MU/time =

Dose/dose rate

2
New cards

What do you use to know the dose at the point?

The PDD for SSD and TPR for SAD

3
New cards

Machine output is a measure of the amount of radiation produced by a unit; usually stated as ___ rate or _____

exposure rate

absorbed dose rate

4
New cards

What is included in absorbed dose?

Primary dose and dose due to scatter from the medium

5
New cards

____ is used if in air

___ is used if in a medium

Exposure rate

Absorbed dose rate

6
New cards

____ is useful when using TAR and is only useful for beam energies up to Co60

Units

Exposure rate

R/min

7
New cards

___ is the measure of ionization in air by photons

Units

Exposure

Roentgen

8
New cards

Exposure rate varies with?

Energy, distance, field size (collimator scatter not patient; no patient)

9
New cards

____ is the energy imparted by ionizing radiation to a medium

Units

Absorbed dose

Gray

1 Gy = 1j/kg

<p>Absorbed dose</p><p>Gray</p><p>1 Gy = 1j/kg</p>
10
New cards

What is f?

Units

The roentgen to rad conversion factor/ exposure to absorbed dose factor

A function of the medium composition and photon energy (depends on energy and type of medium ex bone, muscle, air)

cGy/R

11
New cards

A is always

Corrects for the presence of a

1

Medium

12
New cards

Use f = ___ unless indicated otherwise

0.95cGy/R

13
New cards

___ are measured in a phantom, are referred to as output, needed for higher energy machines

Useful when using PDD or TPR or TMR

Measured at the standard SSD at the depth of dmax in a phantom

Can be measured at any depth

Absorbed dose rate

14
New cards

What is the RDF?

Account for the change in ___ that contributes to the dose rate

-collimator

-patient

Ratio of the dose rate of a given field size to the dose rate of the reference field size, usually 10x10

Scatter

15
New cards

RDF is divided into?

Collimator scatter factor

Phantom scatter factor (scatter from patient)

16
New cards

____ is the ratio of the output in air for a given field size to that for a reference field size (10x10)

Collimator scatter factor

17
New cards

As the field size increases, the interactions against the collimator increases, causing more scatter

This reaches a threshold beyond which?

The field size will not contribute more scatter

18
New cards

For photon beams wherte BSF can be accurately measured, the patient scatter factor Sp may be defined as

The ratio of BSF

BSF (field size)/ BSF (ref)

19
New cards

When can you use Sp without Sc?

At extended SSD (collimators don't change)

Field has significant blocking

20
New cards

The relative exposure factor is a sum of

Sp and Sc

21
New cards

Dosimetric functions used for entire photon energy range

PDD

RDF

Sp, Sc

22
New cards

Dosimetric functions used at Co-60 and below

BSF

TAR

SAR

23
New cards

Dosimetric functions used at Co-60 and above

TMR

TPR

SMR

24
New cards

The factor of converting exposure to absorbed dose is known as the

F factor

25
New cards

Relative exposure factor is only used when there is a change in?

The diaphragm settings

26
New cards

A TAR at the depth of dmax used to correct for scatter of dose is called the

Backscatter factor