provided by Doppler ultrasound

Presence of flow

Direction of flow

Speed of flow

Character of flow

2
New cards

Doppler effect

The change in frequency (and wavelength) caused by motion of a sound source, receiver of reflector

3
New cards

Doppler shift (fD)

The difference between the frequency of the received sound and the frequency of the emitted sound

4
New cards

Doppler shift (fD) =

fR - fT

5
New cards

Doppler shift can determine

Direction and velocity

6
New cards

Doppler shift units

kHz

7
New cards

Doppler equation

fD = 2/c x v x fT x cos(angle)

8
New cards

Doppler modes

Color, power, tissue

9
New cards

Doppler modes cw or pw?

Pulsed wave so location Information can be displayed

Provides qualitative info

10
New cards

Spectral Doppler modes

Pulsed wave, pulsed wave tissue Doppler, continuous wave

Quantitative information

11
New cards

“Spenctral” refers to

Graphical representation of velocity (y axis) over time (x axis)

12
New cards

Doppler angle

Angle between the direction of sound propagation and the direction of flow

Should always be less than 60°

13
New cards

What happens when flow is perpendicular to the sound beam (90°)

No shift would be detected so no flow would be displayed

14
New cards

How to avoid perpendicular incidence

Steer the beam using sequencing or phasing

15
New cards

True or false:steering the color box changes the Doppler angle

True

16
New cards

Sample volume/gate

Region selected where Doppler information will be displayed

17
New cards

CW sample gate

Region of overlap between emitting and receiving beams

18
New cards

Color sample gate

Region of color box

19
New cards

PW spectral Doppler sample gate

Position and size of sample gate

20
New cards

Wall filter/reject

Filter out unwanted motion from the vessel walls

21
New cards

Wall thump/clutter

Movement of vessel walls

Tend to be low velocity, but create very high amplitude reflections

22
New cards

True or false: wall filter is only applicable for color Doppler

False

It is applicable to all Doppler modes

23
New cards

Spectral Doppler flow direction

Above baseline- toward transducer

Below baseline- away from transducer

24
New cards

Invert

Flips color map and spectrum

25
New cards

Autocorrelation

A mathematical process that rapidly determines shift information for each sample time (location) along the scan line

26
New cards

Autocorrelation is associated with what Doppler

Color Doppler

27
New cards

Mean

Average

28
New cards

Variance

The spread around the mean

29
New cards

How many Doppler samples (locations) are used per scan line

100-400

30
New cards

Hue

Color

31
New cards

Saturation

Intensity of color

Expressed as it differs from white

32
New cards

luminance

Brightness

33
New cards

Color map

Gives information about sign, magnitude and variance of the shifts

34
New cards

Color map hue

Determines sign of the shift (direction)

35
New cards

Color map saturation or luminance

shows shift magnitudes (velocity)

36
New cards

Variance map

Represents a change in hue from left to right across the map

37
New cards

Ensemble length/ packet size

Number of pulses used for each color scan line

38
New cards

Common ensemble length

10-20

39
New cards

Higher ensemble length provides

More accurate estimates of mean shift

Improved detecting of slow flow

More complete display of flow information

40
New cards

Relationship between ensemble length and FR

Inverse

41
New cards

Color Doppler persistence

Frame averaging technique that reduces noise similarly to gray scale

42
New cards

Relationship between color box width and FR

Inverse

43
New cards

Do we want to use a bigger or smaller color box

Smallest color box possible

44
New cards

Color priority

Priority selects the gray scale echo strength below which color (instead of gray) will be displayed at each pixel location

45
New cards

Low color priority

Weak

Little color displayed

46
New cards

High color priority

More color displayed

47
New cards

PRF/scale sets the limit at the ______

Color bar extremes

48
New cards

For color Doppler, is PRF controlled by operator/sonographer

Yes, sonographer can adjust PRF scale

49
New cards

Lower PRF scale color Doppler

Allows for display of slow flow

50
New cards

Higher PRF scale color Doppler

Shows high velocity without aliasing

51
New cards

Nyquist limit =

PRF / 2

52
New cards

Color baseline

Black area on color map where no shifts detected

Change in direction

53
New cards

Can color a baseline be adjusted

Yes, can be shifted up or down

54
New cards

Color inversion

Switched color assignment to other side of baseline on color map

55
New cards

Color Doppler gain

Amplification of Doppler echoes

56
New cards

Color Doppler limitations

Angle dependent

Low FR

Cannot use multiple foci

Displays mean shifts, not peak

Gives qualitative information, cannot give quantitative measurements

57
New cards

Color interpretation

Allows us to determine flow abnormalities

(Stenosis, blood clot, plaque)

58
New cards

Power Doppler

Color encodes the strength (power intensity) of the shifts

59
New cards

What is the power of the shifts determined by

The concentration of moving scatterers (blood cells)

60
New cards

What is power Doppler helpful for

Showing presence of flow

61
New cards

Power Doppler advantages

Angle independence

No aliasing

Improved sensitivity

62
New cards

Power Doppler disadvantages

No directional information

No flow speed information

No flow character information

63
New cards

Tissue Doppler imaging (TDI)

Color encodes movement of tissue

Angle dependent

64
New cards

Can you use pulsed wave spectral Doppler to measure velocities of tissue motion?

Yes you can

65
New cards

Is continuous wave damped

No, air is the backing material

66
New cards

Does pulsed wave detect complete Doppler shift

No, obtains samples of it

67
New cards

Continuous wave advantages

No aliasing- shows high velocity

Can use high frequencies

Sensitive to low velocities

68
New cards

continuous wave disadvantages

No depth selectivity

Potential overlap of Doppler signals

69
New cards

Pulsed wave advantages

Depth selectivity

Adjustable sample volume size

70
New cards

Pulsed wave disadvantages

Aliasing limits max velocity

71
New cards

Doppler shift detector aka

Quadrate phase detector

72
New cards

Doppler shift detector

Identifies forward and reverse components of the Doppler signal

Determines direction of motion

73
New cards

Fast Fourier Transform (FFT)

Mathematical process to present Doppler shifts in frequency order in a Doppler frequency spectrum

Breaks down the Doppler signals into individual frequencies

Displays frequencies as vertical lines

74
New cards

Spectral display brightness

Intensity if the echo

75
New cards

Spectral Doppler scale -PRF

Adjusts the velocity displayed in the y axis

76
New cards

When do you increase spectral Doppler scale

For higher velocity or to eliminate aliasing

77
New cards

When do you decrease spectral Doppler scale

For low velocity flow

78
New cards

What does adjusting spectral Doppler baseline correct?

Corrects for aliasing

79
New cards

Spectral Doppler gain

A,polities the Doppler shifts and makes the waveform appear brighter

80
New cards

When to increase spectral Doppler gain

to show low amplitude velocities

81
New cards

When to decrease spectral Doppler gain

To eliminate noise

82
New cards

Evaluation of spectral Doppler waveforms include

Direction of flow

Velocity of flow

Type and character of flow

Various measurements (RI, AT, PI)

83
New cards

Spectral window

Black area under a waveform

Narrow range of velocities

84
New cards

Spectral broadening

Widening of the Doppler shift spectrum

Filling in of the spectral window

Broad range of velocities

85
New cards

causes of spectral broadening

Disturbed or turbulent flow

Large sample volume

86
New cards
<p>Type of flow</p>

Type of flow

Laminar

87
New cards
<p>Type of flow</p>

Type of flow

Disturbed

88
New cards
<p>Type of flow</p>

Type of flow

Turbulent

89
New cards
<p>Type of flow</p>

Type of flow

Laminar

90
New cards
<p>Type of flow</p>

Type of flow

Disturbed

91
New cards
<p>type of flow</p>

type of flow

Turbulent

92
New cards

Peak systolic velocity (PSV)

Highest peak of a waveform corresponding to cardiac systole

93
New cards

End diastolic velocity (EDV)

Velocity at the end of diastole, just before systole

94
New cards

Acceleration time (AT)

Time between beginning of systolic flow and its peak velocity, useful for predicting significant stenosis

95
New cards

Doppler indices

Provides info about vascular resistance and compliance

96
New cards

Resistive index (RI) =

PSV - EDV / PSV

97
New cards

Pulsatility index (PI) =

PSV - EDV / mean velocity

98
New cards

Antegrade

Forward flow

Flows in normal direction for the artery being evaluated

99
New cards

Retrograde

Reverse flow

Flow opposite to the normal direction for the artery being evaluated

100
New cards

Bidirectional

“To-fro”

Blood flow enters and

Eaves a contained space via the same orifice