MU/time =
Dose/dose rate
What do you use to know the dose at the point?
The PDD for SSD and TPR for SAD
Machine output is a measure of the amount of radiation produced by a unit; usually stated as ___ rate or _____
exposure rate
absorbed dose rate
What is included in absorbed dose?
Primary dose and dose due to scatter from the medium
____ is used if in air
___ is used if in a medium
Exposure rate
Absorbed dose rate
____ is useful when using TAR and is only useful for beam energies up to Co60
Units
Exposure rate
R/min
___ is the measure of ionization in air by photons
Units
Exposure
Roentgen
Exposure rate varies with?
Energy, distance, field size (collimator scatter not patient; no patient)
____ is the energy imparted by ionizing radiation to a medium
Units
Absorbed dose
Gray
1 Gy = 1j/kg

What is f?
Units
The roentgen to rad conversion factor/ exposure to absorbed dose factor
A function of the medium composition and photon energy (depends on energy and type of medium ex bone, muscle, air)
cGy/R
A is always
Corrects for the presence of a
1
Medium
Use f = ___ unless indicated otherwise
0.95cGy/R
___ are measured in a phantom, are referred to as output, needed for higher energy machines
Useful when using PDD or TPR or TMR
Measured at the standard SSD at the depth of dmax in a phantom
Can be measured at any depth
Absorbed dose rate
What is the RDF?
Account for the change in ___ that contributes to the dose rate
-collimator
-patient
Ratio of the dose rate of a given field size to the dose rate of the reference field size, usually 10x10
Scatter
RDF is divided into?
Collimator scatter factor
Phantom scatter factor (scatter from patient)
____ is the ratio of the output in air for a given field size to that for a reference field size (10x10)
Collimator scatter factor
As the field size increases, the interactions against the collimator increases, causing more scatter
This reaches a threshold beyond which?
The field size will not contribute more scatter
For photon beams wherte BSF can be accurately measured, the patient scatter factor Sp may be defined as
The ratio of BSF
BSF (field size)/ BSF (ref)
When can you use Sp without Sc?
At extended SSD (collimators don't change)
Field has significant blocking
The relative exposure factor is a sum of
Sp and Sc
Dosimetric functions used for entire photon energy range
PDD
RDF
Sp, Sc
Dosimetric functions used at Co-60 and below
BSF
TAR
SAR
Dosimetric functions used at Co-60 and above
TMR
TPR
SMR
The factor of converting exposure to absorbed dose is known as the
F factor
Relative exposure factor is only used when there is a change in?
The diaphragm settings
A TAR at the depth of dmax used to correct for scatter of dose is called the
Backscatter factor