doppler shift formula
(2 x velocity of blood x transducer frequency x cos θ)/ propagation speed
demodulation definition
the process of extracting the low doppler frequency from the transducer's carrier frequency in order to isolate it as the "doppler shift"
positive doppler shift definition
blood cells moving towards the transducer (antegrade), the reflected frequency is HIGHER than the transmitted frequency
negative doppler shift definition
blood cells moving away from the transducer (retrograde), reflected frequency is LOWER than the transmitted frequency
doppler frequencies indicate _______, not _______.
velocity not speed
speed definition and units
purely a magnitude, indicates the distance that a red blood cell moves in one second; units are any distance divided by time
velocity definition
magnitude and direction
doppler shift is _______ related to the velocity of the blood cells
directly
why is there a 2 in the doppler equation?
two doppler shifts occur during a clinical US exam; the first when sound strikes moving blood cells (transmission) and the second at the transducers reception of the sound wave from the moving blood cells
doppler shift is _______ related to the frequency of the transmitted sound
directly
if the transducer frequency is doubled, how will it affect the measured doppler shift
it will also be doubled
measured velocity formula
true velocity x cos θ
doppler shift is ______ related to the cosine of the angle between the direction of flow and the direction of sound
directly
if the cosine θ is halved, how will the measured doppler shift be affected
it will also be halved
cosine 0
1.0
cosine 30º
0.87
cosine 60º
0.5
cosine 90º
0
cosine 120º
-0.5
cosine 150º
-0.87
cosine 180º
-1.0
phase quadrature definition
commonly used signal processing technique for bidirectional doppler (display graphically like spectral tracing)
Phase quadrature is also known as
quadrature detection
continuous wave doppler advantage and disadvantage
advantage is that it accurately measures high velocities; disadvantage is range ambiguity
how many crystals does a continuous wave doppler contain
two (one transmitting and one receiving)
dedicated continuous wave transducer
-does not create images
-two elements (both shaped like a semi-circle) in which one transmits and one receives
-does not use backing material
duplex imaging definition
simultaneous anatomic imaging and doppler
what are the consequences of CW transducers not using backing material
-undampened transmitted signal
-narrow bandwidth
-high quality factor
-higher sensitivity
pulsed wave doppler advantages and disadvantages
only one PZT necessary; advantages are range resolution/ specificity, disadvantages are aliasing at high velocities; uses backing material
what are the consequences of PW transducers using backing materials
-low quality factor
-lower sensitivity
-wide bandwidth pulses
aliasing definition
false identity; very high velocities in one direction are incorrectly displayed as going in the opposite direction
the top of the display graph represents ___ _______ _____
the nyquist limit
when does aliasing occur
-only with pulsed wave doppler
-when the doppler sampling rate is too low in comparison to the measured blood velocities
nyquist limit formula
PRF/ 2
nyquist limit unit
Hz
1 multiple choice option
how to achieve a higher nyquist limit
-shallower structure
-a higher PRF
how to achieve a lower nyquist limit
-deeper structures
-lower PRF
what conditions create less aliasing
-slower blood velocity
-lower frequency transducer
-shallow SVG
-higher PRF
what conditions create more aliasing
-faster blood velocity
-higher frequency transducer
-deeper SVG
-lower PRF
what five techniques may be used to avoid aliasing artifacts
1. adjust scale to max
2. use a new view where structure is more shallow
3. select lower frequency transducer
4. use baseline shift
5. use continuous wave doppler
at a particular velocity, higher frequency transducers create ____ doppler shifts and lower frequency transducers create ____ doppler shifts
higher; lower
how does increasing the PRF affect the nyquist limit
also increases
what information does color flow doppler provide
info on the location of flow (subject to aliasing) (average/ mean velocities)
color maps definition
converts measured velocities into colors that appear superimposed on the image
what are the two types of color maps
velocity mode and variance mode
velocity mode definition
flow direction and velocity; colors above the baseline are velocities moving toward the transducer and colors below the baseline are moving away from the transducer; colors farther away from the baseline represents faster velocities
variance mode definition
distinguishes laminar and turbulent flow; colors on the left represent laminar flow and colors on the right represent turbulent flow (colors above and below the baseline represent the same as velocity mode)
ghosting/ clutter doppler artifact
very low frequency doppler shift (from slowly moving anatomy, not blood) show on spectral display; can be eliminated by utilizing wall filter
wall filter definition
used to eliminate low frequency doppler shifts from moving anatomy rather than moving blood cells
wall filter is also known as
high pass filters
crosstalk definition
Mirror image artifact on a Spectral Doppler display resulting from the doppler gain being set too high or an incidence angle near 90º
what is spectral analysis
breaking the complex doppler signal down into it's basic "building blocks" and identifies individual velocities that make up a reflected doppler signal
What are the two types of spectral analysis?
-Fast fourier transform (FFT)
-Autocorrelation
fast Fourier transform (FFT) definition
exceedingly accurate way to display all individual velocity components that make up the complex reflected signal; distinguishes laminar and turbulent flow
how will laminar flow appear on the spectral doppler display due to fast Fourier transform (FFT)
distinct velocity profile with spectral window
how will turbulent flow appear on the spectral doppler display due to fast Fourier transform (FFT)
messy waveform with spectral broadening
fast Fourier transform (FFT) is used for
spectral doppler
autocorrelation is used for
color flow doppler
normal incidence
the incident sound beam strikes the boundary at exactly 90 degrees
what makes color appear and not appear in a vessel using color doppler?
an angle other than 90º must be created between the direction of flow and the direction of the sound beam
color doppler gain set too high leads to
color confetti
spectral doppler gain
alters the gray scale content of a spectral display waveform (if too high can create noise)
color aliasing may be reduced by
increasing the Nyquist velocity scale (color scale)
flow reversal
will demonstrate red, black, and blue color flow next to each other on the image (near the baseline of the color map),
aliasing
will demonstrate yellow and light blue next to each other on the image (both farthest from the baseline on the color map)
how are artifacts described
-not real
-not seen on the image
-incorrect shape or size
-incorrect position
-incorrect brightness
cause of artifacts
-violations of assumption
-equipment malfunction
-operator error
-physics of ultrasound
six assumptions of imaging systems
1. sound travels in a straight line
2. sound travels directly to a reflector and back
3. sound travels in soft tissue at 1540 m/sec
4. reflections arise only from structures positioned in the sound beam's main axis
5. the imaging plane is very thing
6. the strength of a reflection is related to the characteristics of the tissue creating the reflections
reverberations definition
multiple equally spaced echoes caused by the bouncing of sound wave between two strong reflectors positioned parallel to the sound beam
reverberation characteristics
appear in multiples, appear equally spaced, located parallel to the sound beam's main axis, located at ever- increasing depths
comet tail definition
solid hyper echoic line directed downward created when closed spaced reverberations merge
comet tail artifact is also known as
ring down artifact
comet tail characteristics
single long hyper echoic echo, located parallel to the sound beam's main axis
3 multiple choice options
shadow artifact definition
hypo echoic or anechoic region extending downward from a highly attenuating structure
When does shadow artifact occur
when attenuation is higher in the tissue above the shadow than in the surrounding tissue
shadow artifact characteristics
-hypo- or anechoic
-result of too much attenuation
-located beneath the structure with abnormally high attenuation
-prevents visualization of true anatomy on the scan
edge shadow artifact definition
special form of shadowing that appears as a hypo echoic region extending down from the edge of a curved structure due to a decrease in intensity
edge shadow artifact characteristics
-hypo- or anechoic
-results when the beam spreads after striking a curved reflector
-extends downward from the curved reflectors edge parallel to the beam
-prevents visualization of true anatomy on the scan
enhancement artifact definition
hyper echoic region beneath tissue with abnormally low attenuation
enhancement artifact characteristics
-hyperechoic
-result of too little attenuation
-located beneath a structure with abnormally low attenuation
focal enhancement definition
special form of enhancement in which a side- to-side region of an image appears brighter than tissues at other depths
focal enhancement is also known as
focal banding
focal enhancement characteristics
-hyper echoic side to side region
-results from increased intensity at the focus
mirror image artifact definition
Duplicates a structure on the opposite side of a strong reflector, appears deeper than the true reflector. The true reflector & artifact are at an equal distance from the mirror
3 multiple choice options
mirror image artifact characteristics
-second copy of a true reflector
-artifact appears deeper than the true reflector
-bright reflector lies on a straight line between artifact and transducer
-true reflector and artifact are equidistant from the mirror
speed error definition
created when a sound wave propagates through a medium at a speed other than that of soft tissue (1540m/s) so the reflector appears at incorrect depths
speed error artifact is also known as
range error artifact
what happens if the medium's speed is slower than that in soft tissue
system assumes the reflectors are far from the transducer and reflectors are placed too deep on an image
what happens if the medium's speed is faster than that in soft tissue
system assumes the reflectors are close to the transducer and reflectors are placed too shallow on an image
speed error characteristics
-correct number of reflectors but improper depth
-appears as a step off
lobe artifact definition
appear when sound energy is transmitted in a direction other than along the sound beams main axis
lobe artifact characteristics
-second copy of the true reflector
-the artifact and the true reflector are located side by side at the same depth
side lobes are created by
a single crystal transducer (mechanical)
grating lobes are created by
array transducers
subdicing definition
Reduces grating lobes by dicing crystal into groups of smaller crystals. Electronically joined
apodization definition
process of differential excitation to reduce grating lobes in which the subdiced elements are excited with different voltages
refraction definition
when sound pulse changes direction during transmission because it strikes a boundary obliquely and the media on either side of the boundary have different propagation speeds
refraction characteristics
-second copy of the reflector
-copy is side by side or at the same depth as the reflector
slice thickness artifact
occurs as a result of the beam not being razor thin; thus unintended echoes may appear in the image as the beam slices through structures adjacent to intended reflectors
slice thickness artifact is also known as
partial volume artifact