rate response goals

maintain minimum HR

maintain AV synchrony

provide appropriate HR variability

not every patient needs it

all about r-wave

2
New cards

max sensor rate

max rate that paces atrium

patients with CI

single/dual/CRT: device with sinus brady therapy

3
New cards

max track rate

max rate that paces ventricles

rate at which upper rate behaviors begin to occur

patients with AV block

dual/CRT: only for device when AV synchrony is goal (VDD/DDD)

4
New cards

upper rate behaviors

pacemaker wenckebach and 2:1 block)

5
New cards

chronoropic incomepetence

unable to obtain adequate HR increase with exercise or stress

6
New cards

what percent of patients have CI

30%

7
New cards

causes of chronotropic incompetence

sinus node dysfunction

autonomic dysfunction

AFib with/without CHB

beta blocker

heart transplant

8
New cards

where to program MSR for elderly PPM

age

9
New cards

chronotropic competence

no need for rate response

10
New cards

cardiac output equation

heart rate x stroke volume

11
New cards

which devices help SV

CRT not single or dual devices

12
New cards

how to increase CO

increase HR; CI patients cannot

13
New cards

what percent of AV synchrony contributes to CO

20-30%

14
New cards

what percent of CI in HF and CRT patients

45% HF and 70% CRT

15
New cards

which patients are good candidates for rate response

poor LV function

increase HR to help patient with low SV

16
New cards

non-exercise needs of rate reponse

mental stress

emotional response

anticipation

17
New cards

monitoring devices and time

loop 24-48 hr

mobile cardiac telemetry 30day

loop 2-6yr

18
New cards

how to establish need to activate rate response

indication/patient history

rate histogram (holter, mct, loop)

heart rate trend (holter, mct, loop)

activity level (holter, mct, loop)

symptoms

hall walk

[physician preferences

19
New cards

ideal HR variability histogram

step

as HR increases, time decreases

20
New cards

how to identify CI

exercise stress test prior to implant

CIED after implant

21
New cards

poor HR distribution

blunted; no variability

22
New cards

how to change blunted historgram in DDD

turn on rate variability and adjust settings

23
New cards

AFib histogram

right shifted blunted atrium

good stair step and bell curve in ventricle

24
New cards

do atrium and ventricle histogram always match

no

25
New cards

heart rate trends

show A and V HR overtime (dual or single)

overall average

day vs night

26
New cards

rate response sensor must

detect physical or psycholigocal parameter related to metabolic demand

Algorithm that relates changes in sensed parameter to change in pacing rate

must be able to achieve clinically desired rate

sensor needs to closely resemble synthetic sinus node

27
New cards

what must a sensor closely resemble

synthetic sinus node

28
New cards

open loop sensor

abbott, boston, medtronic

change in physiological/physical parameter

sensor

algorithm (physician input)

change rate

29
New cards

closed loop sensor

biotronik only

change in rate to change physiological parameter to sensor to algorithm to change rate

30
New cards

major classes of sensors

vibration and acceleration

impedance (minute ventilation and closed loop)

31
New cards

vibration and acceleration sensing

activity (every company)

32
New cards

impedance sensing

minute ventilation (BSX only)

ventricular contractility (Closed loop sensing BIO only)

33
New cards

activity sensing

body movements create vibrations detected by sensor inside device

34
New cards

3 types of activity sensors

piezoelectric crystal

piezoelectric accelerometer

Piezoresistive accelrometer

35
New cards

piezoelectric sensor

electric charge accumulates in crystals

36
New cards

piezoelectric crystal

vibrations deform crycstal which induce voltage proportional to amount of deformation

37
New cards

piezoelectric accelerometer

still use

measures rate in change of velocity

38
New cards

piezoresistive accelerometer

still use

measures changes in electrical resistances

39
New cards

triaxially measured changes

activity sensing based off movement (sideways, up/down, in/out)

40
New cards

programmable parameters (mode and 3 things)

DDDR, DDIR, AAIR, VVIR (APVS or APVP)

sensor thresholds- how much movement is seen; sensor has nothing to do with sensing)

rate response slopes

reaction and recovery time

41
New cards

passive mode

sensor programmed to this to tell what the histogram could look like if you turn on rate response

collects data

42
New cards

MDT sensor threshold parameter

low, medium/low, medium/high, high

43
New cards

abbott sensor threshold parameter

threshold 1-7

1= most sensitive

7= least sensitive

threshold: manual or automatic

44
New cards

boston threshold parameter

very low, low, medium low, medium, medium high, high, very high

45
New cards

low/1

detect low movement/activity with minimal exertion for sedentary patients

46
New cards

high/7

requires vigorous body activities for active individuals

47
New cards

how to make more sensitive

bring sensor threshold in to increase rate response

ex: medium/low to low

48
New cards

sensor slope

how fast you increase HR

how fast/slow you reach MSR

linear or triphasic curve

49
New cards

reaction time- activity acceleration parameter

initial timeframe to determine when to begin increasing sensor rate (30sec nominal)

50
New cards

recovery time- activity deceleration parameter

determines timeframe for pacing rate to decrease down to target rate or LRL

51
New cards

which companies only have accelerometer

medtronic and abbott

52
New cards

accelerometer challenges

movement

rocking chair or bumpy road

stationary bike

53
New cards

boston scientific minute ventilation

transthoracic impedance (fluid monitor) changes between inspiration and expiration

frequent measurements of thoracic impedance allow alteration of pacing rate

goal is to detect increased metabolic demand even in the absence of physical activity

not reliant upon motion

54
New cards

Ohm’s law

V = I x R

voltage = current (amp) x resistance (ohm)

minute ventilation uses microamps

55
New cards

minute ventilation equation

respiration rate (speed) x tidal volume (depth)

56
New cards

boston scientific minute ventilation devices

dual sensor requires input from accelerometer

ventilation threshold chosen according to age, gender, four level of patient fitness

arm exercises on same side as ppm will cause change in impedance not related to respiratory changes leading to faster pacing

pulmonary issues (asthma and COPD) cannot use MV or with intubation

requires bipolar lead

57
New cards

CLS receptors

aortic arch and carotid artery

58
New cards

how CLS works

autonomic tone changes to determine rate response

impedance changes in cardiac tissue

myocardium inotropy: squeeze harder and autonomic tone changes

goal: squeeze harder and beat faster

59
New cards

stroke volume

blood leaving LV

60
New cards

CLS mode and more

DDD-CLS not DDDR to designate rate response

impedance rises with contraction

self calibrates week worth HB to adjust

sub-threshold measurements

extra sensitive telemetry

61
New cards

CLS reference waveform

baseline impedance waveform and exercise impedance waveform converted into sensor pacing rate

magnitude of rate response determined by programmable CLS-response parameter

62
New cards

resting rate control- CLS

pacing during autonomic changes in absence of activity

cannot faint

63
New cards

CLS for neurocardiogenic stroke

myocardial contractility changes occur before syncope

cls responds quickly to increase heart rate before syncope

can forsee 15min prior

64
New cards

medtronic micra leadless

detects motion and atrial mechanical sensing

VDD: atrial mechanical sensing/AV synchronous

  • VDIR during mode switch

  • VVIR non-AV synchronous

triphasic rate response curve

LRL, ADLR, SURL

applied when in VDIR/VVIR

65
New cards

atrial mechanical sensing (MDT Micra)

accelerometer not used for rate response in VDD mode but used for sensing atrial motion

filter cardiac from body movement

66
New cards

abbott leadless ppm

temperature sensor (0.01)

initially drops in temp due to peripheral blood rushing into RA

as blood circulates, temp rises then device reacts based on parameter settings

67
New cards

patient cant get HR up while walking

activity threshold too high

68
New cards

patient feels like heart racing while walking

activity threshold too low

69
New cards

patient HR escalates quickly just by walking to door

slope too high

70
New cards

patient HR take long time to increase during exercise

slope too low

71
New cards

patient experiences DOE at first but feels better after awhile

reaction time too slow

72
New cards

patient feels light headed when sitting down after exertion

recovery time too quick