rate response goals
maintain minimum HR
maintain AV synchrony
provide appropriate HR variability
not every patient needs it
all about r-wave
max sensor rate
max rate that paces atrium
patients with CI
single/dual/CRT: device with sinus brady therapy
max track rate
max rate that paces ventricles
rate at which upper rate behaviors begin to occur
patients with AV block
dual/CRT: only for device when AV synchrony is goal (VDD/DDD)
upper rate behaviors
pacemaker wenckebach and 2:1 block)
chronoropic incomepetence
unable to obtain adequate HR increase with exercise or stress
what percent of patients have CI
30%
causes of chronotropic incompetence
sinus node dysfunction
autonomic dysfunction
AFib with/without CHB
beta blocker
heart transplant
where to program MSR for elderly PPM
age
chronotropic competence
no need for rate response
cardiac output equation
heart rate x stroke volume
which devices help SV
CRT not single or dual devices
how to increase CO
increase HR; CI patients cannot
what percent of AV synchrony contributes to CO
20-30%
what percent of CI in HF and CRT patients
45% HF and 70% CRT
which patients are good candidates for rate response
poor LV function
increase HR to help patient with low SV
non-exercise needs of rate reponse
mental stress
emotional response
anticipation
monitoring devices and time
loop 24-48 hr
mobile cardiac telemetry 30day
loop 2-6yr
how to establish need to activate rate response
indication/patient history
rate histogram (holter, mct, loop)
heart rate trend (holter, mct, loop)
activity level (holter, mct, loop)
symptoms
hall walk
[physician preferences
ideal HR variability histogram
step
as HR increases, time decreases
how to identify CI
exercise stress test prior to implant
CIED after implant
poor HR distribution
blunted; no variability
how to change blunted historgram in DDD
turn on rate variability and adjust settings
AFib histogram
right shifted blunted atrium
good stair step and bell curve in ventricle
do atrium and ventricle histogram always match
no
heart rate trends
show A and V HR overtime (dual or single)
overall average
day vs night
rate response sensor must
detect physical or psycholigocal parameter related to metabolic demand
Algorithm that relates changes in sensed parameter to change in pacing rate
must be able to achieve clinically desired rate
sensor needs to closely resemble synthetic sinus node
what must a sensor closely resemble
synthetic sinus node
open loop sensor
abbott, boston, medtronic
change in physiological/physical parameter
sensor
algorithm (physician input)
change rate
closed loop sensor
biotronik only
change in rate to change physiological parameter to sensor to algorithm to change rate
major classes of sensors
vibration and acceleration
impedance (minute ventilation and closed loop)
vibration and acceleration sensing
activity (every company)
impedance sensing
minute ventilation (BSX only)
ventricular contractility (Closed loop sensing BIO only)
activity sensing
body movements create vibrations detected by sensor inside device
3 types of activity sensors
piezoelectric crystal
piezoelectric accelerometer
Piezoresistive accelrometer
piezoelectric sensor
electric charge accumulates in crystals
piezoelectric crystal
vibrations deform crycstal which induce voltage proportional to amount of deformation
piezoelectric accelerometer
still use
measures rate in change of velocity
piezoresistive accelerometer
still use
measures changes in electrical resistances
triaxially measured changes
activity sensing based off movement (sideways, up/down, in/out)
programmable parameters (mode and 3 things)
DDDR, DDIR, AAIR, VVIR (APVS or APVP)
sensor thresholds- how much movement is seen; sensor has nothing to do with sensing)
rate response slopes
reaction and recovery time
passive mode
sensor programmed to this to tell what the histogram could look like if you turn on rate response
collects data
MDT sensor threshold parameter
low, medium/low, medium/high, high
abbott sensor threshold parameter
threshold 1-7
1= most sensitive
7= least sensitive
threshold: manual or automatic
boston threshold parameter
very low, low, medium low, medium, medium high, high, very high
low/1
detect low movement/activity with minimal exertion for sedentary patients
high/7
requires vigorous body activities for active individuals
how to make more sensitive
bring sensor threshold in to increase rate response
ex: medium/low to low
sensor slope
how fast you increase HR
how fast/slow you reach MSR
linear or triphasic curve
reaction time- activity acceleration parameter
initial timeframe to determine when to begin increasing sensor rate (30sec nominal)
recovery time- activity deceleration parameter
determines timeframe for pacing rate to decrease down to target rate or LRL
which companies only have accelerometer
medtronic and abbott
accelerometer challenges
movement
rocking chair or bumpy road
stationary bike
boston scientific minute ventilation
transthoracic impedance (fluid monitor) changes between inspiration and expiration
frequent measurements of thoracic impedance allow alteration of pacing rate
goal is to detect increased metabolic demand even in the absence of physical activity
not reliant upon motion
Ohm’s law
V = I x R
voltage = current (amp) x resistance (ohm)
minute ventilation uses microamps
minute ventilation equation
respiration rate (speed) x tidal volume (depth)
boston scientific minute ventilation devices
dual sensor requires input from accelerometer
ventilation threshold chosen according to age, gender, four level of patient fitness
arm exercises on same side as ppm will cause change in impedance not related to respiratory changes leading to faster pacing
pulmonary issues (asthma and COPD) cannot use MV or with intubation
requires bipolar lead
CLS receptors
aortic arch and carotid artery
how CLS works
autonomic tone changes to determine rate response
impedance changes in cardiac tissue
myocardium inotropy: squeeze harder and autonomic tone changes
goal: squeeze harder and beat faster
stroke volume
blood leaving LV
CLS mode and more
DDD-CLS not DDDR to designate rate response
impedance rises with contraction
self calibrates week worth HB to adjust
sub-threshold measurements
extra sensitive telemetry
CLS reference waveform
baseline impedance waveform and exercise impedance waveform converted into sensor pacing rate
magnitude of rate response determined by programmable CLS-response parameter
resting rate control- CLS
pacing during autonomic changes in absence of activity
cannot faint
CLS for neurocardiogenic stroke
myocardial contractility changes occur before syncope
cls responds quickly to increase heart rate before syncope
can forsee 15min prior
medtronic micra leadless
detects motion and atrial mechanical sensing
VDD: atrial mechanical sensing/AV synchronous
VDIR during mode switch
VVIR non-AV synchronous
triphasic rate response curve
LRL, ADLR, SURL
applied when in VDIR/VVIR
atrial mechanical sensing (MDT Micra)
accelerometer not used for rate response in VDD mode but used for sensing atrial motion
filter cardiac from body movement
abbott leadless ppm
temperature sensor (0.01)
initially drops in temp due to peripheral blood rushing into RA
as blood circulates, temp rises then device reacts based on parameter settings
patient cant get HR up while walking
activity threshold too high
patient feels like heart racing while walking
activity threshold too low
patient HR escalates quickly just by walking to door
slope too high
patient HR take long time to increase during exercise
slope too low
patient experiences DOE at first but feels better after awhile
reaction time too slow
patient feels light headed when sitting down after exertion
recovery time too quick